Afinion manual us qxd: Layout 1



Yüklə 230,92 Kb.
Pdf görüntüsü
səhifə5/8
tarix08.03.2018
ölçüsü230,92 Kb.
#30910
1   2   3   4   5   6   7   8

16

GETTING STARTED

Setting date

The correct date should always be set because the date and time for the analysis are stored and displayed in

the patient and control records.The date format is YYYY:MM:DD, where YYYY is the year, MM is the month

(01 to 12), and DD is the day (01 to 31).

Configuration menu

Setting time

The correct time should always be set because the date and time for the analysis are stored and displayed in

the patient and control records.The time format is hh:mm, where hh is the hour from 00 to 23 and mm is

minutes from 00 to 59.

Configuration menu

1

Touch


to enter

date/time settings.



2

Touch


to enter

date setting.



3

Enter today's date.

Touch

to confirm



and return to previous view.

1

Touch


to enter

date/time settings.



2

Touch


to enter

time setting.



3

Enter time.

Touch

to confirm and



return to previous view.

09-38555



17

GETTING STARTED

Adjusting screen contrast

The screen contrast can be adjusted.

Configuration menu

Aligning screen

The screen must be realigned if the buttons do not respond accurately when touched. It is not necessary to

align screen as a part of the initial configuration.

Configuration menu



1

Touch


to enter

screen/beeper settings.



2

Touch


to enter

screen contrast setting.



3

Adjust screen contrast by

touching

or

Touch



to accept

and return to previous view.



1

Touch


to enter

screen/beeper settings.



2

Touch


to enter

screen alignment function.



3

Tap the cross-hair object (+)

in the upper left corner using

a blunt pencil to be precise.

Repeat for the object appea-

ring in the lower right corner

and in the center of the screen.

The previous screen view will

automatically return.

09-38555



18

GETTING STARTED

Adjusting beeper volume

One beep means that the test is completed and result displayed.Two beeps means that an information

or error code is displayed.The beeper volume can be adjusted. If adjusted to a minimum, no beep will

be heard.

Configuration menu

How to switch OFF the Analyzer

Switch off the Analyzer by pressing the ON/OFF button (Figure 1).

The Analyzer should be switched off after the end of a working day.

Please note:

• When the power is turned off, a closing down procedure is initiated.The cartridge carriage will move

to a safe position and the display will be active a few seconds until the Analyzer shuts down.

The Analyzer can be switched off, or the power supply disconnected, without loss of stored results.

• The Analyzer can only be switched off when the cartridge chamber is empty and the lid is closed. If the

ON/OFF button is pressed and the lid is open, the message “Close lid” will appear on the screen.



1

Touch


to enter

screen/beeper settings.



2

Touch


to enter

beeper volume setting.



3

Adjust beeper volume by

touching

or

Touch



to accept

and return to previous view.

09-38555



19

QUALITY CONTROL

Why quality control testing?

Quality control testing should be done to confirm that your Afinion™ AS100 Analyzer System is working

properly and providing reliable results. Only when controls are used routinely and the values are within the

acceptable ranges can accurate results for patient samples be assured.



Choosing control material

Controls recommended by Axis-Shield should be used for quality control of your the Afinion™ AS100

Analyzer System.These control kits contain control materials with established acceptable ranges for the

Afinion™ AS100 Analyzer System.



Handling and testing controls

Consult the Package Insert that comes with each control kit for detailed instructions on handling

and storage of the control material.

To run a control, follow the procedure in the section “Testing procedures”, page 20-28.

The measured value should be within the acceptable range stated on the control vial label or in the control

package insert. If the control results are within the acceptable ranges, patient samples may be tested and

results reported.

If the result obtained for a control is outside the acceptable limits, make sure that:

- The control vial has not passed its expiry date.

- The control vial has not passed its open vial expiry date.

- The control vial and Afinion™ Test Cartridges have been stored according to recommendations.

- There is no evidence of bacterial or fungal contamination of the control vial.

Correct any procedural error and re-test the control material. If no procedural errors are detected, it is

recommended to examine the laboratory’s quality control record to investigate the frequency of control

failures. Ensure that there is no trend in out-of-range quality control results. Re-test the control material

using a new control vial.

Patient results must be declared invalid when controls do not perform as expected. Contact your

customer service representative (1-877-4-Afinion or 1-877-423-4646) for advice before analyzing

patient samples.

Frequency of control testing

Controls should be analyzed:

- With each new shipment of Afinion™ test kits.

- With each new lot of Afinion™ test kits.

- Users with a low frequency of testing should analyze controls at least every 30 days.

- When training new operators in correct use of the Afinion™ AS100 Analyzer.

- Anytime an unexpected test result is obtained.

- After software upgrade of the Afinion™ AS100 Analyzer.

The controls should always be analyzed if an unexpected test result is obtained (see the Afinion™ test

Package Insert, section Test result reporting). If local, state and/or federal regulations require more frequent

testing of control materials, then quality control should be performed in compliance with these regulations.

Each laboratory site can benefit from establishing a quality control plan.The laboratory director should

determine whether additional testing is appropriate for their laboratory.

09-38555



Yüklə 230,92 Kb.

Dostları ilə paylaş:
1   2   3   4   5   6   7   8




Verilənlər bazası müəlliflik hüququ ilə müdafiə olunur ©genderi.org 2024
rəhbərliyinə müraciət

    Ana səhifə