14  

CONFIGURATION

Operator configuration

The Operator ID function is disabled as a default setting by the manufacturer. 

Touch 

 in the configuration menu to enter the operator configuration menu.



Operator ID enable/disable

Touch 


 in operator configuration menu to enable/disable operator ID.

Select 


  to disable the operator ID function.

Select 


  to enable operator ID. Any operator ID is accepted.

Select 


 to enable operator ID with verification. 

     · To enable this function at least one supervisor is required to be present in the operator list. 

     · When operator ID verified is enabled, instrument configuration will only be available to the supervisors. 

     ·  To log in, the operator ID entered is required to be present in the operator list (see “Operator management”, 

bottom of the page).

Touch 


  to accept and return to the configuration menu.

Operator login expiration

Touch 

 in the operator configuration menu to set automatic logout of the operator. 



Enter the number of minutes before automatic logout of operator. 

The operator will automatically be logged out after the configured number of minutes after analysis of the test 

is complete.

Touch 


 to confirm and return to previous view.

Operator management

Touch 

 in operator configuration menu to enter operator list.



1

 

Touch 



 to add new operator. 

Touch desired operator ID and touch 

 to delete or

 

 to edit the highlighted operator.



2

Enter new/edit operator ID

Enter new/edit operator ID and touch 

 to enter. Both letters and numbers can be entered (maximum 

16 characters). 

If a barcode reader is connected to the Analyzer, a barcoded operator ID can be entered.



  15

CONFIGURATION__3'>CONFIGURATION

3

Configure the operator level:

Select whether this operator will be a user or supervisor.

1) User


2) Supervisor

Configure tests accessible by checking the appropriate test boxes for this operator.

Check off the test accessible for this operator.   

Touch 


 to return and edit the operator ID.

Touch 


 to accept and store new operator in the operator list. The operator list can store 500 operator 

IDs.


Supervisors will be marked with * in the operator list. When instrument is configured to Operator ID verified, 

configuration of instrument settings will only be available to the supervisors.

Copy operator list 

It is possible to copy operator lists between instruments using a USB flash drive. Insert USB flash in  

instrument USB port. Touch 

 to export operator list from instrument to USB flash. Move USB to new 

instrument and touch to 

 import operator list. Any existing operator list on the instrument will be 

deleted.

Choosing language

Touch 


  in the configuration menu to enter the language selection. The default setting by manufacturer is English.  

Other languages are available.

Touch the arrow in the window to view other options. Scroll down until you find the preferred language.  

Touch 


 to accept and return to the configuration menu.

Setting date and time

The correct date and time should always be set because the date and time for the analyses are stored and displayed in the patient 

and control records. The date format is YYYY:MM:DD, where YYYY is the year, MM is the month (01 to 12), and DD is the day (01 to 

31). The time format is hh:mm, where hh is the hour from 00 to 23 and mm is minutes from 00 to 59.

Touch 

 in the configuration menu to enter date/time setting.



1

Touch 


 to enter date setting. 

Touch 


 to enter time setting.

2

Enter today’s date or time. 

Touch 

 to confirm and return to the previous view.




16  

CONFIGURATION

Adjusting screen/beeper volume settings

Touch 


 in the configuration menu to enter the screen/beeper volume settings. 

The screen contrast can be adjusted.

Touch 

  to enter the screen contrast setting.



Touch 

  to enter the screen alignment setting.

Touch 

  to enter the beeper volume setting.



Screen contrast setting 

Adjust the screen contrast by touching 

 or 

  

Touch 



 to confirm and return to previous view.

Screen alignment setting 

Tap the cross-hair object (+) in the upper left corner using a blunt pencil to be precise. Repeat for the object 

appearing in the lower right corner and in the center of the screen. The previous screen view will automatical-

ly return.

Beeper volume setting

Adjust the beeper volume by touching 

 or 


  

Touch 


 to confirm and return to the previous view.


  17

CONFIGURATION

QC lockout configuration

The QC lockout function is disabled as a default setting by the manufacturer. 

Touch 

 in the configuration menu to enter the QC lockout setting menu.



Touch 

  to configure QC lockout for the assay selected.

Touch 

  to configure QC lockout interval.



Touch 

  to view/add/delete stored control lots in the control lot database.



1

Select assay type

Touch the arrow in the window to open the drop down menu to select the assay type. 

Touch the assay to select.



2

QC lockout

Select  

  to disable the QC lockout function. No QC runs will be required for this assay 

Select  

  to enable the QC lockout function. It is required to run 

ONE passed control,  

control level C I 

OR C II, to reset the QC lockout interval. 

Select  


 to enable the QC lockout function. It is required to run 

TWO passed control,  

both control level C I 

AND C II, to reset the QC lockout interval.

Touch 

 to confirm and return to the previous view.



3

QC lockout interval

Select 

 to configure QC lockout interval by number of runs.



Select 

 to configure QC lockout interval by hours.

Touch 

 to enter/edit number of runs/hours to QC lockout.



 

displays the number of runs/hours configured in the QC lockout interval.

Touch 

 to confirm and return to the previous view.



4

Control lot database

To add a control to the control lot database, the Alere Afinion™ Control Data is required.

The Alere Afinion™ Control Data is a numeric data string which contains all lot spesific data:

• Alere Afinion™ control lot number 

• Control type (assay) 

• Control level (C I or C II) 

• Control expiry date 

• Acceptable control range 

• CRC (check sum to validate the previous data)

The Alere Afinion™ Control Data and its accompanying barcode is found in the Alere Afinion™ Control kit  

package insert. If the Alere Afinion™ Control Data is not available, contact your local supplier. 

Touch 

 and either manually enter the Control Data or if a barcode reader is connected to the Analyzer 



(recommended), scan the barcode. The Alere Afinion™ Control Data may also be entered before, during or 

after a control run. The control lot will automatically be stored in the database (see page 25).

Touch 

 to delete selected lot number from the control lot database. When a control lot has reached its 



expiry date, the control will automatically be deleted from the instrument control database. The control lot 

database can store 100 control lots.