23
Using the control ID function
In quality control testing, a suitable control ID must always be entered. The lot number of the control material is recommended as
a suitable control ID. The control ID function cannot be disabled.
Entering Control ID
It is recommended to enter the control ID during processing of the Test Cartridge in the Analyzer. Entering the control ID will not
interrupt the processing. It is also possible to enter the control ID before and after processing. Both letters and numbers can be en-
tered (maximum 16 characters). The control ID will be displayed in the result records and appear on print outs and data transferred
to data management systems.
To enter the control ID during processing, do the following:
1
Touch
to enter the control ID option.
2
Enter control ID by numbers and/or touch
to enter letters.
Touch
to confirm and return to the previous view.
3
The entered control ID will appear on the screen.
The control ID touch button will remain in the view and make corrections possible.
TESTING PROCEDURES
24
Using the QC lockout function
When the QC lockout function is enabled for one or more assays, approved control testing is required within the configured interval.
If the interval expires, patient testing for the assay will be locked. A passed control run must be performed according to configuration,
to reset the interval or to unlock the assay for patient testing. A failed control run will disable patient testing (see “Configuration of QC
lockout” page 17).
QC lockout status
The status of the enabled QC lockouts is presented with a QC lockout status button (padlock symbol) visible in the Start-up menu.
This gives the operator the status of QC lockout before he/she attempts to run any tests.
The padlock symbol will only be visible if QC lockout function is enabled for one or more assay types.
The padlock symbols used are:
1
Enabled-unlocked
All controls are within the configured interval. It is possible to run patient tests for all assays.
2
Warning-unlocked
All controls are within the configured interval. When one or more of the assays has 10 % or less of the
configured interval remaining the warning icon will be displayed. It is possible to run patient tests for all
assays.
3
Expired-locked
One or more controls have expired according to the configured interval. Patient testing on the expired
assay has been locked.
4
Touch the QC lockout status button (padlock symbol) in the
Start-up menu to enter the QC lockout status view.
Status
The information is displayed as a list.
Only the assays with QC lockout activated are displayed in this window.
Red text indicates expired assays and orange text indicates assays within warning period.
Control level
How to reset QC lock interval and/or unlock expired assays.
If no control level is specified, it is required to run
one passed control, control level C I or C II, to reset the QC
lockout interval and unlock the assay for patient testing.
E.g.
HbA1c: #0
If the control level is specified it is required to run
two passed controls, both control level C I and C II, to reset
the QC lockout interval and unlock the assay for patient testing.
E.g.
ACR C I: 00.00.00
ACR C II: 00.00.00
Remaining time/runs
Remaining time (dd:hh:mm) or number of runs for each assay with active QC lockout is shown. dd is the
number of days, hh is the number of hours, and mm is the number of minutes until the assay will be locked.
# is number of patient tests.
TESTING PROCEDURES
25
TESTING PROCEDURES
Running controls with enabled QC lockout function
When running controls with the QC lockout function enabled, the Alere Afinion™ Control Data is required to be entered or
previously stored in the instrument control lot database (see “QC lockout configuration” page 17).
1) The Alere Afinion™ Control Data is entered before, during or after the control run. If a barcode scanner is connected (recommended)
the Control Data barcode may be scanned. The control lot will automatically be stored in the instrument control database.
2) If the Alere Afinion™ Control Data is previously stored in the instrument control database, the operator will simply need to enter
the 8 digit control lot number before, during or after the control run.
If the instrument is configured to QC lockout and the control lot number is not found in the Alere Afinion™ control database or
the Alere Afinion™ Control Data entered is not valid, the instrument will present an option to retry the input or discard the control run
result. If discarded, the result will not be stored in the instrument result records.
Passed (within the acceptable control range)
The result of the control is checked against the acceptable ranges for the corresponding lot number.
If the result is within the limits, a pass symbol
is displayed on the screen and the QC lock interval is reset
according to the QC lockout configuration. The result is stored in the instrument and is sent to the data
management system if connected.
If QC lockout is configured to require two control levels (C I
and C II), both levels must pass to reset the
lockout interval. Only the interval for the control level used in the test is reset.
Failed (above or below the acceptable control range)
When a control result is not within the acceptable ranges specified for the control lot, a failed symbol
is shown on the screen. The result is stored in the instrument and is sent to the data management system
if connected. The QC lockout interval will not be reset.
The arrow symbol will specify whether the result is above or below the acceptable ranges (see
“Handling and testing controls”, page 18).