Methods for impurity profiling
31
Sample preparation: All glassware is silylated. Add 200 mg of the heroin sample to
4 ml of 1M sulphuric acid and 4 ml of a 3:2 mixture of diethyl ether and
dichloromethane, mix thoroughly. Centrifuge to separate phases and transfer 3 ml of
the organic phase into a clean reaction vial. Dry under dry nitrogen at 60° C.* Add
300 µl of the derivatization solution and 60 µl of the internal standard. Seal the reac-
tion vial tightly, mix thoroughly and heat at 70° C for 30 min.
If method retention times shift more than 0.2 min, the instrument is recalibrated and/or
the retention time of the internal standard is adjusted by adjusting the carrier gas flow.
Rationale for use: The method allows a total impurity profile of all alkaloidal trace
impurities. Several modifications of the above method are in use worldwide. This ver-
sion should provide the sensitivity and the specificity necessary for the analysis of
highly refined heroin samples, that is, highly refined hydrochloride samples from
South-East Asia.
Outcome: Indication of general source region (South-East Asia, South-West Asia,
Mexico, South America). Sample comparisons for discrimination and evaluation of
samples for case-to-case evidential purposes (linkage determinations). Provides addi-
tional information required to confirm links between samples, that is, the method
should be used in conjunction with a major component analysis.
*In this form, a tightly capped reaction vial can be stored for up to a week at -5° C prior to
further analysis.
4.
Residual solvent analysis
Method B5:
Dynamic headspace (thermal desorption) gas chromatography
(Conventional GC instrument)
Source: J. Cartier, O. Guéniat and M. D. Cole, “Headspace analysis of solvents in
cocaine and heroin samples”, Science and Justice, vol. 37, No. 3 (1997), pp. 175-181.
Operating conditions:
Detector:
FID, at 45 ml/min hydrogen and 450 ml/min air
Column:
DB-1 or equivalent , 60 m x 0.33 mm x 3.0 µm
Carrier gas:
Helium at 0.7 ml/min (column pressure 17 psig)
Injection
technique:
1 µl; split, 30:1
Temperatures:
Injector: 280° C
Detector: 280° C
Oven: 35° C for 14 min, to 100° C at 5° C/min, to 245° C
at 7° C/min, no final hold (if petrol/gasoline was detected, the
temperature of the column was held at 245° C for 10 min)
32
Methods for impurity profiling of heroin and cocaine
Internal standard: None
Sample preparation:* Weigh 250-300 mg of the powdered heroin sample into a 2-ml
GC derivatization vial. Add a 0.25-ml insert, containing approximately 100 mg of
activated carbon (charcoal strip). After sealing the vial, heat the contents for 60 min
at 80° C and allow to cool. Extract the activated carbon with 50 µl of carbon disul-
phide and inject 1 µl of this solution** into the GC.
Rationale for use: The method provides for the detection of 12 solvents in heroin sam-
ples, at detection limits between 2-15 parts per million (ppm) for 250-300 mg powder
samples. No special injector device is required; the method uses a conventional GC
instrument. Solvent analyses may be carried out on samples up to two years old.
Outcome: Provides additional information and an independent means of confirming
links from organic impurity and inorganic ion determinations.
Method B6:
Static headspace-GC/MS
(Special injector device required)
Source: Modified from D. R. Morello and R. P. Meyers, “Qualitative
and quantita-
tive determination of residual solvents in illicit cocaine HCl and heroin HCl”, Journal
of Forensic Sciences, vol. 40, No. 6 (1995), pp. 957-963.
Operating conditions:
Detector:
Mass spectrometer, 70eV EI, 20-220 amu, 0.5 sec scan cycle
Column:
DB-1, 60 m x 25 mm x 1.0 µm
Carrier gas:
Helium, constant velocity approx. 30 cm/s
Injector
technique:
Split 38:1, Tekmar 7000/7050 headspace autosampler
22 ml vials
Equilibration:
14 min at 85° C
2-ml loop; injection time: 0.3 min
Temperatures:
Loop: 175° C
Injector: 190° C
Injector transfer line: 175° C
MS transfer line: 190° C
Oven: 35° C to 150° C at 6° C/min to 180° C at 15° C/min,
hold for 10 min
*Preparation of headspace samples must be performed under clean laboratory conditions to
avoid solvent cross-contamination.
**The remainder of this sample can be reused for ICP-MS analysis.
Methods for impurity profiling
33
Internal standard stock (ISS):*
d
6
-acetone at 6 mg/ml, d
9
-2-chloro-2-methylpropane at
3.5 mg/ml, d
14
-n-hexane at 1.5 mg/ml, d
8
-isopropanol at 7.5 mg/ml and d
8
-toluene at
1.5 mg/ml in dimethylsulfoxide (DMSO).** All standards should be +99.5% pure.
22% sodium sulfate solution: 220 gm of +99% anhydrous sodium sulfate in one litre
of deionized water. Between use, store at room temperature.
Dilute internal standard solution (DISS): Add 100 µl of the ISS for each 100 ml of
22% Na
2
SO
4
solution. Prepare daily.
Standard stock solutions (SSS): Methanol, ethanol, acetone, isopropanol,
n-butanol and
isobutanol at 0.4 mg/ml; methyl ethyl ketone, ethyl acetate, chloroform, methylene
chloride, 1,1,2-dichloroethane, ethyl ether, methyl acetate, cyclopentane, isobutyl
acetate, n-butylethyl acetate and 1,1,1-trichloroethane at 0.2 mg/ml; cyclohexane, ben-
zene, toluene, n-hexane, o-xylene, p-xylene, mesitylene and mesityl oxide at 0.08
mg/ml. Two additional calibrations solutions are created by making two 1:5 serial
dilutions.*** All dilutions are made using a solution of DMSO:water (5:1). Label the
solutions SSSH, SSSM, SSSL respectively from the highest to the lowest concentra-
tion solution.
Calibration solutions: Add 5 ml of DISS to four 22-ml headspace vials, then add 50
µl of SSSH to the first vial, 50 µl of SSSM to the second vial and 50 µl of SSSL to
the third vial. The fourth vial is used as a run blank (standard zero point) as it con-
tains only the DISS. Immediately seal each vial tightly with a silicon/polytetrafluo-
roethylene (PTFE)-lined cap.
Sample preparation: Accurately weigh approximately 50 mg equivalent of heroin base
into a 22-ml headspace vial, add 5 ml of DISS. Immediately seal each vial tightly
with a silicon/PTFE-lined cap.
Rationale for use: The method provides facile detection to the 0.00001% level and
reproducible quantification to the low 0.0001% level relative to the heroin content
and, with minor modification, is utilized for both heroin and cocaine samples. The
method requires a special injector device. Also requires the acquisition of appropri-
ately deuterated standard materials; however, a few millilitres of the standard materi-
als will last for a considerable period of time.
Outcome: Allows access to solvents trapped within the crystal matrices of samples
without decomposition of solvents or other sample components. The method provides
nearly comprehensive detection of all sample solvents and, very importantly, allows
the accurate assessment of the relative importance of the detected solvents.
*The internal standard stock is stable for up to two months if stored in the dark at -10° C
between uses.
**Deuterated standards were from the Cambridge Isotope Laboratories, 50 Frontage Rd.,
Andover, Massachusetts, United States.
***The calibration standards solutions were found to be stable up to four weeks if maintained
in the dark, at -10° C, between uses.