International


© ISO/IEC 2017 – All rights reserved  5



Yüklə 0,67 Mb.
Pdf görüntüsü
səhifə22/68
tarix31.12.2021
ölçüsü0,67 Mb.
#82101
1   ...   18   19   20   21   22   23   24   25   ...   68
ISO-IEC 17025-2017

 

© ISO/IEC 2017 – All rights reserved 



5


 

ISO/IEC 17025:2017(E)

6.3  Facilities and environmental conditions

6.3.1 

The  facilities  and  environmental  conditions  shall  be  suitable  for  the  laboratory  activities  and 

shall not adversely affect the validity of results.

NOTE 


Influences that can adversely affect the validity of results can include, but are not limited to, microbial 

contamination, dust, electromagnetic disturbances, radiation, humidity, electrical supply, temperature, sound 

and vibration.

6.3.2 

The requirements for facilities and environmental conditions necessary for the performance of 

the laboratory activities shall be documented.

6.3.3 

The laboratory shall monitor, control and record environmental conditions in accordance with 

relevant specifications, methods or procedures or where they influence the validity of the results.

6.3.4 

Measures to control facilities shall be implemented, monitored and periodically reviewed and 

shall include, but not be limited to:

a)  access to and use of areas affecting laboratory activities;

b)  prevention of contamination, interference or adverse influences on laboratory activities;

c)  effective separation between areas with incompatible laboratory activities.



6.3.5 

When the laboratory performs laboratory activities at sites or facilities outside its permanent 

control, it shall ensure that the requirements related to facilities and environmental conditions of this 

document are met.




Yüklə 0,67 Mb.

Dostları ilə paylaş:
1   ...   18   19   20   21   22   23   24   25   ...   68




Verilənlər bazası müəlliflik hüququ ilə müdafiə olunur ©genderi.org 2024
rəhbərliyinə müraciət

    Ana səhifə