Incremental Techniques 

|

 



Section 6 

9.  Rinse the electrode (ATC probe, stirrer probe and reference electrode, if applicable) with 

distilled water, blot dry with a lint-free tissue and place the electrode(s) into the sample. 

When the electrode and sample are ready, press the f2 (next) key.  

10. Wait for the reading to stabilize. When the reading is stable, press the f2 (yes) key. 

11. Use the numeric keypad and decimal key to enter the volume of the standard, in mL, to be 

added to the sample and press the f2 (accept) key. 

12. Add the standard to the sample and press the f2 (yes) key. 

13. Wait for the reading to stabilize. When the reading is stable, press the f2 (yes) key. 

14. The meter will display the result of the single known addition. Press the f3 (log/print) key to 

save and end the single known addition, export the data to the data log and export the data 

to a printer or computer, if a printer or computer is connected to the meter and enabled in 

the setup menu. 

Note: Press the f1 (escape) key to return to the measurement mode without saving the single 

known addition.  

15. Press the f2 (next) key.  

16. To perform a single known addition using the values that were entered for the previous 

single known addition, press the f2 (yes) key. To perform a single known addition and enter 

new values, press the f3 (no) key. To end the incremental technique and return to the 

measurement mode, press the f1 (escape) key. 

 

 



 

Thermo Scientific 

Orion Dual Star pH/ISE Benchtop Meter User Manual 

|

 



53 


Section 6 

|

 



Incremental Techniques 

Double Known Addition 

  For most double known addition analysis, it is recommended that:  



o  The sample volume should be 100 mL.  

o  The concentration of the standard added should be 100 times the expected sample 

concentration.  

o  The volume of the first addition should be 1 mL and the volume of the second addition 

should be 10 mL.  

  Refer to the appropriate electrode user guide for instructions for preparing the electrode, 



required solutions (ionic strength adjusters, standards, etc) and any special requirements.  

  Ensure that the electrode, ATC probe and reference electrode (if applicable) are connected 



to the meter inputs that correspond with the selected channel. 

 

Double Known Addition Procedure  

1.  Prepare the sample as directed in the electrode user guide.  

2.  In the measurement mode, press the setup key. Press the ▲ / ▼ keys to highlight 

Incremental Techniques and press the f2 (select) key. 

3.  Dual channel display only: Press the ▲ / ▼ keys to select the channel to perform the 

double known addition on and press the f2 (accept) key. 

4.  Press the ▲ / ▼ keys to highlight Double Known Addition and press the f2 (select) key. 

5.  Use the numeric keypad and decimal key to enter the sample volume, in mL, and press the 

f2 (accept) key. 

6.  Use the numeric keypad and decimal key to enter the total volume, in mL, and press the 



f2 (accept) key. The total volume includes the sample volume, ISA volume and diluting 

agent volume, but not the standard volume. 

7.  Use the numeric keypad and decimal key to enter the concentration of the standard and 

press the f2 (accept) key.  

8.  Rinse the electrode (ATC probe, stirrer probe and reference electrode, if applicable) with 

distilled water, blot dry with a lint-free tissue and place the electrode(s) into the sample. 

When the electrode and sample are ready, press the f2 (next) key. 

9.  Wait for the reading to stabilize. When the reading is stable, press the f2 (yes) key. 

10. Use the numeric keypad and decimal key to enter the volume of the first increment of 

standard, in mL, to be added to the sample and press the f2 (accept) key. 

11. Add the first increment of standard to the sample and press the f2 (yes) key. 

12. Wait for the reading to stabilize. When the reading is stable, press the f2 (yes) key. 



54 

|

 



Orion Dual Star pH/ISE Benchtop Meter User Manual 

Thermo Scientific 




Incremental Techniques 

|

 



Section 6 

13. Use the numeric keypad and decimal key to enter the volume of the second increment of 

standard, in mL, to be added to the sample and press the f2 (accept) key. 

14. Add the second increment of standard to the sample and press the f2 (yes) key. 

15. Wait for the reading to stabilize. When the reading is stable, press the f2 (yes) key. 

16. The meter will display the result of the double known addition. Press the f3 (log/print) key to 

save and end the double known addition, export the data to the data log and export the data 

to a printer or computer, if a printer or computer is connected to the meter and enabled in 

the setup menu. 

Note: Press the f1 (escape) key to return to the measurement mode without saving the double 

known addition.  

17. Press the f2 (next) key.  

18. To perform a double known addition using the values that were entered for the previous 

double known addition, press the f2 (yes) key. To perform a double known addition and 

enter new values, press the f3 (no) key. To end the incremental technique and return to the 

measurement mode, press the f1 (escape) key. 

 

 



 

Thermo Scientific 

Orion Dual Star pH/ISE Benchtop Meter User Manual 

|

 