Specifications
Spectral Range 7,500 to 370 cm-1 , with standard KBr beamsplitter
10,000 to 400 cm-1 , optional, with broadband KBr beamsplitter
5,000 to 200 cm-1 , optional with CsI optics
6,000 to 500 cm-1 , optional with "High Humidity" ZnSe optics
Resolution Better than 1 cm-1 (apodized), optional better than 0.5 cm-1 (apodized)
Wavenumber Accuracy Better than 0.01 cm-1 @ 2,000 cm-1
Photometric Accuracy Better than 0.1% T
Signal-to-Noise (Minimum) 5 Sec: >6,000:1 (= <7.2*10-5AU noise) peak-to-peak, 4 cm-1 resolution
Signal-to-Noise (Achievable) 5 Sec: >8,000:1 (= <5.4*10-5AU noise) peak-to-peak, 4 cm-1 resolution1
Min: >45,000:1 (= <9.7*10-6AU noise) peak-to-peak, 4 cm-1 resolution
Interferometer RockSolidTM, Permanent aligned, high stability
Scan Speed 3 velocities, 2.2 - 20 kHz (1.4 - 12.7 mm/sec opd),
optional 4 velocities, 2.2 - 40 kHz (1.4 - 25.5 mm/sec opd)
Detector DigiTectTMdetector system, High sensitivity DLATGS,
optional LN2cooled MCT detector, easy exchange
A/D Converter 24 bit
Validation IVU internal validation unit
Spectrometer Size 66.5(W) x 43.4(D) x 28.1(H) cm
Sample Compartment Size 25.5(W) x 27(D) x 16(H) cm
Weight 37 Kg
Spectrometer Power 85 - 265 VAC, 45 - 67 Hz, 70 W
Computer Interface Ethernet connection
Spectroscopy Software OPUSTM: easy to use, full GLP compliant, full 21 CFR part 11 compliant
Options
Spectrometer Options
- Switch mirror for output port
- Beam Bender port multiplexer, for up to four different output ports
- Dual channel acquisition
MIR Accessories for the Sample Compartment
- Accessory kit including ATR, diffuse & specular reflectance
- Liquid cells, demountable and fixed
- Liquid autosampler
- Gas cells, variable and fixed pathlength
- Diffuse reflectance accessories
- Specular reflectance accessories
- Variable angle specular reflectance
- ATR, single bounce
- Diamond ATR
- Horizontal ATR, multiple bounce
- Photoacoustic cell
- Beam condensers
and more...
External MIR Accessories
- HYPERIONTMseries, FT-IR microscope
- HTS-XT, High Throughput Screening eXTension
- TGA-, and GC-FTIR coupling
- PMA 50, Polarization Modulation Accessory for VCD and PM-IRRAS
- MICOR-ID, Micro-organisms identification
- External sample compartment
Software
OPUS, OPtics User Software, consists of a suite of softwarepackages for all standard and specialized applications.Optional OPUS packages:
- OPUS/SEARCH, Software package for the creation of userlibraries and expanded library search capabilities
- OPUS/STRUCTURE, Software package for the drawingand search of chemical structures
- OPUS/IDENT, Software package for the identification ofraw materials
- OPUS/QUANT, Software package for the quantification ofingredients
- OPUS/CHROM, Software package for TG- and GC-FTIRcoupling
- OPUS/MAP, Software package for measurements with amapping stage
- OPUS/VIDEO, Software package for video supportedmeasurements
- OPUS/3D, Software package for the evaluation of 3D-data
- OPUS/LAB, Software package for QA/QC
Libraries
- BRUKER/MERCK (3125 spectra)
- BRUKER/BASF (3900 spectra)
- BRUKER POLYMER (1300 spectra)
- BRUKER/PHARM (400 spectra)
- BRUKER/CROP (185 spectra)
- BRUKER/KIDNEY STONES (>5000 spectra)
- EPA Vapor Phase (3300 spectra)
- Georgia Crime Lab (740 spectra)
- Extended Sadtler libraries
Services
- Remote Diagnostics
- Services Contracts
- Validation Packages and Services
Heatable Variable Pathlength Gas Cell Specifications
Model Sp-M-4-10-H metal-bodied variable cell in heatable configuration. Cell of
approx 2.5 liters in volume, with base path of 10 inches and variable from 1 to
10 meters. Nickel-coated aluminum endplates and internal hardware. Glass
internal mirror blanks with Protected-gold optical coatings. KCl windows
included, SS Swagelock valves & fittings included. (includes 30/30 compound
gauge) Open-air transfer optics for FTIR Spectrometer . Viton o-ring seals on
windows and endplates. Laser device for cell alignment and pathlength
verification is included. A heating jacket and a digital temperature controller are
included.
Dostları ilə paylaş: |