Gaia Data Release 1 Documentation release 0



Yüklə 5,01 Kb.
Pdf görüntüsü
səhifə90/125
tarix02.01.2018
ölçüsü5,01 Kb.
#19053
1   ...   86   87   88   89   90   91   92   93   ...   125

Figure 5.11: The left panel shows the selection done to produce the fitting shown in the right panel for SDSS stripe
82 stars processed in PhotPipe cycle 1.
Figure 5.12: Johnson-Cousins relationships using Landolt standard stars observed with Gaia.
237


Figure 5.13: Fitted relationship for stars common in M4 Hubble and Gaia DR1 data.
5.4.1
Pre-processing
Author(s): Francesca De Angeli
The ingestion and pre-processing activities include a number of steps that are necessary for any further processing
of the data.
The ingestion phase in particular converts the data as received by the DPC into a data format that is designed and
optimized for the photometric processing. This requires matching information contained in di
fferent tables using
the available cross-match information. Trivial unit conversions are also applied in this stage.
Bias correction and computation of predicted positions are important modules in the pre-processing stage.
Bias correction for the SM and AF observations is taken care within the IDT pre-processing (see Section 2.3.5).
Only BP and RP observations need to be corrected for bias in PhotPipe. This is done using the same algorithm and
software as the one used in IDT.
Predicted positions of sources at the time of observation are required for the calibration of flux loss and of the
geometry of the BP
/RP instruments. The computation of predicted positions at the desired accuracy requires the
availability of astrophysical coordinates for all sources at a higher accuracy than what is normally available from
ground. These were not available yet for this cyclic processing (at the next cycle of operations PhotPipe will start
making use of Gaia astrometric results, the accuracy of which will improve at each cycle). For this reason, for
Cycle 01 processing no flux loss calibration could be performed. In order to allow the calibration of the geometry
of the BP and RP instruments, centroid coordinates were extrapolated from the AF observations onto the BP and
RP CCDs to provide a prediction of the location of the source in the window reference system (WRS) at the
observation time.
238


5.4.2
BP
/RP processing
Author(s): Giorgia Busso, Francesca De Angeli
The BP
/RP processing includes the background calibration and removal, the geometric calibration and its applica-
tion, the di
fferential dispersion function calibration and the flux and LSF calibration. The differential dispersion
function calibration was run for Cycle 01 for validation but results were not applied: only nominal dispersion
functions were used convert sample positions into absolute wavelengths. The flux and LSF calibration did not run
for Cycle 01. Spectra will be part of future releases as planned within DPAC.
The calibration models have been already defined in section Section 5.3. In this section, more detail will be given
on the processing aspects of these calibrations.
5.4.2.1
Background removal
In the current release only the most dominant contribution to the background is calibrated and then subtracted, that
is the one from the stray light. For this, the models described in 5.3.1 are used. If in the maps there are empty
bins, interpolation between the neighbours is performed to fill them. The completed map is then used to build
a bicubic spline interpolator which, given the heliotropic spin phase and the AC coordinate of the transit to be
corrected, computes the background value in electron
/pixel/second. This value is converted to the appropriate one,
depending on the window sampling (1D or 2D), and then subtracted to the transit to be corrected.
As the calibration for the charge release is not yet available, only the transits with a distance from the charge
injection bigger than 50 TDI, are processed at this stage.
5.4.2.2
Geometric calibration
The di
fferential dispersion function and geometric calibrations are carried out simultaneously. The differential
AL geometric and dispersion calibration is determined with respect to an initial fixed arbitrary reference sample
position (ARS) and using observations of a sub-sample of the sources selected in a narrow range of colour to ensure
that they are of similar spectral types (and their spectra are therefore quite similar).
The calibration proceeds through the following steps:
• Alignment
– Initial estimate of the set of AL position shifts and flux scaling factors to be applied to the
observed spectra in a given calibration unit to align them with each other. All the suitable
observed spectra are first normalized to a given magnitude (set in the configuration). Within
each calibration unit then, all spectra are cross-correlated using a spectrum that is observed at
AC position close to the CCD centre (and therefore is representative of an average dispersion)
as reference. The result of this process is a set of AL position shift and scaling factor, one per
input spectrum.
– Refinement of the set of shifts and scaling factors found in the previous step. This is achieved
by fitting a spline to all the observed spectra (after applying the adjustments found in the previ-
ous step) in one calibration unit. There will be one spline (reference spectrum) per calibration
unit. A correction to the shift and scaling factor first estimates is computed by fitting each
239


of the observed spectra back to the corresponding reference spectrum spline. This is an itera-
tive process where at each iteration a new reference spectrum is computed. The result of this
process is a new set of shifts and scaling factors.
• Di
fferential dispersion calibration In this step the algorithm runs on a set of binned spectra (one
per calibration unit) defined using the adjusted spectra obtained from the previous step. Each binned
spectrum is computed on a fixed grid, the value of the binned spectrum at each bin is computed as
the weighted mean of all observed samples (from all available transits) falling within the bin range.
The binned spectra are thus compared to compute a set of additional AL position shifts (one per
calibration unit) and a set of corrections for the nominal di
fferential dispersion functions (again one
per calibration unit). This is achieved in an iterative process where each iteration consists of two
steps:
– a spline is fitted to all the input binned spectra by iteratively adjusting the set of AL position
shifts and scaling factors (one per calibration unit);
– an LSQ fit is done with all the input binned spectra and the fitted spline to obtain as a result a
set of AL position shifts and corrections for the nominal di
fferential dispersion functions (one
per calibration unit).
• Di
fferential geometric calibration The differential AL geometric calibration is finally derived as
the median of the di
fference between the location of an arbitrary reference sample (ARS) in the
observed spectrum (given by the location of the ARS in the reference system adjusted by all the
shifts computed so far, the AL position shift per epoch spectrum computed in the alignment and
the additional shift per calibration unit computed in the di
fferential dispersion calibration) and the
predicted AL data space coordinate for each transit.
It is probably necessary to spend a few more words on the definition of the arbitrary reference sample
(ARS). This is the AL location of an arbitrary e
ffective wavelength. For the internal calibration it
is irrelevant to know what this wavelength is. The di
fferential AL geometric calibration is defined
with respect to the location of the ARS. The application of the geometric calibration determines the
observed location of the ARS expressed in pixels within the window.
In future cycles, the location of the ARS will be optimized to make sure that the di
fference between
e
ffective wavelength and nominal wavelength is at its minimum in the vicinity of the ARS for dif-
ferent spectral types. Each sample position corresponds to a given nominal wavelength through the
dispersion function. However in general this wavelength is di
fferent from the effective wavelength
(which is the average wavelength weighted by the spectrum shape and therefore clearly depends on
the spectral type). In order to make sure that the chosen ARS is consistent over a large range of
spectral types, we need to select a sample position where the di
fference between the nominal and
e
ffective wavelength has a minimum.
For Cycle 01, no optimization was done and the nominal values for reference sample and corre-
sponding wavelength were adopted.
The application is done on all observed spectra. It consists of two steps:
• Di
fferential geometric calibration application: defines for each observed spectrum the AL loca-
tion of the reference e
ffective wavelength. This is referred to as the observed ARS.
• Di
fferential dispersion calibration application: converts the observed sample AL positions into
the pseudo-wavelength system. As mentioned only nominal dispersion functions were used for
Cycle 01 processing (and for the generation of the data included in Gaia DR1).
240


Yüklə 5,01 Kb.

Dostları ilə paylaş:
1   ...   86   87   88   89   90   91   92   93   ...   125




Verilənlər bazası müəlliflik hüququ ilə müdafiə olunur ©genderi.org 2024
rəhbərliyinə müraciət

    Ana səhifə