Orion versa star user Manual


   |    pH, pH/ISE and pH/LogR Measurement Modules



Yüklə 6,2 Kb.
Pdf görüntüsü
səhifə19/44
tarix05.01.2018
ölçüsü6,2 Kb.
#19795
1   ...   15   16   17   18   19   20   21   22   ...   44

Chapter 4  
|
  
pH, pH/ISE and pH/LogR Measurement Modules 
17.
 
The meter will display the calculated sample concentration and summary of the incremental 
technique performed. Press the f1 (Esc) key to proceed to the measurement mode or press 
the f2 (Next) key to perform another analysis. 
Double Known Addition 
Double Known Addition Recommendations 
The sample volume should be 100 mL. 
The concentration of the standard should be 100 times the expected sample concentration. 
The volume of the first addition should be 1 mL and the volume of the second addition should be 10 mL. 
1.
 
Prepare the ion selective electrode (ISE) according to the instructions in the electrode user 
manual. Prepare all required solutions (ionic strength adjusters, standards, etc) and review 
any special requirements outlined in the electrode user manual. Connect the ISE, and any 
other electrodes to be used, to the module inputs on the selected channel.  
2.
 
In the measurement modepress the channel key until the channel with the pH/ISE module 
with ISE connected is shown. If needed, press the mode key until ISE Measurement is 
shown as the measurement mode and use the
 ISE Mode Menu 
to set the Electrode Type.  
3.
 
In the measurement mode, press the setup key.  
4.
 
Press the ◄ or ► key to highlight the channel setup menu that the ISE is connected with 
(Channel 1, Channel 2, Channel 3 or Channel 4) and press the f3 (Select) key.  
5.
 
Press the ◄ or ►key to highlight Incremental Techniques and press the f3 (Select) key.  
6.
 
Press the  or  key to highlight Double Known Addition and press the f3 (Select) key.  
7.
 
Enter the sample volume using the numeric keypad and press the f2 (Accept) key. 
8.
 
Enter the total volume (sample volume plus added ISA volume) using the numeric keypad 
and press the f2 (Accept) key. 
9.
 
Enter the standard concentration using the numeric keypad and press the f2 (Accept) key. 
10.
 
Rinse the ISE and any other electrodes in use with distilled water or rinse solution required 
for that electrode. Blot the electrode dry with a lint-free tissue or shake the electrode, place 
into the prepared sample and press the f2 (Next) key.  
a.
 
Follow all electrode rinsing and drying instructions or exceptions in the ISE user 
manual. Do not wipe or rub the sensing element of the electrode. 
11.
 
Wait for the mV value on the meter to stabilize and press the f2 (Accept) key. 
12.
 
Enter the first standard volume using the numeric keypad and press the f2 (Accept) key. 
13.
 
Add the first specified volume of the standard to the sample and press the f2 (Yes) key. 
14.
 
Wait for the mV value on the meter to stabilize and press the f2 (Accept) key. 
15.
 
Enter the second standard volume using the numeric keypad and press the f2 (Accept) key. 
16.
 
Add the second specified volume of the standard to the sample and press the f2 (Yes) key. 
17.
 
Wait for the mV value on the meter to stabilize and press the f2 (Accept) key. 
50  
|
  
Orion V
ERSA 
S
TAR
 Meter User Manual 
Thermo Scientific 


Chapter 4  
|
  
pH, pH/ISE and pH/LogR Measurement Modules 
18.
 
The meter will display the calculated sample concentration and summary of the incremental 
technique performed. Press the f1 (Esc) key to proceed to the measurement mode or press 
the f2 (Next) key to perform another analysis. 
Single Known Subtraction 
In known subtraction, each addition of standard subtracts an increment of the species to be 
measured. Enter the reaction ratio of the measured species to the added standard. For 
example, if a sulfide (S
2-
) sample will be subtracted by the addition of silver (Ag
+
), two silver 
ions are required to subtract each sulfide ion and the reaction ratio is 0.5. 
Sample concentration in Molarity or Normality should be known within an order of magnitude 
and the total sample concentration should be approximately halved upon addition of the 
standard. Concentrations are in Molarity or Normality due to the stoichiometric nature of the 
subtraction. Refer to the following table to choose the correct standard concentration and 
volume of addition.  
Amount of standard to be added per 
100 mL of sample 
Standard concentration (in Molarity or Normality) 
compared to expected sample concentration 
1 mL 
50 times more concentrated 
5 mL 
10 times more concentrated 
10 mL 
5 times more concentrated 
1.
 
Prepare the ion selective electrode (ISE) according to the instructions in the electrode user 
manual. Prepare all required solutions (ionic strength adjusters, standards, etc) and review 
any special requirements outlined in the electrode user manual. Connect the ISE, and any 
other electrodes to be used, to the module inputs on the selected channel.  
2.
 
Determine the ISE slope as directed in the electrode user manual. 
3.
 
In the measurement mode, press the channel key until the channel with the pH/ISE module 
with ISE connected is shown. If needed, press the mode key until ISE Measurement is 
shown as the measurement mode and use the
 ISE Mode Menu 
to set the Electrode Type.  
4.
 
In the measurement mode, press the setup key.  
5.
 
Press the ◄ or ► key to highlight the channel setup menu that the ISE is connected with 
(Channel 1, Channel 2, Channel 3 or Channel 4) and press the f3 (Select) key.  
6.
 
Press the ◄ or ►key to highlight Incremental Techniques and press the f3 (Select) key.  
7.
 
Press the  or  key to highlight Single Known Subtraction and press the f3 (Select) key.  
8.
 
Press the f3 (Clear) key as needed to delete the displayed slope value. Use the numeric 
keypad to enter the ISE slope value determined in step 2 and press the f2 (Accept) key.  
9.
 
Enter the sample volume using the numeric keypad and press the f2 (Accept) key. 
10.
 
Enter the total volume (sample volume plus added ISA volume) using the numeric keypad 
and press the f2 (Accept) key. 
11.
 
Enter the standard concentration in Molarity or Normality using the numeric keypad and 
press the f2 (Accept) key. 
Thermo Scientific 
Orion V
ERSA 
S
TAR
 Meter User Manual  
|
  
51
 


Yüklə 6,2 Kb.

Dostları ilə paylaş:
1   ...   15   16   17   18   19   20   21   22   ...   44




Verilənlər bazası müəlliflik hüququ ilə müdafiə olunur ©genderi.org 2024
rəhbərliyinə müraciət

    Ana səhifə