Conductivity Measurement Module
Connecting Electrodes to the Module
To begin using the V
ERSA
S
TAR
meter and module system, prepare the meter and insert the
module into an available channel on the back of the meter. See
Chapter 2 Meter Basics
for
instructions. For information on customizing meter settings, refer to
Chapter 3 Setup Menu
.
Ø
It is recommended that electrodes be placed in the meter-attached electrode stand for easy
movement in and out of beakers during calibration, sample measurement and storage.
VSTAR-CND V
ERSA
S
TAR
Conductivity Module
1.
Connect a conductivity probe to the 8 pin MiniDIN connector on the module. If the
conductivity probe has a built-in temperature probe, the temperature will also be measured
when the conductivity probe is connected.
2.
Prepare the conductivity probe for use as instructed in the probe user manual.
C H A P T E R 5
8 pin MiniDIN
56
|
Orion V
ERSA
S
TAR
Meter User Manual
Thermo Scientific
Chapter 5
|
Conductivity Measurement Module
Channel 1-4 Setup Menus
Mode Menus for Conductivity, TDS, Salinity and Resistivity
Within the Channel 1, Channel 2, Channel 3 and Channel 4 setup menus are Method, Mode
and Temperature submenus, which can be used to customize measurement settings and
parameters for the selected channel. This section provides detailed information on the Mode
menus for each measurement parameter available when using a conductivity module.
1.
In the measurement mode, press the setup key.
2.
Press the ◄ or ► key to highlight Channel 1, Channel 2, Channel 3 or Channel 4 (channel
number the module is connected with) and press the f3 (Select) key.
3.
Press the ◄ or ► key to highlight Mode and press the f3 (Select) key.
4.
Press the ▲ or ▼ key to highlight Conductivity, TDS, Salinity or Resistivity and press the
f3 (Select) key.
5.
Press the ▲ or ▼ key to highlight a menu option and press the ◄ or ► key to highlight a
submenu option. Onscreen text prompts provide instructions for changing the parameters.
a.
Use the popup keypad screen
for
Electrode SI No and
Sample ID. Press the
▲, ▼, ◄ or ► key to highlight
an alphanumeric character (numeric only for
Auto
Increment Sample ID), p
ress the f2 (Enter) key to input
the character and repeat until the desired label is
entered. Press the f1 (Done) key to save the entry.
Thermo Scientific
Orion V
ERSA
S
TAR
Meter User Manual
|
57
Chapter 5
|
Conductivity Measurement Module
Conductivity Mode Menu
Main Menu
Settings
Details
Read Type
Auto-Read, Timed,
Single-Shot, Continuous
For Timed and Single-Shot, set the time in hours,
minutes, seconds using the arrow keys
Cell K
0.475
Enter the nominal cell constant (K) value of the
conductivity probe, value is used when
performing an automatic conductivity calibration
Ref. Temperature
5°C, 10°C, 15°C, 20°C, 25°C
Set the reference temperature that all displayed
conductivity measurements will be reported to
using the selected temperature compensation
Temp.
Compensation
Linear, nLFn, nLFu, EP, Off
Set the temperature compensation type used to
calculate and report conductivity measurements
at the selected reference temperature
Temperature Coeff. 2.10
Enter the temperature coefficient value for linear
temperature compensation
Stability
Smart, Fast, Medium, Slow
Set when a measurement is recognized as stable
Averaging
Off, Automatic Smart
Set averaging for faster measurement stability
Alarm
Limit Alarm
Off, On
Set high and low limit values to activate alarm
Cal Due Alarm Off, On
Set a calibration interval (hours) to activate alarm
Set Point Alarm Off, On
Set a base and offset value to activate alarm
Cell Type
Standard, USP
Select Standard for most conductivity probes or
USP for ultra pure water 2-cell conductivity probe
(Catalog Number 013016MD)
Electrode SI No
----
Use the alphanumeric keypad to enter the serial
number or clear all to shown no value (----)
Sample ID
Off, Manual, Auto Increment
Use the popup keypad screen to enter an
alphanumeric manual ID or a numeric value as
the starting value for auto incremental ID
Detailed information on these features is available in the
Appendix
section of this user manual.
58
|
Orion V
ERSA
S
TAR
Meter User Manual
Thermo Scientific
Chapter 5
|
Conductivity Measurement Module
TDS Mode Menu
Main Menu
Settings
Details
Read Type
Auto-Read, Timed,
Single-Shot, Continuous
For Timed and Single-Shot, set the time in hours,
minutes, seconds using the arrow keys
Cell K
0.475
Enter the nominal cell constant (K) value of the
conductivity probe; value is used when
performing an automatic conductivity calibration
TDS Factor
Linear, ISO/EN 27888
0.49
Set the TDS factor type used to convert
conductivity to TDS measurements; enter a 0.02
to 9.99 value for the Linear TDS factor type
Ref. Temperature
5°C, 10°C, 15°C, 20°C, 25°C
Set the reference temperature that all displayed
TDS measurements will be reported to using the
selected temperature compensation
Temp.
Compensation
Linear, nLFn, nLFu, EP, Off
Set the temperature compensation type used to
calculate and report TDS measurements at the
selected reference temperature
Temperature Coeff. 2.10
Enter the temperature coefficient value for linear
temperature compensation
Stability
Smart, Fast, Medium, Slow
Set when a measurement is recognized as stable
Averaging
Off, Automatic Smart
Set averaging for faster measurement stability
Alarm
Limit Alarm
Off, On
Set high and low limit values to activate alarm
Set Point Alarm Off, On
Set a base and offset value to activate alarm
Cell Type
Standard, USP
Select Standard for most conductivity probes or
USP for ultra pure water 2-cell conductivity probe
Salinity Mode Menu
Main Menu
Settings
Details
Read Type
Auto-Read, Timed,
Single-Shot, Continuous
For Timed and Single-Shot, set the time in hours,
minutes, seconds using the arrow keys
Cell K
0.475
Enter the nominal cell constant (K) value of the
conductivity probe; value is used when
performing an automatic conductivity calibration
Salinity Type
Practical Salinity, Sea Water Set the salinity type used to convert conductivity
to salinity measurements
Stability
Smart, Fast, Medium, Slow
Set when a measurement is recognized as stable
Averaging
Off, Automatic Smart
Set averaging for faster measurement stability
Alarm
Limit Alarm
Off, On
Set high and low limit values to activate alarm
Set Point Alarm Off, On
Set a base and offset value to activate alarm
Cell Type
Standard, USP
Select Standard for most conductivity probes or
USP for ultra pure water 2-cell conductivity probe
Detailed information on these features is available in the
Appendix
section of this user manual.
Thermo Scientific
Orion V
ERSA
S
TAR
Meter User Manual
|
59
Dostları ilə paylaş: |