Conductivity Measurement Module  
Connecting Electrodes to the Module 
To begin using the V
ERSA 
S
TAR
 meter and module system, prepare the meter and insert the 
module into an available channel on the back of the meter. See 
Chapter 2 Meter Basics
 
for 
instructions. For information on customizing meter settings, refer to 
Chapter 3 Setup Menu

Ø
 
It is recommended that electrodes be placed in the meter-attached electrode stand for easy 
movement in and out of beakers during calibration, sample measurement and storage. 
VSTAR-CND V
ERSA 
S
TAR
 Conductivity Module 
 
1.
 
Connect a conductivity probe to the 8 pin MiniDIN connector on the module. If the 
conductivity probe has a built-in temperature probe, the temperature will also be measured 
when the conductivity probe is connected. 
2.
 
Prepare the conductivity probe for use as instructed in the probe user manual.  
 
 
C H A P T E R   5  
8 pin MiniDIN 
56  
|
  
Orion V
ERSA 
S
TAR
 Meter User Manual 
Thermo Scientific 


Chapter 5  
|
  
Conductivity Measurement Module 
Channel 1-4 Setup Menus 
Mode Menus for Conductivity, TDS, Salinity and Resistivity 
Within the Channel 1, Channel 2, Channel 3 and Channel 4 setup menus are Method, Mode 
and Temperature submenus, which can be used to customize measurement settings and 
parameters for the selected channel. This section provides detailed information on the Mode 
menus for each measurement parameter available when using a conductivity module.  
1.
 
In the measurement mode, press the setup key. 
2.
 
Press the  or  key to highlight Channel 1, Channel 2, Channel 3 or Channel 4 (channel 
number the module is connected with) and press the f3 (Select) key. 
 
3.
 
Press the  or  key to highlight Mode and press the f3 (Select) key. 
 
4.
 
Press the  or  key to highlight Conductivity, TDS, Salinity or Resistivity and press the 
f3 (Select) key.  
 
5.
 
Press the  or  key to highlight a menu option and press the  or  key to highlight a 
submenu option. Onscreen text prompts provide instructions for changing the parameters.       
a.
 
Use the popup keypad screen
 for 
Electrode SI No and 
Sample ID. Press the 
▲, ▼, ◄ or ► key to highlight 
an alphanumeric character (numeric only for 
Auto 
Increment Sample ID), p
ress the f2 (Enter) key to input 
the character and repeat until the desired label is 
entered. Press the f1 (Done) key to save the entry.  
Thermo Scientific 
Orion V
ERSA 
S
TAR
 Meter User Manual  
|
  
57
 


Chapter 5  
|
  
Conductivity Measurement Module 
 
Conductivity Mode Menu 
  
 
Main Menu 
Settings 
Details 
Read Type 
Auto-Read, Timed,  
Single-Shot, Continuous 
For Timed and Single-Shot, set the time in hours, 
minutes, seconds using the arrow keys 
Cell K 
0.475 
Enter the nominal cell constant (K) value of the 
conductivity probe, value is used when 
performing an automatic conductivity calibration 
Ref. Temperature 
5°C, 10°C, 15°C, 20°C, 25°C 
Set the reference temperature that all displayed 
conductivity measurements will be reported to 
using the selected temperature compensation 
Temp. 
Compensation 
Linear, nLFn, nLFu, EP, Off 
Set the temperature compensation type used to 
calculate and report conductivity measurements 
at the selected reference temperature 
Temperature Coeff.  2.10 
Enter the temperature coefficient value for linear 
temperature compensation 
Stability 
Smart, Fast, Medium, Slow 
Set when a measurement is recognized as stable 
Averaging 
Off,  Automatic Smart 
Set averaging for faster measurement stability 
Alarm 
Limit Alarm 
Off,  On 
Set high and low limit values to activate alarm  
Cal Due Alarm  Off,  On 
Set a calibration interval (hours) to activate alarm 
Set Point Alarm  Off,  On 
Set a base and offset value to activate alarm 
Cell Type 
Standard,  USP 
Select Standard for most conductivity probes or 
USP for ultra pure water 2-cell conductivity probe 
(Catalog Number 013016MD) 
Electrode SI No 
---- 
Use the alphanumeric keypad to enter the serial 
number or clear all to shown no value (----) 
Sample ID 
Off, Manual, Auto Increment 
Use the popup keypad screen to enter an 
alphanumeric manual ID or a numeric value as 
the starting value for auto incremental ID 
 
Detailed information on these features is available in the 
Appendix 
section of this user manual. 
 
 
58  
|
  
Orion V
ERSA 
S
TAR
 Meter User Manual 
Thermo Scientific 


Chapter 5  
|
  
Conductivity Measurement Module 
TDS Mode Menu 
Main Menu 
Settings 
Details 
Read Type 
Auto-Read, Timed,  
Single-Shot, Continuous 
For Timed and Single-Shot, set the time in hours, 
minutes, seconds using the arrow keys 
Cell K 
0.475 
Enter the nominal cell constant (K) value of the 
conductivity probe; value is used when 
performing an automatic conductivity calibration 
TDS Factor 
Linear,      ISO/EN 27888 
 0.49 
Set the TDS factor type used to convert 
conductivity to TDS measurements; enter a 0.02 
to 9.99 value for the Linear TDS factor type  
Ref. Temperature 
5°C, 10°C, 15°C, 20°C, 25°C 
Set the reference temperature that all displayed 
TDS measurements will be reported to using the 
selected temperature compensation 
Temp. 
Compensation 
Linear, nLFn, nLFu, EP, Off 
Set the temperature compensation type used to 
calculate and report TDS measurements at the 
selected reference temperature 
Temperature Coeff.  2.10 
Enter the temperature coefficient value for linear 
temperature compensation 
Stability 
Smart, Fast, Medium, Slow 
Set when a measurement is recognized as stable 
Averaging 
Off,  Automatic Smart 
Set averaging for faster measurement stability 
Alarm 
Limit Alarm 
Off,  On 
Set high and low limit values to activate alarm  
Set Point Alarm  Off,  On 
Set a base and offset value to activate alarm 
Cell Type 
Standard,  USP 
Select Standard for most conductivity probes or 
USP for ultra pure water 2-cell conductivity probe 
 
Salinity Mode Menu 
Main Menu 
Settings 
Details 
Read Type 
Auto-Read, Timed,  
Single-Shot, Continuous 
For Timed and Single-Shot, set the time in hours, 
minutes, seconds using the arrow keys 
Cell K 
0.475 
Enter the nominal cell constant (K) value of the 
conductivity probe; value is used when 
performing an automatic conductivity calibration 
Salinity Type 
Practical Salinity, Sea Water  Set the salinity type used to convert conductivity 
to salinity measurements  
Stability 
Smart, Fast, Medium, Slow 
Set when a measurement is recognized as stable 
Averaging 
Off,  Automatic Smart 
Set averaging for faster measurement stability 
Alarm 
Limit Alarm 
Off,  On 
Set high and low limit values to activate alarm  
Set Point Alarm  Off,  On 
Set a base and offset value to activate alarm 
Cell Type 
Standard,  USP 
Select Standard for most conductivity probes or 
USP for ultra pure water 2-cell conductivity probe  
 
Detailed information on these features is available in the 
Appendix 
section of this user manual. 
 
Thermo Scientific 
Orion V
ERSA 
S
TAR
 Meter User Manual  
|
  
59
 


Yüklə 6,2 Kb.

Dostları ilə paylaş:
1   ...   18   19   20   21   22   23   24   25   ...   44




Verilənlər bazası müəlliflik hüququ ilə müdafiə olunur ©genderi.org 2024
rəhbərliyinə müraciət

    Ana səhifə