Chapter 5  
|
  
Conductivity Measurement Module 
Conductivity Verification Procedure  
Use the Orion Star series conductivity verification kit, Catalog Number 1010001, to verify the 
accuracy of the conductivity measurements on the V
ERSA 
S
TAR
 meter with conductivity module. 
The conductivity value for each resistor should fall within the relative accuracy of the resistor 
(±0.1 % of the actual conductance of the resistor) plus the relative accuracy of the meter  
(±0.5 % of reading ±1 digit for conductivity readings greater than 3 μS/cm and ±0.5 % of 
reading ±0.01μS/cm for conductivity readings less than or equal to 3 μS/cm). 
1.
 
Ensure that the conductivity resistor kit is certified and calibration date valid. 
2.
 
Press the f1 (Cal) key to start the calibration.  
a.
 
If more than one channel is displayed in the measurement mode, press the  or  
key to highlight the desired channel and press the f3 (Select) key. 
3.
 
Press the f2 (Cell K) key, wait for the cell constant value on the meter to stabilize and press 
the f3 (Clear) key. Use the numeric keypad to enter the cell constant value as “1.000” and 
press the f2 (Accept) key to confirm the entered value. 
4.
 
Press the measure (esc) key to proceed to the measurement mode. 
5.
 
Connect a resistor to the meter, press the measure (esc) key, wait for the measurement 
value to stabilize and record the conductivity value of the resistor. Repeat the procedure for 
all six resistors in the kit. 
 
Resistor Number  Nominal Resistance  Nominal Conductance  Nominal Acceptance Range 
1010001-A 
1000 KΩ 
1 μS 
0.984 to 1.016 μS 
1010001-B 
100 KΩ 
10 μS 
9.930 to 10.07 μS 
1010001-C 
10 KΩ 
100 μS 
99.30 to 100.7 μS 
1010001-D 
1 KΩ 
1000 μS 
993.0 to 1007 μS 
1010001-E 
100 Ω 
10 mS 
9.930 to 10.07 mS 
1010001-F 
10 Ω 
100 mS 
99.30 to 100.7 mS 
 
Note: These are the nominal values for the resistors only. The actual resistance and 
conductance is printed on each resistor and the actual acceptance range must be calculated 
from the actual value on each resistor.  
 
Thermo Scientific 
Orion V
ERSA 
S
TAR
 Meter User Manual  
|
  
63
 


 

RDO/DO Measurement Module  
Connecting Electrodes to the Module 
To begin using the V
ERSA 
S
TAR
 meter and module system, prepare the meter and insert the 
module into an available channel on the back of the meter. See 
Chapter 2 Meter Basics
 for 
instructions. For information on customizing meter settings, refer to 
Chapter 3 Setup Menu

Ø
 
It is recommended that electrodes be placed in the meter-attached electrode stand for easy 
movement in and out of beakers during calibration, sample measurement and storage. 
VSTAR-RD
 
V
ERSA 
S
TAR 
RDO/Dissolved Oxygen Module 
 
1.
 
Connect the dissolved oxygen (DO) probe to the 9 pin MiniDIN connector on the module. 
The temperature will also be measured when the DO probe is connected. 
Ø
 
An Orion RDO optical DO probe or Orion polarographic DO probe can be connected 
and the meter will automatically detect the probe type.  
2.
 
Prepare the DO probe for use as instructed in the probe user manual.  
C H A P T E R   6  
9 pin MiniDIN 
64  
|
  
Orion V
ERSA 
S
TAR
 Meter User Manual 
Thermo Scientific 


Chapter 6  
|
  
RDO/DO Measurement Module 
Channel 1-4 Setup Menus 
Mode Menus for Dissolved Oxygen 
Within the Channel 1, Channel 2, Channel 3 and Channel 4 setup menus are Method, Mode 
and Temperature submenus, which can be used to customize measurement settings and 
parameters for the selected channel. This section provides detailed information on the Mode 
menus for each measurement parameter available when using a RDO/DO module.  
1.
 
In the measurement mode, press the setup key. 
2.
 
Press the  or  key to highlight Channel 1, Channel 2, Channel 3 or Channel 4 (channel 
number the module is connected with) and press the f3 (Select) key. 
 
3.
 
Press the  or  key to highlight Mode and press the f3 (Select) key. 
 
4.
 
Press the  or  key to highlight a menu option and press the  or  key to highlight a 
submenu option. Onscreen text prompts provide instructions for changing the parameters. 
a.
 
Use the popup keypad screen
 for 
Electrode SI No and Sample ID. Press the 
▲, ▼, ◄ or ► 
key to highlight an alphanumeric character (numeric only for 
Auto Increment Sample ID), 
p
ress the f2 (Enter) key to input the character and repeat until the desired label is 
entered. Press the f1 (Done) key to save the entry.     
    
 
 
 