Chapter 8  
|
  
Data Transfer and Software Updates 
Meter Calibration Data from GETCAL # Command 
Mode 
Printed Format 
pH 
Meter Type, Meter Serial Number, Meter Software Revision, Channel Number, Electrode 
ID, Module Serial Number, Method Number, Calibration Date, Calibration Time, User ID, 
Total Calibration Points,  
(repeated for each calibration point) Calibration Point Number, pH Unit, pH Value, mV 
Unit, mV Value, Temperature Value, Temperature Unit, Calibration Type, 
(repeated for each point-to-point segment) Segment Number, Slope Value, Slope Unit, 
(repeated for each point-to-point segment) Segment Number, Offset Value, Offset Unit, 
Average Slope Value, Average Slope Unit, Calibration Log Number 
Example (5 point calibration): VERSA STAR, V00000, 2.73, Ch-1, SENSOR-01, 
VA12345, M-101, 15-03-14, 1:30:30 PM, USER-01, 5, 1, pH, 1.678, mV, 290.4, 25.0, C 
(MAN), Auto, 2, pH, 4.01, mV, 174.3, 25.0, C (MAN), Auto, 3, pH, 6.997, mV, 0.0, 25.0, C 
(MAN), Auto, 4, pH, 10.013, mV, -173.7, 25.0, C (MAN), Auto, 5, pH, 12.46, mV, -289.5, 
25.0, C (MAN), Auto, 1, -49.8, mV, 2, -58.4, mV, 3, -57.6, mV, 4, -47.3, mV, 1, 25.4, mV, 
2, -0.2, mV, 3, -0.2, mV, 4, -31.1, mV, 90.0, %, 101  
RmV 
Meter Type, Meter Serial Number, Meter Software Revision, Channel Number, Electrode 
ID, Module Serial Number, Method Number, Calibration Date, Calibration Time, User ID, 
Relative mV Value, Relative mV Unit, Raw mV Value, Raw mV Unit, Temperature Value, 
Temperature Unit, Relative mV Offset Value, Relative mV Offset Unit, Calibration Log 
Number 
Example: VERSA STAR, V00000, 2.73, Ch-1, SENSOR-01, VA12345, M-101, 15-03-14, 
1:30:30 PM, USER-01, 199.8, RmV, 199.7, mV, 25.0, C (MAN), 0.1, mV, 101  
ORP 
Meter Type, Meter Serial Number, Meter Software Revision, Channel Number, Electrode 
ID, Module Serial Number, Method Number, Calibration Date, Calibration Time, User ID, 
ORP Value, ORP Unit, Raw mV Value, Raw mV Unit,  Temperature Value, Temperature 
Unit, ORP Offset Value, ORP Offset Unit, Calibration Log Number 
Example: VERSA STAR, V00000, 2.73, Ch-1, SENSOR-01, VA12345, M-101, 15-03-14, 
1:30:30 PM, USER-01, 419.5, RmV, 199.8, mV, 25.0, C (MAN), 219.7, mV, 101  
ISE 
Meter Type, Meter Serial Number, Meter Software Revision, Channel Number, Electrode 
ID, Electrode Type, Module Serial Number, Method Number, Calibration Date, 
Calibration Time, User ID, Total Calibration Points,  
(repeated for each calibration point) Calibration Point Number, Concentration Value, 
Concentration Unit, mV Value, mV Unit, Temperature Value, Temperature Unit, 
Calibration Type, 
(repeated for each point-to-point segment) Segment Number, Slope Value, Slope Unit, 
(repeated for each point-to-point segment) Segment Number, Offset Value, Offset Unit, 
Average Slope Value, Average Slope Unit, Blank Function, Blank Value, Blank Unit, 
Calibration Log Number 
Example (5 point calibration): VERSA STAR, V00000, 2.73, Ch-2, SENSOR-02, F-, 
VA12345, M-201, 15-03-14, 1:30:30 PM, USER-01, 5, 1, 0.1, ppm, 174.2, mV, 25.0, C 
(MAN), Manual, 2, 1, ppm, 116.2, mV, 25.0, C (MAN), Manual, 3, 10, ppm, 58.0, mV, 
25.0, C (MAN), Manual, 4, 100, ppm, 0.0, mV, 25.0, C (MAN), Manual, 5, 1000, ppm, -
57.8, mV, 25.0, C (MAN), Manual, 1, -58.0, mV/dec, 2, -58.2, mV/dec, 3, -58.0, mV/dec, 
4, -57.8, mV/dec, 1, 116.2, mV, 2, 116.2, mV, 3, 116.0, mV, 4, 115.6, mV,  -58.0, 
mV/dec, 0, 0.000, ppm, 201  
Thermo Scientific 
Orion V
ERSA 
S
TAR
 Meter User Manual  
|
  
85
 


Chapter 8  
|
  
Data Transfer and Software Updates 
Mode 
Printed Format 
Incremental 
Technique 
(Single 
Known 
Addition and 
Single 
Known 
Subtraction) 
Meter Type, Meter Serial Number, Meter Software Revision, Channel Number, Electrode 
ID, Electrode Type, Module Serial Number, Method Number, Calibration Date, 
Calibration Time, User ID, Incremental Type, Slope Value, Slope Unit, Sample Volume 
Value, Sample Volume Unit, Total Volume Value, Total Volume Unit, Standard 
Concentration Value, Standard Concentration Unit, Reaction Ratio (known subtraction 
only), Measured Sample Value, Measured Sample Unit, Standard Volume Value, 
Standard Volume Unit, Measured Sample and Standard Value, Measured Sample and 
Standard Unit, Temperature Value, Temperature Unit, Calculated Concentration Value, 
Calculated Concentration Unit, Calibration Log Number 
Example: VERSA STAR, V00000, 2.73, Ch-2, SENSOR-02, F-, VA12345, M-201, 15-03-
14, 1:30:30 PM, USER-01, Single Known Addition, -59.2, mV/dec, 100.000, mL, 102.000, 
mL, 100.000, ppm, 0.0, mV, 1.000, mL, -57.8, mV, 25.0, C (MAN), 0.117, ppm, 201 
 
Example: VERSA STAR, V00000, 2.73, Ch-2, SENSOR-02, F-, VA12345, M-201, 15-03-
14, 1:30:30 PM, USER-01,Single Known Subtraction, -59.2, mV/dec, 100.000, mL, 
102.000, mL, 0.100, molar, 0.500, 0.0, mV, 5.000, mL, 58.0, mV, 25.0, C (MAN), 
0.00281, molar, 202  
Incremental 
Technique 
(Double 
Known 
Addition and 
Double 
Known 
Subtraction) 
Meter Type, Meter Serial Number, Meter Software Revision, Channel Number, Electrode 
ID, Electrode Type, Module Serial Number, Method Number, Calibration Date, 
Calibration Time, User ID, Incremental Type, Slope Value, Slope Unit, Sample Volume 
Value, Sample Volume Unit, Total Volume Value, Total Volume Unit, Standard 
Concentration Value, Standard Concentration Unit, Reaction Ratio (known subtraction 
only), Measured Sample Value, Measured Sample Unit, First Standard Volume Value, 
First Standard Volume Unit, Measured Sample and First Standard Value, Measured 
Sample and First Standard Unit, Second Standard Volume Value, Second Standard 
Volume Unit, Measured Sample and Second Standard Value, Measured Sample and 
Second Standard Unit, Temperature Value, Temperature Unit, Calculated Concentration 
Value, Calculated Concentration Unit, Calibration Log Number 
Example: VERSA STAR, V00000, 2.73, Ch-2, SENSOR-02, F-, VA12345, M-201, 15-03-
14, 1:30:30 PM, USER-01, Double Known Addition, -58.0, mV/dec, 100.000, mL, 
102.000, mL, 100.000, ppm, 0.0, mV, 1.000, mL, -17.8, mV, 10.000, mL, -63.3, mV, 25.0, 
C (MAN), 1.040, ppm, 203  
Example: VERSA STAR, V00000, 2.73, Ch-2, SENSOR-02, F-, VA12345, M-201, 15-03-
14, 1:30:30 PM, USER-01, Double Known Subtraction, -58.0, mV/dec, 100.000, mL, 
102.000, mL,0.100, molar, 0.500, 0.0, mV, 1.000, mL, 8.9, mV, 10.000, mL, 32.3, mV, 
25.0, C (MAN), 0.00140, molar, 204  
Conductivity 
Meter Type, Meter Serial Number, Meter Software Revision, Channel Number, Electrode 
ID, Module Serial Number, Method Number, Calibration Date, Calibration Time, User ID, 
Total Calibration Points,  
(repeated for each calibration point) Calibration Point Number, Conductivity Value, 
Conductivity Unit, Conductance Value, Conductance Unit, Temperature Value, 
Temperature Unit, Calibration Type, 
(repeated for each calibration point) Calibration Point Number, Cell Constant Value, Cell 
Constant Unit, 
Average Cell Constant Value, Average Cell Constant Unit, Calibration Log Number 
Example (1 point calibration): VERSA STAR, V00000, 2.73, Ch-3, SENSOR-03, 
VA12345, M-301, 15-03-14, 1:30:30 PM, USER-01, 1, 1, 100.0, uS/cm, 99.98, uS, 25.0, 
C (MAN), Auto, 1, 1.0010, /cm, 1.0010, /cm, 1, 0.00, uS/cm, 301  