Chapter 8  
|
  
Data Transfer and Software Updates 
Computer and Printer Compatibility  
Computer Requirements  
The V
ERSA 
S
TAR
 USB driver and software update program have been tested with Microsoft
®
 
Windows
®
 XP, Windows Vista
®
 and Windows 7 operating systems only. We are always working 
to update our programs; however, they are not compatible with Windows 8 or any other 
operating systems at this time.  
The computer must be up-to-date with the latest Windows updates installed and have .NET 
framework version 4.0 with the latest updates.  
It is strongly recommended that all computer screen savers and power management settings be 
disabled when the meter is interfaced with the computer. The computer must not go into 
sleep/standby mode while the software update is performed. Many laptop computers go into 
sleep/standby mode when the lid is closed, so the lid should remain open throughout the 
software update. Laptop computers must also be plugged into a power supply when running the 
software update – performing the software update while the computer is on battery power is not 
recommended. 
Printer Requirements 
V
ERSA 
S
TAR
 meters can print directly to the Orion Star series RS232 inkjet printer (Catalog 
Number 1010006) or certain models of USB dedicated printers with PCL 5 or PCL 6 printer 
language capability.   
V
ERSA 
S
TAR
 Meter Remote Control Protocols 
Rules for Remote Control Usage  
1.
 
CR (Carriage Return, ASCII 13) is used to terminate a command. Whenever this 
character is received, the internal buffer will be processed. 
2.
 
All characters except for NL (New Line, ASCII 10) are significant. The NL character will 
be ignored. The meter is not case sensitive, though all lower case characters are 
converted to upper case internally. 
3.
 
Only one command can be executed at a time. No new command can issued until the 
current command is finished being processed. When the current command is finished, 
it will issue the “>” prompt to the user, indicating a new command can be entered. 
4.
 
Empty commands (i.e. just a CR) will be ignored and a new prompt will be issued.  
5.
 
Set the meter communication settings using the
 Instrument Settings
 setup menu.  
Communication Settings  Default RS232 Settings  Default USB Settings 
Bits per second 
9600 
115200 
Data bits 


Parity 
None 
None 
Stop bits 


Flow control 
None 
None 
80  
|
  
Orion V
ERSA 
S
TAR
 Meter User Manual 
Thermo Scientific 


Chapter 8  
|
  
Data Transfer and Software Updates 
Remote Control Engine 
The remote control engine receives input from the serial/USB port and processes it as needed. 
Commands sent to the remote control interface will be in the form of “OPCODE 
 
CR
”. Line feeds will be ignored. New commands cannot be issued until the 
previous command has been completed and a prompt is given. A prompt is shown as the 
greater than symbol (“>“‘) followed by a space. 
Remote Control Commands 
GETMEAS 
GETMEAS  
Prints the current displayed channel measurements immediately 
GETMEAS Data Count  
Prints the current display channel measurement for a set number of times 
Example: GETMEAS 2  
GETMEASTIMED CH_Channel Combination, Time Interval  
Set the channels to be measured and the time interval for the measurements 
Channel Combination: Channel numbers to be measured 
Time Interval: The measurement retrieving time interval in seconds 
Example: GETMEASTIMED CH_12 5   
STOP  
Ends the timed measurement command   
GETCAL 
GETCAL #  
Prints all calibration data for the selected mode for all applicable channels 
If no calibration is available, will return “>” to receive next command 
Select calibration data for specific mode as a number: 
PH = 1 
COND = 6 
RMV = 3  DO = 10 
ORP = 4  RDO = 12 
ISE = 5 
INCREMENTAL = 21 
Example: GETCAL 1   
GETLOG 
GETLOG  
Prints all the logged measurement data 
If no data is logged, will return “>” to receive next command 
Example: GETLOG   
GETLOG START, END  
Prints the data in a specified range, according to the selected print format 
If no data is available, will return “>” to receive next command 
Example: GETLOG 10, 100  
SYSTEM 
SYSTEM  
Prints the Meter Type, Serial Number, Software Version 
Example: Versa Star, V00000, 2.73 
Thermo Scientific 
Orion V
ERSA 
S
TAR
 Meter User Manual  
|
  
81
 


Chapter 8  
|
  
Data Transfer and Software Updates 
SETRTC 
SETRTC YY MM DD HH MM SS  
Set the date and time (in 24 hour format) for the meter 
Example: SETRTC 14 03 15 18 15 30 00  
SETMODE 
SETMODE CHANNEL MODE   
Set the channel as: 
1          2          3          4 
Set the channel measurement mode as: 
PH 
COND 
DOSAT  
MV 
TDS  
DOCON 
RMV 
SAL 
ORP 
RES 
ISE 
 
Example: SETMODE 1 PH  
GETMODE 
GETMODE CHANNEL  
Print the channel measurement mode 
Example: GETMODE 1 
ESC 
ESC  
Stops the command being executed 
STOP 
STOP  
Command stops continuous measurement from meter 
 
 
Meter Measurement Data from GETMEAS Command 
Mode 
Printed Format 
pH 
Meter Type, Meter Serial Number, Meter Software Revision, Date, Time, Channel 
Number, Electrode ID, Sample ID, User ID, Module Serial Number, Mode, pH Value, pH 
Unit, mV Value, mV Unit, Temperature Value, Temperature Unit, Slope Value, Slop Unit, 
Method Number  
Example: VERSA STAR, V00000, 2.73, 15-03-14, 1:30:30 PM, Ch-1, SENSOR-01, 1, 
USER-01, VA12345, pH, 7.000, pH, 0.0, mV, 25.0, C (MAN), 100.0, %, M101 
 
mV 
Meter Type, Meter Serial Number, Meter Software Revision, Date, Time, Channel 
Number, Electrode ID, Sample ID, User ID, Module Serial Number, Mode, mV Value, mV 
Unit, Temperature Value, Temperature Unit, Method Number 
Example: VERSA STAR, V00000, 2.73, 15-03-14, 1:30:30 PM, Ch-1, SENSOR-01, 1, 
USER-01, VA12345, MV, 0.0, mV, 25.0, C (MAN), M101  
RmV 
Meter Type, Meter Serial Number, Meter Software Revision, Date, Time, Channel 
Number, Electrode ID, Sample ID, User ID, Module Serial Number, Mode, Relative mV 
Value, Relative mV Unit, Raw mV Value, Raw mV Unit, Temperature Value, Temperature 
Unit, Method Number 
Example: VERSA STAR, V00000, 2.73, 15-03-14, 1:30:30 PM, Ch-1, SENSOR-01, 1, 
USER-01, VA12345, RMV, 0.1, RmV, 0.0, mV, 25.0, C (MAN), M101  
82  
|
  
Orion V
ERSA 
S
TAR
 Meter User Manual 