Chapter 8
|
Data Transfer and Software Updates
Computer and Printer Compatibility
Computer Requirements
The V
ERSA
S
TAR
USB driver and software update program have been tested with Microsoft
®
Windows
®
XP, Windows Vista
®
and Windows 7 operating systems only. We are always working
to update our programs; however, they are not compatible with Windows 8 or any other
operating systems at this time.
The computer must be up-to-date with the latest Windows updates installed and have .NET
framework version 4.0 with the latest updates.
It is strongly recommended that all computer screen savers and power management settings be
disabled when the meter is interfaced with the computer. The computer must not go into
sleep/standby mode while the software update is performed. Many laptop computers go into
sleep/standby mode when the lid is closed, so the lid should remain open throughout the
software update. Laptop computers must also be plugged into a power supply when running the
software update – performing the software update while the computer is on battery power is not
recommended.
Printer Requirements
V
ERSA
S
TAR
meters can print directly to the Orion Star series RS232 inkjet printer (Catalog
Number 1010006) or certain models of USB dedicated printers with PCL 5 or PCL 6 printer
language capability.
V
ERSA
S
TAR
Meter Remote Control Protocols
Rules for Remote Control Usage
1.
CR (Carriage Return, ASCII 13) is used to terminate a command. Whenever this
character is received, the internal buffer will be processed.
2.
All characters except for NL (New Line, ASCII 10) are significant. The NL character will
be ignored. The meter is not case sensitive, though all lower case characters are
converted to upper case internally.
3.
Only one command can be executed at a time. No new command can issued until the
current command is finished being processed. When the current command is finished,
it will issue the “>” prompt to the user, indicating a new command can be entered.
4.
Empty commands (i.e. just a CR) will be ignored and a new prompt will be issued.
5.
Set the meter communication settings using the
Instrument Settings
setup menu.
Communication Settings Default RS232 Settings Default USB Settings
Bits per second
9600
115200
Data bits
8
8
Parity
None
None
Stop bits
1
1
Flow control
None
None
80
|
Orion V
ERSA
S
TAR
Meter User Manual
Thermo Scientific
Chapter 8
|
Data Transfer and Software Updates
Remote Control Engine
The remote control engine receives input from the serial/USB port and processes it as needed.
Commands sent to the remote control interface will be in the form of “OPCODE
CR
”. Line feeds will be ignored. New commands cannot be issued until the
previous command has been completed and a prompt is given. A prompt is shown as the
greater than symbol (“>“‘) followed by a space.
Remote Control Commands
GETMEAS
GETMEAS
Prints the current displayed channel measurements immediately
GETMEAS Data Count
Prints the current display channel measurement for a set number of times
Example: GETMEAS 2
GETMEASTIMED CH_Channel Combination, Time Interval
Set the channels to be measured and the time interval for the measurements
Channel Combination: Channel numbers to be measured
Time Interval: The measurement retrieving time interval in seconds
Example: GETMEASTIMED CH_12 5
STOP
Ends the timed measurement command
GETCAL
GETCAL #
Prints all calibration data for the selected mode for all applicable channels
If no calibration is available, will return “>” to receive next command
Select calibration data for specific mode as a number:
PH = 1
COND = 6
RMV = 3 DO = 10
ORP = 4 RDO = 12
ISE = 5
INCREMENTAL = 21
Example: GETCAL 1
GETLOG
GETLOG
Prints all the logged measurement data
If no data is logged, will return “>” to receive next command
Example: GETLOG
GETLOG START, END
Prints the data in a specified range, according to the selected print format
If no data is available, will return “>” to receive next command
Example: GETLOG 10, 100
SYSTEM
SYSTEM
Prints the Meter Type, Serial Number, Software Version
Example: Versa Star, V00000, 2.73
Thermo Scientific
Orion V
ERSA
S
TAR
Meter User Manual
|
81
Chapter 8
|
Data Transfer and Software Updates
SETRTC
SETRTC YY MM DD HH MM SS
Set the date and time (in 24 hour format) for the meter
Example: SETRTC 14 03 15 18 15 30 00
SETMODE
SETMODE CHANNEL MODE
Set the channel as:
1 2 3 4
Set the channel measurement mode as:
PH
COND
DOSAT
MV
TDS
DOCON
RMV
SAL
ORP
RES
ISE
Example: SETMODE 1 PH
GETMODE
GETMODE CHANNEL
Print the channel measurement mode
Example: GETMODE 1
ESC
ESC
Stops the command being executed
STOP
STOP
Command stops continuous measurement from meter
Meter Measurement Data from GETMEAS Command
Mode
Printed Format
pH
Meter Type, Meter Serial Number, Meter Software Revision, Date, Time, Channel
Number, Electrode ID, Sample ID, User ID, Module Serial Number, Mode, pH Value, pH
Unit, mV Value, mV Unit, Temperature Value, Temperature Unit, Slope Value, Slop Unit,
Method Number
Example: VERSA STAR, V00000, 2.73, 15-03-14, 1:30:30 PM, Ch-1, SENSOR-01, 1,
USER-01, VA12345, pH, 7.000, pH, 0.0, mV, 25.0, C (MAN), 100.0, %, M101
mV
Meter Type, Meter Serial Number, Meter Software Revision, Date, Time, Channel
Number, Electrode ID, Sample ID, User ID, Module Serial Number, Mode, mV Value, mV
Unit, Temperature Value, Temperature Unit, Method Number
Example: VERSA STAR, V00000, 2.73, 15-03-14, 1:30:30 PM, Ch-1, SENSOR-01, 1,
USER-01, VA12345, MV, 0.0, mV, 25.0, C (MAN), M101
RmV
Meter Type, Meter Serial Number, Meter Software Revision, Date, Time, Channel
Number, Electrode ID, Sample ID, User ID, Module Serial Number, Mode, Relative mV
Value, Relative mV Unit, Raw mV Value, Raw mV Unit, Temperature Value, Temperature
Unit, Method Number
Example: VERSA STAR, V00000, 2.73, 15-03-14, 1:30:30 PM, Ch-1, SENSOR-01, 1,
USER-01, VA12345, RMV, 0.1, RmV, 0.0, mV, 25.0, C (MAN), M101
82
|