XM-18

[OK]
and then raise your
hand to save data, and automatically move into the next parameter setting.
Set up upper limit value of temperature control :
Led display as shown in right figure:
 
"P3" is the upper limit of temperature
control. The heating will stop when the temperature reaches this value. If any change is needed,
please press [▲▼] untill it displays the value you required. And press 
[OK]
and then raise your
hand to save data, and automatically move into the next parameter setting.
Set up lower limit value of temperature control:
Led display asshown in right figure:
"P4" is the lower limit of temperature
control. The heating will start up when the temperature droppes to this value. If any change is
needed, please press [▲▼] untill displays the value you required. And press 
[OK]
and then raise
your hand to save data, and automatically move into the next parameter setting. 
Set up low temperature value:
Led display asshown in right figure:
shown in right figure: "P5" is the low
temperature value. The standby heating will start up when the temperature droppes to this value. If
any change is needed, please press [▲▼] untill it displays the value you required. And press 
[OK]
and then raise your hand to save data, and automatically move into the next parameter setting.
Set up insufficient-temperature alarm value:
Led display as shown in right figure: 
 
"P6" is the insufficient-temperature alarm
value. The alarm will be given if the temperature droppes to this value. If any change is needed,
please [▲▼] press untill it displays the value you required. And pres 
[OK]
and then raise your
hand to save data, and automatically move into the next parameter setting.
Set up over-humidity alarm:
Led display as shown in right figure: 
"P7" is the over-humidity alarm value. The
alarm will be given if the humidity reaches this value. If any change is needed, please press [▲▼]
untill it displays the value you required. And press 
[OK]
and then raise your hand to save data, and
automatically move into the next parameter setting. 