Calibration_can_be_also_loaded_from_calibration_data_file_._Energy_calibration_process_Calibration_–_general_principles_10.9'>TUKAN 8k
Calibration can be also loaded from 
calibration data file
.
Energy calibration process
Calibration – general principles
10.9
Copying calibration from spectrum to spectrum
In some cases (ie. series of measurement) is needed to transfer calibration
from one spectrum to the another. This operation can be done in a following
way:
read the spectrum with calibration to 
spectrum container
 and then
read this calibration to the spectrum panel in 
Calibration module
,
press the button Move calib. to program
 
placed on the 
calibration panel
 - calibration from the spectrum will be read to the calibration control
panel
read the target spectrum,
run operation 
Set calibration to spectrum
Calibration – general principles
10.10 Show calibration of main spectrum
The dialog window with calibration preview is showed below. It shows
coefficients of the calibration equations and function plots for whole three types
of calibration.
138
127
40
81
34
123
133
121


135
Calibration
This window is presented after executing operation Spectrum calibration
show, which can be find in ANALYZER and ANALYSIS modules, in menu
Analyse. Calibrations presented on this window always concern to just sowed
spectrum.
Spectrum calibration show in ANALYZER and ANALYSIS modules:
icon:   
menu:  Analysis | Spectrum calibration show
Spectrum calibration show in CALIBRATION module
icon:   
menu:  Calibration | Spectrum calibration show
operation panel:   Show
Calibration
10.11 Calibration of ACQUISITION spectrum
The notion "acquisition path" is defined in Tukan as an ansamble of devices (and
their settings) used to measurement. Elements of  "acquisition path" are:
detector, amplifier gain, measurement geometry etc., and also: spectrum
connected with the buffer of the analyser card. Calibration made for this
"acquisition path" stays a part of its and it is as long as parameters of the
"acquisition path" are constant. 
For simplifying, program defines calibration of the ACQUISITION spectrum. The
following rules are valid:
1. calibration set in spectrum_MEASUREMENT is a calibration of the "acquisition
path",
2. this calibration is stored during closing program and reminded after next
121


136
TUKAN 8k
execution,
3. calibration is automatically set to every next spectrum_MEASUREMENT
Calibration of the "acquisition path" can be read 
calibration control panel
 and
 
set into
 the spectra measured using the same "acquisition path".
Wczytanie kalibracji toru pomiarowego
icon:   
menu:  Calibration | Load acquisition calibration
operation panel ("Acq. calibration" panel):  Load
Calibration
10.12 Calibration removing
Calibration can be removed from spectrum only in module Calibration and only if
spectrum is displayed as a
 Main spectrum 
Calibration removing in spectrum
icon:  
menu:  Calibration | Delete calibration in spectrum
calibration control panel (at the bottom): Delete
This operation removes calibration  in main spectrum (displayed on the screen).
Then, the following message is displayed on screen:
"Are you sure that you want to delete calibration in the spectrum?"
Calibration is removed after "Yes" response.
Setting calibration to spectrum
10.13 Calibration curve
Calibration results are displayed on screen as the coefficient of calibration
equation and calibration curve diagram.
Calibration nodes taken from the calibration data table are displayed on a
diagram. Either diagram axis coordinates automatically adjust to the calibration
curve or, in case, there is no curve, to calibration nodes.
121
133
121
133


137
Calibration
There is a splitting bar between calibration and spectrum panel to change the
heights of both panels. You can zoom eny part of calibration spectra. To do this,
drag the mouse over the part to zoom. Right clicking of the mouse allows
dragging the calibration plot freely horizontally and vertically.
Operation "Print curve" opens a dialog window showing a print preview and
allowing printer selection and setting printing parameters. The curve is printed in
the same shape as in the preview (all printing colors are identical if available on
the printer).
Show calibration of main spectrum
134


138
TUKAN 8k
Setting calibration to spectrum
10.14 Calibration file
The files containing the calibration data have default „.clb" extension. The file
contains a set of full calibration data, selected calibration function type and
coefficients values for all three types of calibration.
Save calibration to a calibration file:
icon:  
 
menu:  Calibration | Save to file...
context menu of calibration panel: Save to file.
Program saves such a dataset, which is displayed at the moment of saving.
By default, calibration file is given the same name as a spectrum from which
calibration is loaded.
Load calibration from calibration file:
icon:  
menu:  Calibration | Load from file...
context menu of calibration panel: Load from file.
All calibration data loaded from a file are displayed on calibration panel. Loading
them to a spectrum takes place after executing Set to spectrum command.
Setting calibration to spectrum
133
133