139
Report
11
Report
Report of measurement may be printed after switching to a REPORT module.
This module consists of two parts:
·
print preview panel representing page of paper and 
·
thumbnail preview panel with miniatures of all pages of the report. 
Thumbnail preview helps with navigation on multi pages report. Report contains
header with information about spectrum and measurement, calibrations
information, table of peaks and figure of the spectrum.
Report content may be easily changed, each part of report is fully configurable
(see next subsections). Report module allows to print generated reports by
windows print dialog. User may scale the print preview by changing Zoom drop
down menu.
Report configuration
Printing report
140
144


140
TUKAN 8k
11.1
Report configuration
There are two ways to configure content of the report:
˙
click on button Configure report
˙
from menu: Configuration | Configure report…
The Report configuration window contains 4 tabs:
General
Acquisition information
Calibrations
Peak table
Each tab corresponds to the section of the report presented in the main panel
and allows to configure the section:
General gives possibility to plug-in and plug-off parts to the report and set
orientation of the paper (only A4 paper size is supported now).
Acquisition information tab controls description of the spectrum and acquisition
path.
Calibration tab controls information about calibration of the spectrum.
The most extended tab is Peak table, which has been described in section:
Peaks table configuration

Clicking OK in the bottom part of the window closes window, stores all setting
and generates new report regarding new settings. User may save current
configuration in the file (*.cfr) - menu Configuration | Save configuration… and
load saved configuration from file - menu Configuration | Load
configuration… Program has built-in Default configuration - menu
Configuration | Default configuration.
141


141
Report
Spectrum printed in the report is configured by:
˙
click on button Configure spectrum
˙
from menu: Configuration | Configure spectrum…
Configuration spectrum window gives possibility to set properties of the
spectrum figure similar to the spectrum panels in other Tukan 8k modules:
zoom, colors, etc. Resizing the configuration window gives the result in the final
report spectrum figures proportion.
It is also possible to save text part of the report to a text file and a spectrum
diagram to another file applying any of graphical formats.
Reports module
Printing report
Peaks table configuration
11.1.1
The window 
Report configuration
 contains the tab Peak table.
Peak table gives possibility to mange the contents of the ROI/peak table. Peak
table contains information about all ROIs in the spectrum. It consists of sort of
columns.  User may choose which column will be displayed in the table and the
unit of the columns. 
36
144
140


142
TUKAN 8k
Columns order is configurable also by "drag and drop" columns in the bottom
part of Peak table tab.
All setting can be save in  
Configuration file
.
Report configuration
Spectrum plot configuration
11.1.2
Report module makes possible configuration of the contents of the figure with
spectrum, which is add
to the report.
Configuration window is run from:
icon:  
143
140


143
Report
menu:  Configuration | Configure spectrum...
User may change spectrum size, colors, add grid etc. (see: 
Plotting a spectrum
).
Changing size of the Configuration window influences on proportion of the
spectrum figure presented in the Report (including print).
Report configuration
Configuration file
11.1.3
Settings of the report configuration may be saved in configuration file. 
It could be done only from menu of the configuration window:
menu of the configuration window: Configuration | Save
configuration...
  
93
140


144
TUKAN 8k
Configuration file may be loaded by: 
Configuration | Load configuration...
Additionally program contains built-in standard configuration settings, loaded
by:
Configuration | Default configuration
Configuration data are saved in files with extension *.cfg. 
These files are saved in working directory by default, but may be saved
anywhere else.
Report configuration
11.2
Printing report
Report print:
icon:  
  (on spectrum panel toolbar)
menu: File | Print report...
140


145
Report
NOTE: Changing orientation in Windows Print… dialog does not effect –
program forces configured orientation.
Report configuration
Reports
140
139


146
TUKAN 8k
11.3
HTML and Text format of the report
The measurement report may be saved as HTML file or as a plain text. Raport
saving may be done using menu or icons.
Save report to a HTML file:
icon:
 
menu
:
 File | Save as HTML...
and:
menu
:
 File | Save as HTML and show in web browser
HTML report looks similar to printed one. Graphical elemnts (mathematical
formulas and spectrum figure) are saved as a JPEG files (Tukan8k version
1.6). Second option, Save as HTML and show in web browser, opens a
defauld browser with just saved report.
Save to a text file:
icon:
  
menu
:
 File | Save as a text file...
and:
menu
:
 File | Save as text file and show in editor
Report is saved as non-formated (plain) text in a file with extention .txt and
does not contain graphical elements.
Seconf option: Save as text file and show in editor saves file and opens its in
editor (ie. Notepad). Contents of the report is editable. 
Report
139


147
Keyboard
12
Keyboard
Markers navigation keys
Displaying spectrum navigation keys
ROI system navigation keys
12.1
Acquisition navigation keys
Acquisition start
Acquisition stop
Acquisition rese
Common acquisition start
Common acquisition stop
Common acquisition reset
Start/Stop/Reset of data acquisition
12.2
Markers navigation keys
selects control to left marker
selects control to right marker
moves one channel to the left
moves one channel to the right
selects control to both (coupled) markers
(the middle key of numeric pad)
147
148
149
56


148
TUKAN 8k
moves coupled markers right
moves coupled markers left
pushes markers aside
brings markers closer
increases twice the number of channels displayed between markers
decreases twice the number of channels displayed between markers
expands a spectrum between markers to the whole width of the
panel
displays spectrum in a full scale
Displaying spectrum navigation keys
ROI system navigation keys
Plotting a spectrum
12.3
Displaying spectrum navigation keys
decreases horizontal scale
increases horizontal scale
 
decreases vertical scale
 
increases vertical scale
moves spectrum to the right
 
moves spectrum to the left
 
moves spectrum up
 
moves spectrum down
expands twice a spectrum selected between markers
squeeze twice a spectrum selected between markers
expands a spectrum between markers to the whole width of the
148
149
93


149
Keyboard
panel
displays spectrum in full scale
Markers navigation keys
ROI system navigation keys
Plotting a spectrum
12.4
ROI system navigation keys
inserts area selected by markers to the ROI table
deletes ROI area selected by markers
moves markers on next ROI to the right
 
moves markers on next ROI to the left
Markers navigation keys
Displaying spectrum navigation keys
Fast selection of ROI
147
149
93
147
148
104


150
Index
  * 
*.wdm     38
*.wds     38
  A 
Acquisition continuation after program shut
down     59
Acquisition navigation keys     147
Acquisition path     37
ACQUISITION spectrum     62
Activity and concentration calculate     113
Adding acquisition path     71
Adding spectra     89
Additional displaying features     97
Advanced analysis     109
Analysis     106
Analysis – general principles     106
Analysis module     35
Analysis of MCS type spectra     69
Analysis options     117
Analyzer internal registers     54
Analyzer module     33
Analyzer parameters in MCS mode     66
Area     43
Automatic acquisition stop criteria     57
Automatic peak search     107
  C 
Calibration     121
Calibration – algorithms     40
Calibration – general principles     121
Calibration control panel     121
Calibration curve     136
Calibration file     138
Calibration module     34
Calibration of ACQUISITION spectrum    
135
Calibration of MCS type spectra     69
Calibration operations     123
Calibration removing     136
Calibration standards editor     131
Colors of the plot panel     99
Common control of acquisition     75
Compressing spectra     90
Configuration file     143
Configuration file of hardware setup     52
Configuration options of measurement     65
Controlling data acquisition     58
Copying calibration from spectrum to
spectrum     134
Ctr     43
  D 
Data files     37
Data in Program Tukan     37
Detection efficiency calculate     129
Detection of analyzer devices     49
Displaying „live” spectrum     63
Displaying options     98
Displaying spectrum navigation keys     148
Dublet separation     111
Dwell Time     66
  E 
Efficiency calculation     42
Efficiency calibration - algorithm     41
Efficiency calibration process     128
Energy calibration - algorithm     40
Energy calibration process     127
Export spectrum to ASCII format     86
  F 
Factory defaults of analyzer     54
Fast selection of ROI     104
First start of the program     16
Fitting the peak with Gauss function     109
Fwhm     43
Fwtm     43
  G 
Geometry     51
  H 
Hardware control     37
Hardware module     32


151
Hardware requirements     11
Hardware setup     48
HTML and Text format of the report     146
  I 
Import spectrum from ASCII format     87
Information about main spectrum     85
Inserting and editing of calibration data    
124
Inserting and editing ROI     103
Inserting data from calibration spectrum    
125
Inserting data from nuclide library     126
Instalation of Tukan8k program     11
Installation     11
Installation for Tukan_8k_PCI card     13
Installation for Tukan-8k-USB     14
Installation for USB Protection Key     15
Integral     43
  K 
Keyboard     147
  L 
Laboratory - description     51
Language version of the program     47
List of reference spectra     83
  M 
Main spectrum     79
Manual calibration data insert     124
Markers     96
Markers navigation keys     147
Mass of sample     51
Mathematical analysis – algorithms     42
Mathematical models of peak analysis     44
MCS - Tukan-8k-PCI     20
MCS - Tukan-8k-USB     25
MCS mode of running     66
Measurement     48
Measurement control in MCS mode     68
Measurement description     51
Measurement session     59
Measurement spectra     75
Measurement spectrum     62
Model f0     43, 44
Model f0c     43
Model f0cm     43
Model f1     44
Model f2     44
Model f3     44
Model f4     44
Model f5     44
Model f6     44
Multi analyzer module     77
  N 
Nuclide library     114
  O 
On-line analysis     64
Opening a spectrum     80
  P 
Parameters of analyzer     52
Peak analysis „on-line”     43
Peak parameters     43
Peak parameters calculate     106
Peak shape calibration - algorithm     41
Peak shape calibration process     128
Peaks identification     112
Peaks table     116
Peaks table configuration     141
Printing a spectrum     91
Printing report     144
Program concept     30
Program configuration     46
Program modules selection     31
Program Tukan8k     28
  R 
Report     139
Report configuration     140
Reports module     36
ROI     102
ROI colors     99
ROI files     105
ROI regions     102


152
ROI system navigation keys     149
ROI Table     103
  S 
Saving a spectrum     80
Selected acquisition path     72
Selection of fitting function models     109
Selection of units     120
Setting calibration to spectrum     133
Show all spectra from container     100
Show calibration of main spectrum     134
Spectra container     81
Spectra displaying - general principles     93
Spectra files formats     38
Spectra management     79
Spectrum "ACQUISITION_MCS"     69
Spectrum copy in memory     84
Spectrum plot configuration     142
Spectrum plot style changing     100
Start/Stop/Reset of data acquisition     56
Subtraction spectra     89
  T 
Time backup of acquisition data     64
Tukan-8k Analyzer     10
Tukan-8k-PCI     17
Tukan-8k-PCI - MCA Mode     18
Tukan-8k-PCI - MCS Mode     20
Tukan-8k-PCI I/O - Connections and LED
Indicators     18
Tukan-8k-USB     22
Tukan-8k-USB - I/O Connections and LED
Indicators     23
Tukan-8k-USB - MCA Mode     24
Tukan-8k-USB - MCS Mode     25
Tukan-8k-USB - powering the device     23
  U 
USB key     47
  V 
Version 1.8 update     28
Version 1.9 update     29
  W 
Work with USB protection key     47
Working with several analyzers     71
Working with the spectra     79
  X 
X scale changing     94
  Y 
Y scale changing     95


Tukan 8k
INSTYTUT PROBLEMÓW JADROWYCH im. Andrzeja Soltana
THE ANDRZEJ SOLTAN INSTITUTE FOR NUCLEAR STUDIES

Document Outline

  • Tukan-8k Analyzer
    • Installation
      • Hardware requirements
      • Instalation of Tukan8k program
      • Installation for Tukan_8k_PCI card
      • Installation for Tukan_8k_USB box
      • Installation for USB Protection Key
      • The first start of the program
    • Tukan-8k-PCI
      • I/O Connections and LED Indicators
      • MCA - Multichannel Analyzer mode
      • MCS - Multichannel Scaling mode
    • Tukan-8k-USB
      • I/O Connections and LED Indicators
      • Powering the device
      • MCA - Multichannel Analyzer mode
      • MCS - Multichannel Scaling mode
  • Program Tukan8k
    • Latest updates
      • Version 1.8 update 
      • Version 1.9 update
    • Program concept
      • Program modules selection
      • Hardware module
      • Analyzer module
      • Calibration module
      • Analysis module
      • Reports module
    • Hardware control
      • Acquisition path
    • Data
    • Calibration – algorithms
      • Energy calibration
      • Peak shape calibration
      • Efficiency calibration
      • Efficiency of detection
    • Mathematical analysis – algorithms
      • Peak analysis 'on-line'
      • Mathematical models of peak analysis
    • Program configuration
    • Language version of the program
    • Work with USB Protection Key
  • Measurement
    • Hardware setup
      • Detection of analyzer devices
      • Measurement description - sample mass
      • Configuration file of hardware setup
    • Parameters of analyzer
      • Factory defaults
      • Analyzer internal registers
    • Start/Stop/Reset of data acquisition
    • Acquisition automatic stop criteria
    • Controling data acquisition
    • Continuation of acquisition after program shut down
    • Measurement session
    • Measurement spectrum
      • Spectrum 'ACQUISITION"
      • Displaying 'live' spectrum
      • On-line analysis
    • Time backup of acquisition data
    • Configuration options of measurement
  • Multi-channel scaling (MCS) mode
    • Analyzer parameters in MCS mode
    • Measurement control in MCS mode
    • Spectrum "ACQUISITION_MCS"
    • Calibration and analysis of MCS type spectra
  • Working with several analyzers
    • Adding acquisition path
      • New features in Hardware module
      • New features in Analyzer module
      • Measurement spectra
    • Coomon control of acquisition
    • "Multi analyzer" module
  • Spectra management
    • Working with the spectra
    • Main spectrum
    • Opening a spectrum
    • Saving a spectrum
    • Spectra container
    • List of reference spectra
    • Spectrum copy in memory
    • Information about main spectrum
    • Export to ASCII format
    • Import from ASCII format
    • Adding and subtraction spectra
    • Compressing spectra
    • Printing a spectrum
  • Plotting a spectrum
    • X scale changing
    • Y scale changing
    • Markers
    • Additional displaying features
    • Displaying options
    • Colors of the plot panel 
    • Spectrum plot style changing
    • Show all spectra from container
  • ROI
    • ROI regions
    • ROI Table
    • Inserting and editing ROI
    • Fast selection of ROI
    • ROI files
  • Analysis
    • Analysis – general principles
    • Peak parameters calculate
    • Automatic peak search
    • Advanced analysis
      • Selection of fitting function models
      • Fitting the peak with Gauss function
      • Dublet separation
    • Peaks identification
    • Activity and concentration calculate
    • Nuclide library editor
    • Peaks table
    • Analysis options
    • Selection of units
  • Calibration
    • Calibration – general principles
    • Calibration control panel
    • Calibration operations
    • Inserting and editing of calibration data
      • Manual data insert
      • Inserting data from calibration spectrum
      • Inserting energy from nuclide library
    • Energy calibration process
    • Peak shape calibration process
    • Efficiency calibration process
      • Detection efficiency calculate
      • Standards editor
    • Setting calibration to spectrum
    • Copying calibration from spectrum to spectrum
    • Show calibration of main spectrum
    • Calibration of ACQUISITION spectrum
    • Calibration removing
    • Calibration curve
    • Calibration file
  • Report
    • Report configuration
      • Peaks table configuration
      • Spectrum plot configuration
      • Configuration file
    • Printing report
    • HTML and Text format of the report
  • Keyboard
    • Acquisition navigation keys
    • Markers navigation keys
    • Displaying spectrum navigation keys
    • ROI system navigation keys

Yüklə 6,31 Kb.

Dostları ilə paylaş:
1   ...   25   26   27   28   29   30   31   32   33




Verilənlər bazası müəlliflik hüququ ilə müdafiə olunur ©genderi.org 2024
rəhbərliyinə müraciət

    Ana səhifə