International



Yüklə 0,72 Mb.
Pdf görüntüsü
səhifə16/62
tarix22.03.2024
ölçüsü0,72 Mb.
#168941
1   ...   12   13   14   15   16   17   18   19   ...   62
IEC60904-9-2020

5.2.3 
Procedure 
a) Spectral irradiance shall be measured at a minimum of 4 locations in the designated test 
area to address potential spatial non-uniformity effects of spectral distribution. The 
selection shall include the locations shown in Figure 1 for a rectangular and a circular 
designated test area. For a rectangular designated test area under a multilamp solar 
simulator, a higher number of locations are deemed necessary if single lamps only 
irradiate a part of the test area. 
Copyrighted material licensed to University of Toronto by Clarivate Analytics (US) LLC, subscriptions.techstreet.com, downloaded on 2020-09-26 08:51:03 +0000 by University of Toronto User.
No further reproduction or distribution is permitted.


IEC 60904-9:2020 © IEC 2020 
– 15 – 
Figure 1 – Locations for spectral irradiance measurement of
a rectangular test area (left) and a circular test area (right) 
b) Spectral irradiance should be reported in tabular form with a minimum of one data point 
per 10 nm interval in the total wavelength range 300 nm to 1 200 nm. Averaging of data 
points is permitted to reduce the number of data points. 
c) The spectral irradiance data taken at all test locations should be integrated (trapezoid type) 
in the total wavelength range (400 nm to 1 100 nm or 300 nm to 1 200 nm) and the 
percentage contribution of the 6 wavelength intervals defined in Table 1 or Table 2 to the 
integrated irradiance determined. 
d) For each test location calculate the spectral match for each wavelength interval, which is 
the ratio of calculated percentage for the simulator spectrum and the AM1.5 reference 
spectral irradiance. 
e) For each test location the data comparison with the reference spectrum shall indicate the 
spectral match classification as per the following: 
Class A+: Spectral match within 0,875 to 1,125 for each wavelength interval, as 
specified in Table 3. This classification can be only assigned if spectral match 
evaluation is performed according to Table 2. 
NOTE Class A+ is restricted for the case that spectral match is evaluated over the extended 
wavelength range of 300 nm to1 200 nm. 
Class A:
Spectral match within 0,75 to 1,25 for each wavelength interval, as specified 
in Table 3. 
Class B: 
Spectral match within 0,6 to 1,4 for each wavelength interval, as specified in 
Table 3. 
Class C:
Spectral match within 0,4 to 2,0 for each wavelength interval, as specified in 
Table 3. 
f) All intervals shown in Table 1 or Table 2 shall agree with the spectral match ratios in 
Table 3 to obtain the respective classes. 
g) The spectral classification of the solar simulator is determined by the test location and 
interval with worst case spectral match. 

Yüklə 0,72 Mb.

Dostları ilə paylaş:
1   ...   12   13   14   15   16   17   18   19   ...   62




Verilənlər bazası müəlliflik hüququ ilə müdafiə olunur ©genderi.org 2024
rəhbərliyinə müraciət

    Ana səhifə