International


  PV module solar simulator



Yüklə 0,72 Mb.
Pdf görüntüsü
səhifə19/62
tarix22.03.2024
ölçüsü0,72 Mb.
#168941
1   ...   15   16   17   18   19   20   21   22   ...   62
IEC60904-9-2020

5.3.2.2 
PV module solar simulator 
Solar simulators are typically used for output power characterization or electrical stabilisation.
Two types of irradiance detector may be used: 
a) Encapsulated c-Si reference cell (single cell detector): The maximum detector size shall 
be 20 cm x 20 cm, preferably a 15,6 cm x 15,6 cm solar cell. 
b) Assembly of detector cells encapsulated in a module (array detector). The maximum 
single detector size is given by the dimensions of a 6-inch solar cell. Spacing between 
cells shall be less than 1 cm. 
The actual grid of the irradiance map will depend on detector size and test area dimensions. 
The ratio of measured area with the uniformity detector and the designated test area shall be 
larger than 80 %. 
For any array detector the relative calibration of cells may change for different irradiance 
levels. Either the calibration shall be performed at different irradiance levels or non-linearity 
impacts shall be considered in the uncertainty budget. 
5.3.2.3 
PV cell solar simulator 
The detector size should not be larger than 1/5 of the smallest dimension of the test area. 
Both single cell detector and assembly array detector with encapsulated cells can be used. 
If one of the dimensions of the test area is smaller than 10 cm, the use of photodiodes or 
masked solar cells as detector for non-uniformity measurement is recommended. 
Photodiodes with small detector area shall be used if a higher spatial resolution is required. 
For photodiodes special attention needs to be given to their linearity at typical irradiances of 
solar simulators. The detector should be of the same dimensions or smaller than the smallest 
sample size expected to be measured on the simulator. 
5.3.3 
Procedure 
5.3.3.1 
General 
The irradiance of the solar simulator may not be constant during the monitoring process for 
non-uniformity. Therefore, a second PV device should be used for monitoring the irradiance
to ensure that temporal fluctuations of irradiance between different measurement positions 
can be corrected to the same target value. This is to be placed typically at a fixed position 
outside the designated test area (stationary cell). It can also be placed inside the designated 
test area, in particular if it is of the same type and dimensions as the movable uniformity 
detector. Readings of the stationary cell and movable uniformity detector should be taken 
simultaneously. Irradiance correction from the stationary cell reading to the target irradiance 
level shall be performed for each measurement of the moving cell. 
If the solar simulator is used for I-V measurement of PV devices, the uniformity measurement 
readings should be taken during the part of the pulse, in which the I-V measurement is 
performed. The resulting non-uniformity pattern shall be referenced to the mean values 
recorded at each test location. If necessary, irradiance correction is to be performed. 
Care should be taken that the test arrangement for measuring spatial uniformity does not 
influence the results. 

Yüklə 0,72 Mb.

Dostları ilə paylaş:
1   ...   15   16   17   18   19   20   21   22   ...   62




Verilənlər bazası müəlliflik hüququ ilə müdafiə olunur ©genderi.org 2024
rəhbərliyinə müraciət

    Ana səhifə