IEC60904-9-2020

5.4.1 
Solar simulators for I-V measurement 
5.4.1.1 
General 
Both short term instability (STI) and long term instability (LTI) need to be evaluated and shall 
be reported. 
For the evaluation of STI, the I-V data acquisition system may be considered an integral part 
of the solar simulator. If a solar simulator does not include the data acquisition system, then 
the simulator manufacturer shall specify the corresponding data sampling time as related to 
the reported STI classification. 
The sampling rate of STI and LTI measurements shall be adequate to measure both 
parameters with sufficient accuracy. The sampling rate of the data acquisition system shall be 
reported. 
External measuring devices used to characterize the simulator (spectroradiometer, I-V data 
acquisition, etc.) may need precise synchronization with the simulator during its critical 
phases of operation. Common practice is that the external measuring device is equipped with 
an optical trigger to the start measurement process.
5.4.1.2 
Determination of STI 
STI relates to the data sampling time in the following measurement concepts: 
a) When there are three separate data input lines that simultaneously store values of 
irradiance, current and voltage, the temporal instability is Class A (if spectral classification 
is performed in the restricted wavelength range 400 nm to 1 100 nm) or Class A+ (if 
spectral classification is performed in the extended wavelength range 300 nm to 1 200 nm) 
for STI. 
NOTE The delay in simultaneous triggering of the three multiple channels is typically less than 10 ns. 
b) When each data set is taken sequentially (irradiance, current, voltage), determine the 
temporal instability as defined below (Figures 2 and 3) 
1) Determine the time for taking two successive data sets (irradiance, current, voltage) 
considering a possible delay time between measurements. 
2) STI is related to the largest irradiance change between successive data sets during the 
time of data acquisition (see Figure 2). 
3) Determine the STI using the data from step 2), formula (2) and Table 3. 
c) For measurement of cells or modules with no irradiance monitoring during I-V 
measurement the STI cannot be determined. Therefore, the LTI related to the time period 
between irradiance determinations shall be used for classification. 
Copyrighted material licensed to University of Toronto by Clarivate Analytics (US) LLC, subscriptions.techstreet.com, downloaded on 2020-09-26 08:51:03 +0000 by University of Toronto User.
No further reproduction or distribution is permitted.


– 20 – 
IEC 60904-9:2020 © IEC 2020 
d) 
For pulsed solar simulators used for I-V measurements but not including an I-V data 
acquisition system, the sections of the pulse to be utilized and the number of evenly 
spaced data points for achieving class A+, A, B, C of STI shall be stated in the 
classification report.