Orion versa star user Manual


A  Channel Specific Settings and



Yüklə 6,2 Kb.
Pdf görüntüsü
səhifə39/44
tarix05.01.2018
ölçüsü6,2 Kb.
#19795
1   ...   36   37   38   39   40   41   42   43   44

 

Channel Specific Settings and 
Features 
Within the Channel 1, Channel 2, Channel 3 and Channel 4 setup menus is the Mode menu, 
which can be used to customize measurement settings and parameters for the selected 
channel. This section provides detailed information on the options available within Mode menus 
when using the various V
ERSA 
S
TAR
 measurement modules. Detailed information on additional 
advanced meter features is also provided.  
 
General Mode Settings 
Sample ID 
If Manual is selected as the type of Sample ID, use the popup alphanumeric keypad screen to 
enter a sample ID up to ten characters long. The entered sample ID will be recorded with 
subsequent measurements and saved to the data log or exported to an external device, 
depending on the meter communication settings in the 
Instrument Settings
 setup menu.  
If Auto Incremental is selected as the type of Sample ID, use the popup numeric keypad screen 
to enter a numeric value up to ten characters long. The entered sample ID will be used as the 
starting value and the meter will automatically increase the value by one for each subsequent 
measurement. The sample IDs will be recorded with each measurement and saved to the data 
log or exported to an external device, depending on the meter communication settings in the 
Instrument Settings
 setup menu. 
Press the f2 (Sample ID) key in the measurement mode to open a shortcut to the Sample ID 
setup menu and enter a new sample ID. 
A P P E N D I X   A  
104  
|
  
Orion V
ERSA 
S
TAR
 Meter User Manual 
Thermo Scientific 


Appendix A  
|
  
Channel Specific Settings and Features 
Stability 
The Stability setting allows the operator to set when a measurement is recognized as stable by 
the meter. When the Smart option is selected, the measurement resolution in conjunction with 
an algorithm that evaluates measurement fluctuations are used to determine when the 
measurement value is stable. When the Fast, Medium or Slow option is selected, only the 
algorithm is used to determine when the measurement value is stable. Approximate mV per 
minute values, estimated based on ideal measurement conditions, are shown below. These 
values are theoretical examples only and actual mV per minute fluctuations will vary based on 
individual measurement conditions. 
Stability Setting 
Approximate mV per Minute Fluctuation 
Smart Stability 
 
0.1 resolution or 1 significant digit 
7.6 
0.01 resolution or 2 significant digits 
2.9 
0.001 resolution or 3 significant digit 
1.0 
0.0001 resolution  or 4 significant digit 
0.4 
Fast Stability 
7.6 
Medium Stability 
2.9 
Slow Stability 
1.0 
 
Averaging 
The Averaging setting allows the operator to select Automatic Smart averaging for faster 
measurement stability or Off for no measurement averaging once the stability criteria is met.  
Alarms 
The Limit Alarm 
allows the operator to s
et a high limit value and low limit value used to evaluate 
measurements. If a measurement goes above the high limit value or below the low limit value, the alarm 
is activated. For example, when the high limit is set to 8.500 pH and the low limit is set to 4.500 pH, the 
Limit Alarm is activated when a measurement of 8.501 pH or higher or a measurement of 4.499 pH or 
lower is observed in the measurement mode. 
The Cal Due Alarm 
allows the operator to s
et a calibration interval in hours. If a calibration is not 
performed within the specified calibration interval, the alarm is activated.  
The Set Point Alarm 
allows the operator to s
et a base value and offset value used to evaluate 
measurements. If the measurement goes above or below the base value by the entered offset value, the 
alarm is activated. For example, when the base value is set to 7.000 pH and the offset value is set to 
2.000 pH, the Set Point Alarm is activated when a measurement of 9.000 pH or higher or a measurement 
of 5.000 pH or lower is observed in the measurement mode.  
Ø
 
Use the Buzzer – Alarm setting in the 
Instrument Settings
 setup menu to enable an audible 
beep whenever an alarm is activated.
 
Thermo Scientific 
Orion V
ERSA 
S
TAR
 Meter User Manual  
|
  
105
 


Appendix A  
|
  
Channel Specific Settings and Features 
pH and ISE Mode Settings 
Buffer Group and Automatic Buffer Recognition 
V
ERSA 
S
TAR
 meters with pH, pH/ISE and pH/LogR modules are capable of automatically 
recognizing pH 1.68, 4.01, 7.00, 10.01 and 12.46 buffers; pH 1.68, 4.01, 6.86, and 9.18 buffers; 
or up to six custom buffers during a pH calibration, depending on the pH buffer set that is 
selected in the setup menu. During a calibration, the meter uses the selected buffer set and the 
raw mV reading of the pH electrode in the buffer to recognize and display the buffer value at the 
measured temperature. The raw mV reading of the pH electrode in the buffer must be within 
one pH unit (± 59 mV) of the buffer’s theoretical mV value for the meter to recognize the buffer. 
USA Buffer Set 
 
DIN Buffer Set 
 
Custom Buffer Set 
pH 
Buffer  mV range 
  pH 
Buffer  mV range 
  Entered pH 
Buffer Value  
mV range (based on 
entered pH value) 
1.68 
+256 to +374 mV    1.68 
+256 to +374 mV    1
st
 buffer value  Theoretical ± 59 mV 
4.01 
+119 to +236 mV    4.01 
+119 to +236 mV    2
nd
 buffer value  Theoretical ± 59 mV 
7.00 
-59 to +59 mV 
  6.86 
-51 to +67 mV 
  3
rd
 buffer value  Theoretical ± 59 mV 
10.01 
-236 to -119 mV    9.18 
-70 to -188 mV 
  4
th
 buffer value  Theoretical ± 59 mV 
12.46 
-264 to -382 mV     
 
  5
th
 buffer value  Theoretical ± 59 mV 
 
 
   
 
  6
th
 buffer value  Theoretical ± 59 mV 
 
Testing a pH Electrode for Automatic Buffer Recognition 
Use the following procedure to verify that the raw mV reading of the pH electrode is within one 
pH unit (± 59 mV) from the theoretical mV reading of the pH buffer, and therefore verify that the 
pH electrode in use is capable of performing automatic buffer recognition. 
1.
 
Prepare the pH electrode according to the electrode user manual. Set the meter 
measurement mode to mV. 
2.
 
Rinse the pH electrode with distilled water, blot it dry with a lint-free tissue and place the pH 
electrode into a pH 4.01 buffer at approximately 25 °C. 
3.
 
Wait for the measurement to stabilize and record the mV value of the pH 4.01 buffer when 
the measurement is stable.  
4.
 
Remove the pH electrode from the pH 4.01 buffer. 
5.
 
Rinse the pH electrode with distilled water, blot it dry with a lint-free tissue and place the pH 
electrode into a pH 7.00 buffer at approximately 25 °C. 
6.
 
Wait for the measurement to stabilize and record the mV value of the pH 7.00 buffer when 
the measurement is stable. 
7.
 
The mV reading of the pH electrode in pH 4 buffer should be +119 to +236 mV and in pH 7 
buffer should be -59 to +59 mV. If the mV readings are in the correct ranges, the pH 
electrode is capable of performing automatic buffer recognition. If the mV readings are not 
in the correct ranges, manually enter the pH buffer values during a pH calibration. 
106  
|
  
Orion V
ERSA 
S
TAR
 Meter User Manual 
Thermo Scientific 


Yüklə 6,2 Kb.

Dostları ilə paylaş:
1   ...   36   37   38   39   40   41   42   43   44




Verilənlər bazası müəlliflik hüququ ilə müdafiə olunur ©genderi.org 2024
rəhbərliyinə müraciət

    Ana səhifə