Orion versa star user Manual


A   |    Channel Specific Settings and Features



Yüklə 6,2 Kb.
Pdf görüntüsü
səhifə43/44
tarix05.01.2018
ölçüsü6,2 Kb.
#19795
1   ...   36   37   38   39   40   41   42   43   44

Appendix A  
|
  
Channel Specific Settings and Features 
Electrode Diagnostics Log 
Electrode 
Model 
Electrode 
Serial # 
Date  Temp.  Resistance  Change in 
Resistance 
Notes 
8102BNWP  
(example)  PV1-27726  1/6/14  24.5°C  340 MΩ 
--- 
Initial Check 
8102BNWP  
(example)  PV1-27726  1/13/14  24.0°C  20 MΩ 
320 MΩ 
Large resistance decrease, 
indicates electrode failure – 
crack in pH bulb observed 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
Thermo Scientific 
Orion V
ERSA 
S
TAR
 Meter User Manual  
|
  
115
 


Appendix A  
|
  
Channel Specific Settings and Features 
Conductivity, TDS, Salinity and Resistivity 
Mode Settings 
Cell Constant (K) and Automatic Conductivity Calibration 
V
ERSA 
S
TAR
 meters with a conductivity module are capable of automatically recognizing 
Thermo Scientific Orion 100 μS/cm conductivity standard, 1413 μS/cm conductivity standard 
and 12.9 mS/cm conductivity standard when the nominal cell constant (K) value of the 
conductivity probe is entered in the setup menu. 
Catalog Number 
Description 
011008 
Orion 100 μS/cm conductivity standard, 5 x 60 mL 
011007 
Orion 1413 μS/cm conductivity standard, 5 x 60 mL 
01100710 
Orion 1413 μS/cm conductivity standard, 10 individual use pouches 
011006 
Orion 12.9 mS/cm conductivity standard, 5 x 60 mL 
01100610 
Orion 12.9 mS/cm conductivity standard, 10 individual use pouches 
 
The nominal cell constant (K) values for Thermo Scientific Orion conductivity probes compatible 
with V
ERSA 
S
TAR
 meters with a conductivity module are listed below. 
Catalog 
Number 
Description 
Measurement 
Range 
Nominal Cell 
Constant 
013005MD 
Orion DuraProbe 4-cell conductivity probe with built-
in temperature, 1.5 meter cable 
1 μS/cm to 
200 mS/cm 
K = 0.475 cm
-1
 
013010MD 
Orion DuraProbe 4-cell conductivity probe with built-
in temperature, 3 meter cable 
1 μS/cm to 
200 mS/cm 
K = 0.475 cm
-1
 
013020MD 
Orion DuraProbe 4-cell conductivity probe with built-
in temperature, 6 meter cable 
1 μS/cm to 
200 mS/cm 
K = 0.475 cm
-1
 
013025MD 
Orion DuraProbe 4-cell conductivity probe with built-
in temperature, 10 meter cable 
1 μS/cm to 
200 mS/cm 
K = 0.475 cm
-1
 
013605MD 
Orion DuraProbe 4-cell conductivity probe with built-
in temperature, 1.5 meter cable 
10 μS/cm to 
200 mS/cm 
K = 0.55 cm
-1
 
013610MD 
Orion DuraProbe 4-cell conductivity probe with built-
in temperature, 3 meter cable 
10 μS/cm to 
200 mS/cm 
K = 0.55 cm
-1
 
013016MD 
Orion ultra pure water 2-cell conductivity probe with 
built-in temperature, 1.5 meter cable 
0.01 μS/cm to 
300 μS/cm 
K = 0.1 cm
-1
 
011510MD 
Orion 2-cell conductivity probe with built-in 
temperature, 3 meter cable 
10 μS/cm to 
200 mS/cm 
K = 1.0 cm
-1
 
011050MD 
Orion 2-cell conductivity probe with built-in 
temperature, 1.5 meter cable 
1 μS/cm to  
20 mS/cm 
K = 1.0 cm
-1
 
018020MD 
Orion high range 2-cell conductivity probe, 1.5 meter 
cable 
10 μS/cm to 
2000 mS/cm 
K = 10 cm
-1
 
 
116  
|
  
Orion V
ERSA 
S
TAR
 Meter User Manual 
Thermo Scientific 


Appendix A  
|
  
Channel Specific Settings and Features 
Orion Conductivity Standard Values from 0 °C to 50 °C 
Temperature 
(°C) 
Orion 
111.9mS/cm 
Conductivity 
Standard 
Cat. No. 
011005, 
01100510 
(mS/cm) 
Orion 
12.9mS/cm 
Conductivity 
Standard 
Cat. No. 
011006, 
01100610 
(mS/cm) 
Orion 
1413μS/cm 
Conductivity 
Standard 
Cat. No. 
011007, 
01100710 
(μS/cm) 
Orion 
147μS/cm 
Conductivity 
Standard 
Cat. No. 
01100910 
(μS/cm) 
Orion 
100μS/cm 
Conductivity 
Standard 
Cat. No. 
011008 
(μS/cm) 

65.10 
7.135 
776 
81 
54 

66.84 
7.344 
799 
83 
56 

68.59 
7.555 
822 
86 
58 

70.35 
7.768 
846 
88 
59 

72.12 
7.983 
870 
91 
61 

73.91 
8.200 
894 
93 
63 

75.70 
8.418 
918 
96 
64 

77.50 
8.638 
943 
98 
66 

79.32 
8.860 
968 
101 
68 

81.15 
9.084 
992 
103 
70 
10 
82.98 
9.309 
1017 
106 
72 
11 
84.83 
9.535 
1043 
108 
73 
12 
86.69 
9.763 
1068 
111 
75 
13 
88.56 
9.993 
1094 
114 
77 
14 
90.45 
10.22 
1119 
116 
79 
15 
92.34 
10.46 
1145 
119 
81 
16 
94.24 
10.69 
1171 
122 
83 
17 
96.15 
10.93 
1198 
125 
85 
18 
98.08 
11.16 
1224 
127 
87 
19 
100.0 
11.40 
1251 
130 
88 
20 
102.0 
11.64 
1277 
133 
90 
21 
103.9 
11.88 
1304 
136 
92 
22 
105.9 
12.12 
1331 
138 
94 
23 
107.9 
12.36 
1358 
141 
96 
24 
109.9 
12.61 
1386 
144 
98 
25 
111.9 
12.85 
1413 
147 
100 
Thermo Scientific 
Orion V
ERSA 
S
TAR
 Meter User Manual  
|
  