Annex C – (normative) – Circuit for measuring leakage currents
A suitable circuit for measuring leakage currents is shown in Figure C.1.
The circuit comprises a rectifier with germanium diodes D, a moving-coil meter M, resistors and a
capacitor C for adjusting the characteristics of the circuit, and a

make-before-break

switch S for
adjusting the current range of the instrument.
Germanium diodes are used since they have a lower voltage drop than other types of diodes, resulting in
a more linear scale, preference being given to gold bonded types. The rating of the diodes has to be
chosen to suit the desired maximum range of the complete instrument. However, this range must not
exceed 25 mA, since diodes suitable for higher currents have a high voltage drop.
The capacitor may be constructed by selecting capacitors having preferred values and using a
series/parallel arrangement.
It is recommended that the switch is arranged so that it automatically returns to the position giving the
highest current range, in order to prevent inadvertent damage to the instrument.
The measuring circuit has a total resistance of 1 750
Ω ±
250

, and is shunted by a capacitor so that the
time constant of the circuit is 225 µs
±
15 µs. The circuit may be protected against overcurrents, but the
method chosen is not to affect the characteristics of the circuit.
The most sensitive range of the complete instrument is not to exceed 1,0 mA, higher ranges being
obtained by shunting the coil of the meter by non-inductive resistors R
s
and simultaneously adjusting the
series resistors R
v
so as to maintain the total resistance R
1
+ R
v
+ R
m
of the circuit at the value specified.
The resistance R
m
is calculated from the voltage drop measured across the rectifier arrangement at 0,5
mA, the resistance R
v
being then adjusted to give the total resistance of the circuit for each range.
The basic calibration points, at a sinusoidal frequency of 50 Hz or 60 Hz, are 0,25 mA, 0,5 mA and 0,75
mA.