Chapter 6  
|
  
RDO/DO Measurement Module 
RDO and DO Mode Menu 
  
 
Main Menu 
Settings 
Details 
Electrode SI No 
---- 
Use the alphanumeric keypad to enter the serial 
number or clear all to shown no value (----) 
Sample ID 
Off, Manual, Auto Increment 
Use the popup keypad screen to enter an 
alphanumeric manual ID or a numeric value as 
the starting value for auto incremental ID 
Measurement Unit 
mg/L, 
  % 
Set the displayed measurement units 
Resolution 
0.1, 0.01 
1,  0.1 
Set the resolution of the measurement value 
Stability 
Smart, Fast, Medium, Slow 
Set when a measurement is recognized as stable 
Averaging 
Off,  Automatic Smart 
Set averaging for faster measurement stability 
Read Type 
Auto-Read, Timed,  
Single-Shot, Continuous 
For Timed and Single-Shot, set the time in hours, 
minutes, seconds using the arrow keys 
Pressure Comp 
Auto,   Manual 
775.0  
Set the barometric pressure compensation 
source as Auto to use the built-in barometer or 
Manual to enter a value. Select the Auto option 
to calibrate the barometer pressure reading 
Salinity Correction 
Auto,   Manual 
Set the salinity correction source as Auto to use 
a conductivity module and probe to measure the 
sample salinity or Manual to enter a value 
Alarm 
Limit Alarm 
Off,  On 
Set high and low limit values to activate alarm  
Cal Due Alarm  Off,  On 
Set a calibration interval (hours) to activate alarm 
Set Point Alarm  Off,  On 
Set a base and offset value to activate alarm 
Detailed information on these features is available in the 
Appendix
 
section of this user manual. 
RDO Electrode Information 
The serial number, remaining cap life and date of manufacture for each RDO optical cap can be 
viewed in the RDO Electrode Information menu when the RDO probe is connected to the meter 
and recognized in the measurement mode. 
1.
 
Verify that the RDO probe is connected to the meter. 
2.
 
In the measurement mode, press the setup key. 
3.
 
Press the  or  key to highlight Channel # (channel 
with the RDO probe) and press the f3 (Select) key. 
4.
 
Press the ◄ or ► key to highlight RDO Electrode 
Information and press the f3 (Select) key. 
66  
|
  
Orion V
ERSA 
S
TAR
 Meter User Manual 
Thermo Scientific 


Chapter 6  
|
  
RDO/DO Measurement Module 
Dissolved Oxygen Calibration 
Note: A polarographic DO probe must be polarized prior to use. The probe is continuously 
polarized when connected to the meter. If the polarographic DO probe is new, has been 
serviced or has not been connected to the meter: connect the probe to the meter, power on the 
meter and wait 30 minutes for the probe to polarize. 
The V
ERSA 
S
TAR
 meter with RDO/DO module offers the following calibration options:  
·
 
Water Saturated Air – This is the simplest and most accurate method and uses the 
calibration sleeve included with most DO probes. For best accuracy, the calibration 
temperature should match the expected sample temperature. Moisten the sponge in the 
calibration sleeve with distilled water and insert the probe into the sleeve. A BOD bottle can 
be used with just enough distilled water to cover the bottom without touching the probe.  
·
 
Air Saturated Water – This method uses water that is 100% saturated with air. Bubble air 
into a water sample for an extended period, preferably overnight. 
·
 
Manual (Winkler) – This method uses a water sample with a known concentration of 
dissolved oxygen and is typically used to calibrate the DO probe to the value achieved by a 
Winkler titration. Due to possible titration errors, this method is inherently less accurate. 
·
 
Set Zero – This method uses an oxygen free solution to add a zero point to an existing 
Water Saturated Air or Air Saturated Water calibration. This calibration is generally 
recommended when measurements are below 10% saturation or 1 mg/L.  
Prepare the DO probe according to the instructions in the probe user manual. Connect the 
probe and any other electrodes to be used to the module inputs on the selected channel.  
In the measurement mode, press the channel key until the channel with the RDO/DO module 
with DO probe connected is shown.  
Water Saturated Air Calibration 
1.
 
Press the f1 (Cal) key to start the calibration.  
a.
 
If more than one channel is displayed in the measurement mode, press the  or  
key to highlight the desired channel and press the f3 (Select) key. 
2.
 
Press the  or  key to highlight Water Saturated Air and press the f3 (Select) key. 
3.
 
Prepare the DO probe and calibration sleeve and allow them to reach equilibrium.   
4.
 
When the DO probe and calibration sleeve are ready, press the f3 (Start) key. 
5.
 
Wait for the dissolved oxygen value on the meter to stabilize and press the f2 (Accept) key. 
a.
 
With a polarographic DO probe, 102.3 % will be displayed when the reading stabilizes. 
b.
 
With an RDO optical DO probe, 100.0 % will be displayed when the reading stabilizes. 
6.
 
Press the f3 (Cal Done) key to save and end the calibration.  
7.
 
The meter will display the calibration summary and export the data to the calibration log. 
Press the measure (esc) key to proceed to the measurement mode. 
Thermo Scientific 
Orion V
ERSA 
S
TAR
 Meter User Manual  
|
  
67
 