Chapter 5  
|
  
Conductivity Measurement Module 
Resistivity Mode Menu 
Main Menu 
Settings 
Details 
Read Type 
Auto-Read, Timed,  
Single-Shot, Continuous 
For Timed and Single-Shot, set the time in hours, 
minutes, seconds using the arrow keys 
Cell K 
0.475 
Enter the nominal cell constant (K) value of the 
conductivity probe; value is used when 
performing an automatic conductivity calibration 
Ref. Temperature 
5°C, 10°C, 15°C, 20°C, 25°C 
Set the reference temperature that all displayed 
resistivity measurements will be reported to using 
the selected temperature compensation 
Temp. 
Compensation 
Linear, nLFn, nLFu, EP, Off 
Set the temperature compensation type used to 
calculate and report resistivity measurements at 
the selected reference temperature 
Temperature Coeff.  2.10 
Enter the temperature coefficient value for linear 
temperature compensation 
Stability 
Smart, Fast, Medium, Slow 
Set when a measurement is recognized as stable 
Averaging 
Off,  Automatic Smart 
Set averaging for faster measurement stability 
Alarm 
Limit Alarm 
Off,  On 
Set high and low limit values to activate alarm  
Set Point Alarm  Off,  On 
Set a base and offset value to activate alarm 
Cell Type 
Standard,  USP 
Select Standard for most conductivity probes or 
USP for ultra pure water 2-cell conductivity probe 
 
Detailed information on these features is available in the 
Appendix
 
section of this user manual. 
 
 
60  
|
  
Orion V
ERSA 
S
TAR
 Meter User Manual 
Thermo Scientific 


Chapter 5  
|
  
Conductivity Measurement Module 
Conductivity Calibration 
One to six conductivity standards can be used for calibration (up to three for an automatic 
conductivity calibration, up to six for a direct conductivity calibration and one for a manual 
conductivity calibration). Always use fresh standards and select standards that are near the 
expected sample conductivity.  
Automatic or Direct Conductivity Calibration  
Prepare the conductivity probe according to the instructions in the probe user manual. Connect 
the probe and any other electrodes to be used to the module inputs on the selected channel. 
In the measurement mode, press the channel key until the channel with the conductivity 
module with conductivity probe connected is shown. If needed, press the mode key until Cond 
Measurement is shown as the measurement mode.  
For an automatic calibration, use the 
Cell Constant Entry
 procedure or 
Conductivity Mode Menu
  
to enter the nominal cell constant of the conductivity probe before the calibration. Use Thermo 
Scientific

 Orion

 100 μS/cm, 1413 μS/cm and/or 12.9 mS/cm conductivity standards. 
1.
 
Press the f1 (Cal) key to start the calibration.  
a.
 
If more than one channel is displayed in the measurement mode, press the  or  
key to highlight the desired channel and press the f3 (Select) key. 
2.
 
Rinse the conductivity probe and any other electrodes in use with distilled water, blot dry 
with a lint-free tissue and place into the standard. 
3.
 
When the conductivity probe and standard are ready, press the f3 (Start) key. 
a.
 
If using a stirrer probe, the stirrer probe will start stirring when the f3 (Start) key is 
pressed and stop stirring when the reading stabilizes. 
4.
 
Wait for the conductivity value on the meter to stabilize and perform one of the following actions: 
a.
 
Press the f2 (Accept) key to accept the displayed value. 
or 
b.
 
Press the f3 (Edit) key, press the f3 (Clear) key and use the numeric keypad to enter 
the conductivity value of the standard at the measured temperature. Press the  
f2 (Accept) key to confirm the entered value. 
5.
 
Press the f2 (Next) key to proceed to the next standard and repeat steps 2 through 4 or 
press the f3 (Cal Done) key to save and end the calibration. 
6.
 
The meter will display the calibration summary and export the data to the calibration log. 
Press the measure (esc) key to proceed to the measurement mode. 
 
 
Thermo Scientific 
Orion V
ERSA 
S
TAR
 Meter User Manual  
|
  
61
 


Chapter 5  
|
  
Conductivity Measurement Module 
Conductivity Calibration Editing 
Once the f3 (Cal Done) key is pressed and the calibration is saved, press the f2 (Cal Edit) key 
to edit the calibration data and fix individual points without a full recalibration. In the calibration 
edit mode, press the  or  key to highlight a calibration point and press the f2 (Delete) key 
to delete the point or press the f3 (Re-Measure) key to re-measure the point. Follow the 
onscreen text prompts and instructions to edit the selected calibration point. 
Orion Conductivity Standard Values from 15 °C to 30 °C   
Refer to the 
Appendix
 section of this user manual for additional values from 0 to 50°C.  
Temperature 
(°C) 
Orion 111.9mS/cm 
Standard 
Cat. No. 011005, 
01100510 
(mS/cm) 
Orion 12.9mS/cm 
Standard 
Cat. No. 011006, 
01100610 
(mS/cm) 
Orion 1413μS/cm 
Standard 
Cat. No. 011007, 
01100710 
(μS/cm) 
Orion 147μS/cm 
Standard 
Cat. No. 
01100910 
(μS/cm) 
Orion 100μS/cm 
Standard 
Cat. No. 011008 
(μS/cm) 
15 
92.34 
10.46 
1145 
119 
81 
16 
94.24 
10.69 
1171 
122 
83 
17 
96.15 
10.93 
1198 
125 
85 
18 
98.08 
11.16 
1224 
127 
87 
19 
100.0 
11.40 
1251 
130 
88 
20 
102.0 
11.64 
1277 
133 
90 
21 
103.9 
11.88 
1304 
136 
92 
22 
105.9 
12.12 
1331 
138 
94 
23 
107.9 
12.36 
1358 
141 
96 
24 
109.9 
12.61 
1386 
144 
98 
25 
111.9 
12.85 
1413 
147 
100 
26 
113.9 
13.10 
1441 
150 
102 
27 
115.9 
13.35 
1468 
153 
104 
28 
117.9 
13.59 
1496 
156 
106 
29 
120.0 
13.84 
1524 
159 
108 
30 
122.0 
14.09 
1552 
161 
110 
 
Cell Constant Entry (Manual Conductivity Calibration) 
1.
 
Press the f1 (Cal) key to start the calibration.  
a.
 
If more than one channel is displayed in the measurement mode, press the  or  
key to highlight the desired channel and press the f3 (Select) key. 
2.
 
Press the f2 (Cell K) key and wait for the cell constant value on the meter to stabilize.  
3.
 
Press the f2 (Accept) key to accept the displayed value or press the f3 (Edit) key, press the 
f3 (Clear) key, use the numeric keypad to enter the cell constant value of the conductivity 
probe and press the f2 (Accept) key to confirm the entered value. 
4.
 
The meter will save the entered cell constant value as the nominal cell constant of the 
conductivity probe and return to the main calibration display. Press the measure (esc) key 
to proceed to the measurement mode or perform the 
Automatic or Direct Conductivity 
Calibration 
procedure. 
62  
|
  
Orion V
ERSA 
S
TAR
 Meter User Manual 