Orion versa star user Manual


   |    pH, pH/ISE and pH/LogR Measurement Modules



Yüklə 6,2 Kb.
Pdf görüntüsü
səhifə20/44
tarix05.01.2018
ölçüsü6,2 Kb.
#19795
1   ...   16   17   18   19   20   21   22   23   ...   44

Chapter 4  
|
  
pH, pH/ISE and pH/LogR Measurement Modules 
12.
 
Enter the reaction ratio using the numeric keypad and press the f2 (Accept) key. 
13.
 
Rinse the ISE and any other electrodes in use with distilled water or rinse solution required 
for that electrode. Blot the electrode dry with a lint-free tissue or shake the electrode, place 
into the prepared sample and press the f2 (Next) key.  
a.
 
Follow all electrode rinsing and drying instructions or exceptions in the ISE user 
manual. Do not wipe or rub the sensing element of the electrode. 
14.
 
Wait for the mV value on the meter to stabilize and press the f2 (Accept) key. 
15.
 
Enter the standard volume using the numeric keypad and press the f2 (Accept) key. 
16.
 
Add the standard to the sample and press the f2 (Yes) key. 
17.
 
Wait for the mV value on the meter to stabilize and press the f2 (Accept) key. 
18.
 
The meter will display the calculated sample concentration and summary of the incremental 
technique performed. Press the f1 (Esc) key to proceed to the measurement mode or press 
the f2 (Next) key to perform another analysis. 
Double Known Subtraction  
Double Known Subtraction Recommendations 
Ensure that the subtraction ability of the chosen standard concentration and volume added does not 
exceed the concentration of the sample. 
The sample volume should be 100 mL. 
The concentration of the standard in Molarity or Normality should be 50 times the expected sample 
concentration when the reaction ratio is 1. 
Calculate the volume of standard required to halve the initial sample concentration and divide by six.  
The volume of the first addition should be 5/6 the total standard mL and the volume of the second 
addition should be 1/6 the total standard mL.  
If available, consult a printout from a previous analysis of a similar sample. 
1.
 
Prepare the ion selective electrode (ISE) according to the instructions in the electrode user 
manual. Prepare all required solutions (ionic strength adjusters, standards, etc) and review 
any special requirements outlined in the electrode user manual. Connect the ISE, and any 
other electrodes to be used, to the module inputs on the selected channel.  
2.
 
In the measurement mode, press the channel key until the channel with the pH/ISE module 
with ISE connected is shown. If needed, press the mode key until ISE Measurement is 
shown as the measurement mode and use the 
ISE Mode Menu 
to set the Electrode Type.  
3.
 
In the measurement mode, press the setup key.  
4.
 
Press the ◄ or ► key to highlight the channel setup menu that the ISE is connected with 
(Channel 1, Channel 2, Channel 3 or Channel 4) and press the f3 (Select) key.  
5.
 
Press the ◄ or ►key to highlight Incremental Techniques and press the f3 (Select) key.  
6.
 
Press the or  key to highlight Double Known Subtraction and press the f3 (Select) key.  
7.
 
Enter the sample volume using the numeric keypad and press the f2 (Accept) key. 
8.
 
Enter the total volume (sample volume plus added ISA volume) using the numeric keypad 
and press the f2 (Accept) key. 
52  
|
  
Orion V
ERSA 
S
TAR
 Meter User Manual 
Thermo Scientific 


Chapter 4  
|
  
pH, pH/ISE and pH/LogR Measurement Modules 
9.
 
Enter the standard concentration in Molarity or Normality using the numeric keypad and 
press the f2 (Accept) key. 
10.
 
Enter the reaction ratio using the numeric keypad and press the f2 (Accept) key. 
11.
 
Rinse the ISE and any other electrodes in use with distilled water or rinse solution required 
for that electrode. Blot the electrode dry with a lint-free tissue or shake the electrode, place 
into the prepared sample and press the f2 (Next) key.  
a.
 
Follow all electrode rinsing and drying instructions or exceptions in the ISE user 
manual. Do not wipe or rub the sensing element of the electrode. 
12.
 
Wait for the mV value on the meter to stabilize and press the f2 (Accept) key. 
13.
 
Enter the first standard volume using the numeric keypad and press the f2 (Accept) key. 
14.
 
Add the first specified volume of the standard to the sample and press the f2 (Yes) key. 
15.
 
Wait for the mV value on the meter to stabilize and press the f2 (Accept) key. 
16.
 
Enter the second standard volume using the numeric keypad and press the f2 (Accept) 
key. 
17.
 
Add the second specified volume of the standard to the sample and press the f2 (Yes) key. 
18.
 
Wait for the mV value on the meter to stabilize and press the f2 (Accept) key. 
19.
 
The meter will display the calculated sample concentration and summary of the incremental 
technique performed. Press the f1 (Esc) key to proceed to the measurement mode or press 
the f2 (Next) key to perform another analysis. 
Single Analate Addition  
In single analate addition, the first measurement is in a known standard solution containing the 
same type of analate of interest as the sample. The second measurement follows the addition 
of the unknown sample. The concentration of the unknown sample is then calculated. Refer to 
the following table to choose the correct standard concentration and volume of addition. 
Amount of analate to be added per 100 mL 
of standard 
Standard concentration compared to expected 
sample concentration 
1 mL 
100 times more concentrated 
5 mL 
20 times more concentrated 
10 mL 
10 times more concentrated 
1.
 
Prepare the ion selective electrode (ISE) according to the instructions in the electrode user 
manual. Prepare all required solutions (ionic strength adjusters, standards, etc) and review 
any special requirements outlined in the electrode user manual. Connect the ISE, and any 
other electrodes to be used, to the module inputs on the selected channel.  
2.
 
Determine the ISE slope as directed in the electrode user manual. 
3.
 
In the measurement mode, press the channel key until the channel with the pH/ISE module 
with ISE connected is shown. If needed, press the mode key until ISE Measurement is 
shown as the measurement mode and use the 
ISE Mode Menu 
to set the Electrode Type.  
4.
 
In the measurement mode, press the setup key.  
Thermo Scientific 
Orion V
ERSA 
S
TAR
 Meter User Manual  
|
  
53
 


Yüklə 6,2 Kb.

Dostları ilə paylaş:
1   ...   16   17   18   19   20   21   22   23   ...   44




Verilənlər bazası müəlliflik hüququ ilə müdafiə olunur ©genderi.org 2024
rəhbərliyinə müraciət

    Ana səhifə