Chapter 4  
|
  
pH, pH/ISE and pH/LogR Measurement Modules 
LogR Temperature Calibration (pH/LogR Module Only) 
The V
ERSA 
S
TAR
 meter with pH/LogR measurement module offers the option for temperature 
calibration of a pH electrode, allowing most standard glass-bulb pH electrodes to be used for 
automatic temperature compensation. The type and quality of pH electrode used to perform 
LogR temperature measurements will contribute to the overall temperature accuracy of the 
system. To maximize LogR performance and accuracy, a Thermo Scientific

 Orion
™ 
ROSS 
Ultra

 or Thermo Scientific

 Orion
™ 
ROSS

 pH electrode is recommended. 
LogR technology is based on using the electrical resistance of a pH electrode’s pH-sensing 
glass-bulb as the temperature source. The logarithm of the resistance of the bulb varies almost 
linearly with the reciprocal of the absolute temperature. Almost all standard glass-bulb pH 
electrodes exhibit a similar decrease in resistance with increasing temperature.  
LogR Temperature Calibration Recommendations 
·
 
When using the meter for the first time or restarting the meter in the LogR temperature 
mode, a LogR temperature calibration must be performed. A LogR temperature calibration 
should be performed at least once per week to ensure accurate temperature compensation 
when using the LogR temperature mode. Recalibrate the LogR temperature if the meter is 
disconnected from the power supply or the pH electrode is disconnected from the meter.  
·
 
Up to a three point LogR temperature calibration can be performed. It is recommended that 
at least two calibration points be used to meet optimum system accuracy.  
·
 
Choose a temperature calibration range that is 5 °C to 20 °C apart, depending on the 
expected sample temperature range, and evenly bracket the expected temperature range. 
For example, if sample temperatures are around 25 °C, calibrate at 15 °C and 35 °C. For 
sample temperatures that differ by more than 20 °C, a three point LogR temperature 
calibration is recommended. For many samples, a one point LogR calibration near the 
expected sample temperature will yield adequate results.  
·
 
A constant temperature bath or a hot plate is recommended to hold the temperature of the 
calibrating solution.  
·
 
For best results during temperature calibration, it is recommended that a pH buffer or tap 
water be used as the temperature calibrating solution. Do not use distilled water because 
the conductivity is too low to obtain a good temperature calibration.  
·
 
LogR temperature calibration can be performed using a NIST-traceable thermometer or 
ATC probe. The most accurate method is to use a NIST-traceable thermometer. It is 
recommended that two NIST-traceable thermometers be used to measure and verify the 
temperature value. The fastest method is to use an ATC probe. If an ATC probe is 
connected with the pH/LogR module when a LogR temperature calibration is performed, 
the meter will import the temperature value from the ATC probe as the reference 
temperature during the calibration. The temperature source used to perform the LogR 
temperature calibration will contribute to the overall temperature accuracy of the system.  
Detailed information on this feature is available in the 
Appendix
 section of this user manual. 
46  
|
  
Orion V
ERSA 
S
TAR
 Meter User Manual 
Thermo Scientific 


Chapter 4  
|
  
pH, pH/ISE and pH/LogR Measurement Modules 
LogR Temperature Calibration Procedure 
Prepare the glass-bulb pH electrode according to the instructions in the electrode user manual. 
Connect the pH electrode to the pH/LogR module input on the selected channel. If an ATC 
probe will be used as the temperature source, prepare and connect the ATC probe to the 
pH/LogR module input on the same channel. Use a constant temperature bath or a hot plate to 
attain the desired temperatures of the calibration solutions.  
1.
 
In the measurement mode, press the setup key.  
2.
 
Press the ◄ or ► key to highlight Channel 1, Channel 2, Channel 3 or Channel 4 and 
press the f3 (Select) key.  
3.
 
Press the ◄ or ►key to highlight Temperature and press the f3 (Select) key.  
4.
 
Press the  or  key to highlight Temperature Input and press the  or  key to 
highlight the pH electrode that will be used for the LogR mode (i.e. Ch2-LogR).  
Ø
 
This enables the LogR mode for the selected channel and allows a LogR temperature 
calibration to be performed. 
5.
 
Press the  or  key to highlight Temperature Cal and press the f3 (Select) key. 
6.
 
Place the pH electrode and NIST-traceable thermometers or ATC probe into a calibration 
solution (pH buffer or tap water) with a known, stable temperature. Moderately stir the 
solution using an 
Orion Star stirrer probe
 (Catalog Number 096019) or stir plate with stir bar. 
7.
 
Press the f3 (Start) key and wait for the temperature value on the meter to stabilize.  
a.
 
If using a stirrer probe, the stirrer probe will start stirring when the f3 (Start) key is 
pressed and stop stirring when the reading stabilizes. 
8.
 
Once the temperature value is stable, perform one of the following actions: 
a.
 
Press the f2 (Accept) key to accept the displayed value. 
or 
b.
 
Press the f3 (Edit) key, press the f3 (Clear) key, use the numeric keypad to enter the 
temperature value and press the f2 (Accept) key to confirm the entered value. 
9.
 
Press the f2 (Next) key to proceed to the next calibration point and repeat steps 6 through 8 
or press the f3 (Cal Done) key to save and end the calibration. 
 
 
 
 
 
 
Thermo Scientific 
Orion V
ERSA 
S
TAR
 Meter User Manual  
|
  
47
 