Chapter 4  
|
  
pH, pH/ISE and pH/LogR Measurement Modules 
Example pH Calibration Displays 
 
 
 
Example 1: When a pH calibration is started
onscreen text prompts provide instructions for 
how to perform each calibration step. 
  Example 2: A flashing Stabilizing icon is shown 
on the display as the pH buffer reading 
fluctuates and becomes stable. 
 
 
 
Example 3: When the pH buffer reading is stable, 
onscreen text prompts and function keys update 
with the appropriate instructions and options.   
  Example 4: While each pH buffer is being read
the display shows the previous calibration points 
with corresponding mV and temperature values.  
 
 
 
Example 5: Once the pH calibration is complete, 
the calibration summary is shown on the display. 
  Example 6: The calibration summary can also 
be viewed as a graph. 
 
 
 
Example 7: The calibration edit function can be 
used to edit calibration data and fix individual 
points without performing a full recalibration. 
  Example 8: The calibration summaries for the 
ten most recent calibrations per parameter can 
be viewed and printed in the calibration log. 
Thermo Scientific 
Orion V
ERSA 
S
TAR
 Meter User Manual  
|
  
41
 


Chapter 4  
|
  
pH, pH/ISE and pH/LogR Measurement Modules 
Relative mV Calibration 
One standard can be used for relative mV calibration. Always use fresh standard for calibration.  
Prepare the electrode according to the instructions in the electrode user manual. Connect the 
electrode and any other electrodes to be used to the module inputs on the selected channel. 
In the measurement mode, press the channel key until the channel with the pH, pH/ISE or 
pH/LogR module with electrode connected is shown. If needed, press the mode key until RmV 
Measurement is shown as the measurement mode.  
1.
 
Press the f1 (Cal) key to start the calibration.  
a.
 
If more than one channel is displayed in the measurement mode, press the  or  
key to highlight the desired channel and press the f3 (Select) key. 
2.
 
Rinse the electrode and any other electrodes in use with distilled water, blot dry with a lint-
free tissue and place into the standard. 
3.
 
When the electrode and standard are ready, press the f3 (Start) key.  
a.
 
If using a stirrer probe, the stirrer probe will start stirring when the f3 (Start) key is 
pressed and stop stirring when the reading stabilizes. 
4.
 
Wait for the mV value on the meter to stabilize and perform one of the following actions: 
a.
 
Press the f2 (Accept) key to accept the displayed value. 
or 
b.
 
Press the f3 (Edit) key and use the f3 (Clear) key and numeric keypad to enter the 
relative mV value. Press the f2 (Accept) key to confirm the entered value. 
5.
 
The meter will display the calibration summary including the mV offset. Press the  
f3 (Cal Done) key to save the calibration and export the data to the calibration log. Press 
the measure (esc) key to proceed to the measurement mode. 
 
 
42  
|
  
Orion V
ERSA 
S
TAR
 Meter User Manual 
Thermo Scientific 


Chapter 4  
|
  
pH, pH/ISE and pH/LogR Measurement Modules 
ORP Calibration 
V
ERSA 
S
TAR
 meters with pH, pH/ISE or pH/LogR modules offer a simple, automatic calibration 
to the E
H
 value (relative to the Standard Hydrogen Electrode, 420 mV at 25°C) in the ORP 
mode when using the following items: 
·
 
Thermo Scientific Orion ORP standard (Catalog Number 967901 or 967961) 
·
 
Thermo Scientific Orion ORP electrode (Catalog Number 9678BNWP, 9778BNWP, 
9180BNMD or 9179BNMD)  
·
 
Thermo Scientific Orion 4 M KCl filling solution (Catalog Number 900011) or 4 M KCl gel 
(in the 9179BNMD electrode) 
Orion ORP standard is nonhazardous, stable and will not change over time, so it is an ideal 
standard for any ORP measurement system. One ORP standard can be used for calibration. 
Always use fresh ORP standard. Prepare the ORP electrode according to the instructions in the 
electrode user manual. Connect the ORP electrode and any other electrodes to be used to the 
module inputs on the selected channel. 
In the measurement mode, press the channel key until the channel with the pH, pH/ISE or 
pH/LogR module with ORP electrode connected is shown. If needed, press the mode key until 
ORP Measurement is shown as the measurement mode.  
1.
 
Press the f1 (Cal) key to start the calibration.  
a.
 
If more than one channel is displayed in the measurement mode, press the  or  
key to highlight the desired channel and press the f3 (Select) key. 
2.
 
Rinse the ORP electrode and any other electrodes in use with distilled water, blot dry with a 
lint-free tissue and place into the standard. 
3.
 
When the electrode and standard are ready, press the f3 (Start) key.  
a.
 
If using a stirrer probe, the stirrer probe will start stirring when the f3 (Start) key is 
pressed and stop stirring when the reading stabilizes. 
4.
 
Wait for the mV value on the meter to stabilize, indicated by the flashing stabilizing or solid 
Ready icons. When the solid Ready icon appears, press the f2 (Accept) key to confirm the 
ORP standard value.  
5.
 
The meter will display the calibration summary including the mV offset. Press the  
f3 (Cal Done) key to save the calibration and then press the measure (esc) key to return to 
measurement mode. 
 
 
 
Thermo Scientific 
Orion V
ERSA 
S
TAR
 Meter User Manual  
|
  
43
 