Chapter 4  
|
  
pH, pH/ISE and pH/LogR Measurement Modules 
VSTAR-PH V
ERSA 
S
TAR
 pH Module 
 
1.
 
Connect a pH electrode or ORP electrode to the BNC connector on the module. 
2.
 
Connect an ATC temperature probe to the 8 pin MiniDIN connector on the module. 
3.
 
Connect a reference half-cell electrode to the pin tip connector on the module (for use when 
a half-cell sensing electrode with BNC connector is used). 
4.
 
Prepare the electrodes for use as instructed in the electrode user manuals.  
VSTAR-ISE V
ERSA 
S
TAR
 pH/ISE Module 
1.
 
Connect an ion selective electrode (ISE), pH electrode or ORP electrode to the BNC 
connector on the module. 
2.
 
Connect an ATC temperature probe to the 8 pin MiniDIN connector on the module. 
3.
 
Connect a reference half-cell electrode to the pin tip connector on the module (for use when 
a half-cell sensing electrode with BNC connector is used). 
4.
 
Prepare the electrodes for use as instructed in the electrode user manuals.  
VSTAR-LR V
ERSA 
S
TAR
 pH/LogR Module 
 
1.
 
Connect a pH electrode or ORP electrode to the BNC connector on the module. LogR 
technology allows the temperature to be measured using the glass bulb of a pH electrode. 
2.
 
Connect an ATC temperature probe to the 8 pin MiniDIN connector on the module. 
3.
 
Connect a reference half-cell electrode to the pin tip connector on the module (for use when 
a half-cell sensing electrode with BNC connector is used). 
4.
 
Prepare the electrodes for use as instructed in the electrode user manuals.  
8 pin MiniDIN 
BNC 
Pin Tip 
8 pin MiniDIN 
BNC 
Pin Tip 
Thermo Scientific 
Orion V
ERSA 
S
TAR
 Meter User Manual  
|
  
35
 


Chapter 4  
|
  
pH, pH/ISE and pH/LogR Measurement Modules 
Channel 1-4 Setup Menus 
Mode Menus for pH, mV, Relative mV, ORP and ISE 
Within the Channel 1, Channel 2, Channel 3 and Channel 4 setup menus are Method, Mode 
and Temperature submenus, which can be used to customize measurement settings and 
parameters for the selected channel. When using a pH/ISE module that is set to the ISE 
measurement mode, an Incremental Techniques submenu is also available. This section 
provides detailed information on the Mode menus for each measurement parameter available 
when using a pH, pH/ISE or pH/LogR module.  
1.
 
In the measurement mode, press the setup key. 
2.
 
Press the  or  key to highlight Channel 1, Channel 2, Channel 3 or Channel 4 (channel 
number the module is connected with) and press the f3 (Select) key. 
 
3.
 
Press the  or  key to highlight Mode and press the f3 (Select) key. 
 
4.
 
Press the  or  key to highlight pH, mV, Relative mV, ORP or ISE (pH/ISE module only) 
and press the f3 (Select) key.  
 
5.
 
Press the  or  key to highlight a menu option and press the  or  key to highlight a 
submenu option. Onscreen text prompts provide instructions for changing the parameters. 
36  
|
  
Orion V
ERSA 
S
TAR
 Meter User Manual 
Thermo Scientific 


Chapter 4  
|
  
pH, pH/ISE and pH/LogR Measurement Modules 
a.
 
Use the popup keypad screen
 for 
Electrode SI No 
and Sample ID. Press the 
▲, ▼, ◄ or ► key to 
highlight an alphanumeric character (numeric 
only for 
Auto Increment Sample ID), p
ress the f2 
(Enter) key to input the character and repeat 
until the desired label is entered. Press the f1 
(Done) key to save the entry.  
 
pH Mode Menu 
  
 
Main Menu 
Settings 
Details 
Electrode SI No 
---- 
Use the alphanumeric keypad to enter the electrode 
serial number or clear all to shown no value (----) 
Sample ID 
Off, Manual, Auto Increment 
Use the popup keypad screen to enter an 
alphanumeric manual ID or a numeric value as the 
starting value for auto incremental ID 
Resolution 
0.1,  0.01,  0.001 
Set the resolution of the measurement value 
Stability 
Smart, Fast, Medium, Slow  Set when a measurement is recognized as stable 
Averaging 
Off,  Automatic Smart 
Set averaging for faster measurement stability 
Read Type 
Auto-Read, Timed,  
Single-Shot, Continuous 
For Timed and Single-Shot, set the time in hours, 
minutes, seconds (HH:MM:SS) using the arrow keys 
Buffer Group 
USA, DIN, CUSTOM 
USA:   1.68, 4.01, 7.00, 10.01, 12.46 
DIN:   1.68, 4.01, 6.86, 9.18 
Custom: Enter 1 to 6 buffer values ≥ 1.00 pH apart 
Isopotential Value  Off,  On  
Default value is 7.000 
Linear Regression  Off,  On 
Set point-to-point (Off) or single best fit (On) slope 
Alarm 
Limit Alarm 
Off,  On 
Set high and low limit values to activate alarm  
Cal Due Alarm  Off,  On 
Set a calibration interval (hours) to activate alarm 
Set Point Alarm  Off,  On 
Set a base and offset value to activate alarm 
Detailed information on these features is available in the 
Appendix
 
section of this user manual. 