Chapter
3
|
Meter Setup Menus
Icon
Setup Menu
Description
Channel 1
Channel 2
Channel 3
Channel 4
Use the channel-specific setup menus to
customize the measurement
parameter settings for each module connected to the meter. The type of
module will define which measurement parameters and settings are
available for the channel. Each channel setup menu icon shows the current
measurement mode (i.e. pH, ISE, Cond, RDO) for the module.
Instrument
Settings
Use the instrument settings setup menu to update meter settings for the
display, sound, date
and time, language, data transfer, stirrer speed and
channel assignment, screen saver and auto-shutoff feature.
Log View
Access the Log View menu to view, print or delete saved data in the data
log and view or print the ten most recent calibrations per parameter in the
calibration log.
Diagnostics
Access the Diagnostics menu
to initiate a meter reset, test electrode
stability, perform a meter self test or view meter and module serial numbers
and software revisions.
Channel 1-4 Setup Menus
Within the Channel 1, Channel 2, Channel 3 and Channel 4 setup menus are Method, Mode
and Temperature submenus, which can be used to customize measurement settings and
parameters for the selected channel. The number of channel-specific setup menus and the
options within each channel-specific setup menu will vary depending on the number and types
of modules that are connected to the meter.
·
Access the Method menu to load, copy, edit, create or delete methods.
·
Select the Mode menu to review and update the measurement mode and parameter
settings for the selected channel and module. Refer to the individual module chapters for
detailed information on the available settings for each measurement mode.
·
Use the Temperature menu to set the temperature units, set the temperature input source,
perform a temperature calibration on an ATC temperature probe, conductivity probe with
built-in temperature or dissolved oxygen probe with built-in temperature when connected to
the meter or to enter a manual temperature value. The settings shown below may be
different depending on the type and number of modules connected to the meter.
Thermo
Scientific
Orion V
ERSA
S
TAR
Meter User Manual
|
23
Chapter
3
|
Meter Setup Menus
·
When using an ion selective electrode (ISE) and V
ERSA
S
TAR
pH/ISE module, use the Incremental Techniques menu to
perform a single
known addition, single known subtraction,
double known addition, double known subtraction, single
analate addition or single analate subtraction analysis.
·
When using a Thermo Scientific
™
Orion
™
RDO
™
optical
dissolved oxygen probe and V
ERSA
S
TAR
RDO/DO module, use
the RDO Electrode Info menu to view the serial number,
remaining useful life and date of manufacture of the optical cap.
Main Menu
Submenu
Settings
Method
Load
Method list from selected channel to load a method
Copy
Method list from selected
channel to copy a method
Edit/Create
Method list from selected channel to edit existing
method or create a new method
Delete
Method list from selected channel to delete a method
Mode
Module dependent
Module dependent, refer to individual module chapters
Temperature
Temperature Unit
°C °F
Temperature Input
Ch1-ATC Ch2-ATC Ch3-ATC Ch4-ATC
MAN
Temperature Cal
MAN Ch#-ATC Ch#-LogR All ATC
(Module and setting dependent)
Incremental
Techniques
(Module
dependent)
Single Known Addition
The Incremental Techniques menu is displayed only
when the measurement mode
is set to ISE using the
pH/ISE module.
Refer to
Chapter 4 pH, pH/ISE and pH/LogR
Measurement Modules
for detailed information on the
incremental technique functions.
Single Known Subtraction
Double Known Addition
Double
Known Subtraction
Single Analate Addition
Single Analate Subtraction
RDO Electrode
Information
(Module
dependent)
Serial No:
Cap Life Left:
Date of Mfg:
The RDO Electrode Information menu is displayed only
when an RDO probe is connected to the meter using
the RDO/DO module.
Note: If a new module is connected to the meter, return to the measurement mode, wait for the
Channel ID display to update with the new module and then access the Main Setup menu. This
ensures the channel-specific setup menu for the new module will be displayed corrected.
24
|
Orion V
ERSA
S
TAR
Meter User Manual
Thermo Scientific
Chapter
3
|
Meter Setup Menus
Method Menu
Up to ten channel-specific methods can be saved in the Method menu within the Channel 1,
Channel 2, Channel 3 and Channel 4 setup menus for fast and easy recollection of specific
measurement modes and channel-specific setup parameters.
The open method is updated whenever the measurement mode and channel setup parameters
are changed and is not password protected. Protected methods require a user name (up to nine
characters) and password (up to six characters) to be created.
When a
protected method is created, loaded and active in the measurement mode and a
calibration is performed, that calibration will be saved to the method so every time the method is
loaded, the corresponding calibration will also be loaded. This feature is helpful when using two
or more electrodes on one channel. For example, a pH electrode and an ion selective electrode
(ISE) used on the same VSTAR-ISE pH/ISE module or a low range conductivity probe and a
standard range conductivity probe used on the same VSTAR-CND conductivity module.
Channel
Open Method
Protected Methods
Channel 1 M100
M101, M102, M103, M104, M105, M106, M107, M108, M109, M110
Channel 2 M200
M201, M202, M203, M204, M205, M206, M207, M208, M209, M210
Channel 3 M300
M301, M302, M303, M304, M305, M306, M307, M308, M309, M310
Channel 4 M400
M401, M402, M403, M404, M405, M406, M407, M408, M409, M410
Load
Use the load function to activate a method for use in the measurement mode. If no protected
method has been created, the open method is active. Loading a blank method (date and time
shown as dashes on the method list) will activate the default measurement mode and channel
setup parameters for that channel. A password is not required to load a protected method.
Copy
Use the copy function to save the active measurement mode and channel setup parameters to
a new or different method. If the copy function is used to create a new protected method, a user
name and password must be created before the new protected method can be edited. If the
copy function is used to overwrite an existing protected method, the password for the existing
method must be entered before the action can be completed. Once a method is copied, it is
automatically loaded and activated for use in the measurement mode.