Chapter
3
|
Meter Setup Menus
c.
When viewing individual data log points, press the
◄ or
► key to scroll through
additional data log points.
3.
To print the data log:
a.
Review the meter output settings (Communication, Printing, Printing Format) in the
Instrument Settings
setup menu to ensure the meter output settings are correct and
interface the meter with a printer or computer.
b.
Press the
▲ or
▼ key to highlight Print single point, Print range
or
Print all and press
the
f3 (Select) key. For a single point or range, enter the data point value(s) and press
the
f3 (Print) key.
4.
To delete the data log:
a.
Press the
▲ or
▼ key to highlight
Delete all and press the
f3 (Select) key.
b.
Press the
f1 (Yes) key to delete the log or press the
f3 (No) key to cancel the function.
5.
To view statistics:
a.
Press the
▲ or
▼ key to highlight
Statistics
and press the
f3 (Select) key.
b.
Press the
▲ or
▼ key to select a measurement parameter. Statistical values for all
data log points with the selected parameter will be displayed.
Calibration Log
The calibration log contains up to the ten most recent calibrations per pH, RmV, ORP, ISE,
incremental technique, conductivity, DO and RDO parameter.
1.
Press the
◄ or
► key to highlight Cal Log.
2.
To view the calibration log:
a.
Press the
▲ or
▼ key to highlight the desired calibration parameter and press
the
f3 (Select) key. Details for up to the last ten calibrations by parameter can be
displayed. Press the
▲ or
▼ key to view each calibration for the selected parameter.
3.
To print a calibration log:
a.
Review the meter output settings (Communication, Printing, Printing Format) in the
Instrument Settings
setup menu to ensure the meter output settings are correct and
interface the meter with a printer or computer.
b.
Press the ▲ or ▼ key to highlight the desired calibration
log and press the
f3 (Print) key.
32
|
Orion V
ERSA
S
TAR
Meter User Manual
Thermo
Scientific
Chapter
3
|
Meter Setup Menus
Diagnostics Menu
Use the Diagnostics Menu to
initiate a meter reset, test electrode stability, perform a meter self
test or view meter and module serial numbers and software revisions.
·
Factory Reset – All meter settings are reset to factory defaults. Data log, calibration log and
methods are deleted.
·
User Reset – All settings for the selected channel are reset to factory defaults. Data log
and calibration log are retained. Methods without a password are erased.
·
Stability Test – Tests the stability of a pH electrode, ORP electrode or ion selective
electrode (ISE) by measuring the mV per minute drift and noise of the electrode.
·
Self Test – Tests the keypad, meter accuracy and stirrer inputs. A self test is also
performed automatically by the meter at each power up.
·
About Meter – Displays the meter and module software revisions and serial numbers.
Main Menu
Submenu
Notes
Factory Reset
Password
Default password is 1111.
User Reset
Channel 1
The channel setup menu for the
selected channel will be
reset to default
settings and methods without a
password will be erased.
Channel 2
Channel 3
Channel 4
Stability Test
Channel #
Only channels with
BNC connectors are
displayed in the stability test submenu.
Channel #
Self Test
Keypad check
The accuracy check is performed on
channels with BNC connectors only.
A stirrer probe (part # 096019) is
required to perform the stirrer check.
Accuracy check
Stirrer check
About Meter
Serial No. SW Rev.
The meter serial number and software
revision will
be different from the module
serial numbers and software revisions.
Meter
######
#.##
Channel 1 ######
#.##
Channel 2 ######
#.##
Channel 3 ######
#.##
Channel 4 ######
#.##
Thermo Scientific
Orion V
ERSA
S
TAR
Meter User Manual
|
33
4
pH, pH/ISE and pH/LogR
Measurement Modules
Connecting
Electrodes to the Module
To begin using the V
ERSA
S
TAR
meter and module system, prepare the meter and insert the
module into an available channel on the back of the meter. See
Chapter 2 Meter Basics
for
instructions. For information on customizing meter settings, refer to
Chapter 3 Setup Menu
.
Ø
It is recommended that electrodes be placed in the meter-attached electrode stand for easy
movement in and out of beakers during calibration, sample measurement and storage.
Module Catalog #
VSTAR-PH
VSTAR-ISE
VSTAR-LR
Label
pH
pH/ISE
pH/LogR
Measurement Modes pH, mV, RmV, ORP pH, mV, RmV, ORP, ISE pH, mV, RmV, ORP
Temperature Modes
Automatic, Manual
Automatic, Manual
Automatic, Manual, LogR
C H A P T E R 4
34
|
Orion V
ERSA
S
TAR
Meter User Manual