Chapter 3  
|
  
Meter Setup Menus 
Instrument Settings Setup Menu  
  
 
Use the Instrument Settings menu to review and update meter settings. The default settings are 
in bold or shown with the selection box checked in the following table. To navigate the menu
press the  or  key to highlight a main menu option and press the  or  key to highlight 
a submenu option. Onscreen text prompts provide instructions for changing parameters.    
·
 
Display – Adjust the display brightness, select the information shown on the measurement 
display and set which channel combinations are shown when the Channel key is pressed.   
·
 
Communication – Select the method used for information transfer to an external device and 
set the baud rate when RS232 is selected as the transfer method.   
·
 
Buzzer – Turn the sound on or off each time a key is pressed, when the main 
measurement value becomes stable and when an alarm is triggered.      
·
 
Stirrer – Connect one or two Orion Star stirrer probes (Catalog Number 096019) to the 
meter, assign a stirrer probe to each measurement channel and set the stirring speed of 
each stirrer probe from 1 (slowest) to 5 (fastest).  
·
 
Calendar – Enter the time, set the time format as 12 hours (am/pm) or 24 hours, enter the 
date and set the date format as Day-Month-Year (all numeric), Month-Day-Year (all 
numeric) or Day-Month-Year (DD-MMM-YY, three letter short form of month). 
·
 
Language – Set the language used for the meter interface.  
·
 
Data log – Turn the data log function on or off. 
·
 
Printing – Turn the printing function to an external device on or off. 
·
 
Printing Format – Select how information will be transferred to an external device, as 
comma delimited (CSV) or standard text (Printer). 
·
 
Screen Saver – Turn the screen saver on or off. The screen saver is shown when the 
meter enters auto-shutoff mode or when the meter is manually put into standby mode by 
pressing the power key. The time is displayed when the screen saver is on and a blank 
display is shown when the screen saver is off.   
·
 
Auto-Shutoff – Turn the meter automatic shutoff feature on or off.  
 
 
Thermo Scientific 
Orion V
ERSA 
S
TAR
 Meter User Manual  
|
  
29
 


Chapter 3  
|
  
Meter Setup Menus 
Main Menu 
Submenu 
Settings 
Display 
Brightness 
Level  1 
Level   2 
Level  3 
Level  4 
Level  5 
Display View 
Channel 
þ
 
Measurement Mode 
þ
 
Method 
þ
 
Stable 
þ
 
Temperature 
þ
 
Secondary Parameter 
þ
 
Calibration Details 
þ
 
User ID 
¨
 
Sample ID 
þ
 
Electrode Serial No. 
þ
 
Display Format 
Channel 1 
þ
 
Channel 2 
þ
 
Channel 3 
þ
 
Channel 4 
þ
 
Channel 1 & 2 
þ
 
Channel 1 & 3 
þ
 
Channel 1 & 4 
þ
 
Channel 2 & 3 
þ
 
Channel 2 & 4 
þ
 
Channel 3 & 4 
þ
 
Channel 1 & 2 & 3 
þ
 
Channel 1 & 2 & 4 
þ
 
Channel 1 & 3 & 4 
þ
 
Channel 2 & 3 & 4 
þ
 
Channel 1 & 2 & 3 & 4 
þ
 
Communication 
USB 
 
RS232 
1200       2400  4800  9600  19200  38400    57600    115200 
USB PRINTER   
Buzzer 
Key Press 
Off   
On 
Ready 
Off   
On 
Alarm 
Off   
On 
Stirrer 
 
 
Channel 1  
Channel 2 
Channel 3  
Channel 4  
Speed   
Stirrer 1 
 ¨ 
 ¨ 
 þ 
 þ 
1  2  3  4  5 
Stirrer 2 
 þ 
 þ 
 ¨ 
 ¨ 
1  2  3  4  5 
Calendar 
Time 
 
Hour  Minutes 
 
12 Hrs  ## 
## 
AM/PM 
 
Hour  Minutes  
24 Hrs  ## 
## 
Date 
DD-MM-YY 
Day  Month  Year 
MM-DD-YY 
Month  Day  Year 
DD-MMM-YY 
Day  Month  Year 
Language 
Current setting  English   Spanish   German   Italian   French   Chinese   Portuguese 
Data Log 
 
Off   
On 
Printing 
 
Off   
On 
Print Format 
 
CSV   
Printer 
Screen Saver 
 
Off   
On 
Auto Shut Off 
 
Off   
On 
 
 
 
 
30  
|
  
Orion V
ERSA 
S
TAR
 Meter User Manual 
Thermo Scientific 


Chapter 3  
|
  
Meter Setup Menus 
Log View Menu 
 
Use the Log View menu to access the data log and calibration log. All information in the data 
and calibration logs is protected by the nonvolatile memory, so stored information is saved in 
the meter even when it is not connected to a power supply.  
Data Log 
V
ERSA 
S
TAR
 meters offer a 2000 point data log. Each point includes measurements from one to 
four channels, depending on which channels were actively displayed in the measurement mode 
when the point was saved. When the data log function is turned on, the read type selected for 
each displayed channel (Auto-Read, Timed, Single-Shot or Continuous) determines how the 
point is saved to the data log. Refer to the 
Measurement Read Types
 section for detailed 
information sending measurements to the data log.  
1.
 
Press the  or  key to highlight Data Log. 
2.
 
To view the data log: 
a.
 
Press the  or  key to highlight View and press the f3 (Select) key.  
 
b.
 
A table will be displayed with the data log number, main measurement value of each 
channel, date and time for each point. Press the  or  key to highlight a point and 
press the enter key to view detailed data for that point.  
 
Thermo Scientific 
Orion V
ERSA 
S
TAR
 Meter User Manual  
|
  
31
 