Orion versa star user Manual


   |    pH, pH/ISE and pH/LogR Measurement Modules



Yüklə 6,2 Kb.
Pdf görüntüsü
səhifə16/44
tarix05.01.2018
ölçüsü6,2 Kb.
#19795
1   ...   12   13   14   15   16   17   18   19   ...   44

Chapter 4  
|
  
pH, pH/ISE and pH/LogR Measurement Modules 
ISE Calibration (pH/ISE Module Only) 
One to six standards can be used for ISE (ion selective electrode) calibration. If more than one 
standard is used for calibration, start with the lowest concentration standard and work up to the 
highest concentration standard last. Always use fresh standards. Select standards that bracket 
the sample concentration and are one decade (10 times) apart in concentration. To accurately 
prepare calibration standards from a stock solution, serial dilution is recommended using 
calibrated pipettes. If ISA will be added to samples, the same ISA must be added to all 
standards prior to calibration to ensure a consistent dilution factor.  
Prepare the ISE according to the instructions in the electrode user manual. Connect the ISE 
and any other electrodes to be used to the module inputs on the selected channel. 
In the measurement mode, press the channel key until the channel with the pH/ISE module 
with ISE connected is shown. If needed, press the mode key until ISE Measurement is shown 
as the measurement mode and use the 
ISE Mode Menu 
to set the Electrode Type.  
1.
 
Press the f1 (Cal) key to start the calibration. 
a.
 
If more than one channel is displayed in the measurement mode, press the  or  
key to highlight the desired channel and press the f3 (Select) key. 
2.
 
Rinse the ISE and any other electrodes in use with distilled water or rinse solution required 
for that electrode. Blot the electrode dry with a lint-free tissue or shake the electrode and 
then place into the standard. Do not wipe or rub the sensing element of the electrode. 
Follow all electrode rinsing and drying instructions or exceptions in the ISE user manual. 
3.
 
When the electrode and standard are ready, press the f3 (Start) key. 
a.
 
If using a stirrer probe, the stirrer probe will start stirring when the f3 (Start) key is 
pressed and stop stirring when the reading stabilizes. 
4.
 
Wait for the concentration value to stabilize and perform one of the following actions: 
a.
 
Press the f2 (Accept) key to accept the displayed value. 
or 
b.
 
Press the f3 (Edit) key and use the numeric keypad to enter the concentration value. 
Press the f2 (Accept) key to confirm the entered value. 
5.
 
Press the f2 (Next) key to proceed to the next standard and repeat steps 2 through 4 or 
press the f3 (Cal Done) key to save and end the calibration. 
6.
 
The meter will display the calibration summary including the slope and export the data to the 
calibration log. Press the measure (esc) key to proceed to the measurement mode. 
 
ISE Calibration Editing 
Once the f3 (Cal Done) key is pressed and the calibration is saved, press the f2 (Cal Edit) key 
to edit the calibration data and fix individual points without a full recalibration. In the calibration 
edit mode, press the  or  key to highlight a calibration point and press the f2 (Delete) key 
to delete the point or press the f3 (Re-Measure) key to re-measure the point. Follow the 
onscreen text prompts and instructions to edit the selected calibration point. 
44  
|
  
Orion V
ERSA 
S
TAR
 Meter User Manual 
Thermo Scientific 


Chapter 4  
|
  
pH, pH/ISE and pH/LogR Measurement Modules 
Temperature Calibration 
V
ERSA 
S
TAR
 meters have a relative temperature accuracy of ±0.1°C. ATC probes have varying 
relative accuracies, usually ±0.5°C to ±2°C. Use the temperature calibration function only 
when necessary. Since the temperature offset calculated during the calibration is applied to all 
future temperature measurements, recalibrate the temperature if a different ATC probe is used.   
ATC Probe Temperature Calibration 
ATC (Automatic Temperature Compensation) probes are defined as both separate temperature 
probes and those built into electrodes. A temperature calibration can be performed for individual 
ATC probes or simultaneously for all ATC probes connected to the meter. Prepare and connect 
all ATC probes to be calibrated to the appropriate module inputs on the selected channels.  
1.
 
In the measurement mode, press the setup key.  
2.
 
Press the ◄ or ► key to highlight Channel 1, Channel 2, Channel 3 or Channel 4 and 
press the f3 (Select) key.  
3.
 
Press the ◄ or ►key to highlight Temperature and press the f3 (Select) key.  
4.
 
Press the  or  key to highlight Temperature Input and press the  or  key to 
highlight an ATC probe (i.e. Ch2-ATC) as the temperature source for the selected channel. 
5.
 
Press the  or  key to highlight Temperature Cal and press the  or  key to highlight 
the individual ATC probe (i.e. Ch2-ATC) or All ATC and press the f3 (Select) key. 
6.
 
Place the ATC probe(s) into a solution with a known, stable temperature. It is recommended 
that two NIST-traceable thermometers be used to measure and verify the temperature. 
7.
 
Press the f3 (Edit) key, use the numeric keypad to enter the temperature of the solution and 
press the f2 (Accept) key to save the temperature value and update the offset. 
8.
 
If applicable, press the  or  key to highlight the reference temperature of the next ATC 
probe and repeat steps 6 and 7 for each ATC probe connected to the meter.  
9.
 
Once all ATC probes are calibrated, press the f1 (Back) key to exit the calibration. 
Manual Temperature Entry 
A manual temperature value can be entered when MAN is selected as the temperature input.  
1.
 
In the measurement mode, press the setup key.  
2.
 
Press the ◄ or ► key to highlight Channel 1, Channel 2, Channel 3 or Channel 4 and 
press the f3 (Select) key.  
3.
 
Press the ◄ or ►key to highlight Temperature and press the f3 (Select) key.  
4.
 
If needed, press the  or  key to highlight Temperature Input and press the  or  key 
to highlight MAN. 
5.
 
Press the  or  key to highlight Temperature Cal and press the f3 (Select) key. Press 
the f3 (Clear) key, input the temperature value using the numeric keypad and then press 
the f2 (Enter) key to save the value.  
Thermo Scientific 
Orion V
ERSA 
S
TAR
 Meter User Manual  
|
  
45
 


Yüklə 6,2 Kb.

Dostları ilə paylaş:
1   ...   12   13   14   15   16   17   18   19   ...   44




Verilənlər bazası müəlliflik hüququ ilə müdafiə olunur ©genderi.org 2024
rəhbərliyinə müraciət

    Ana səhifə