28
TESTING PROCEDURES
Viewing result records
The patient and control results are stored in the memory of the Afinion™ AS100 Analyzer.The last 500
patient results and the last 500 control results are saved in separate records.When exceeding the capacity
of 500 results, the oldest result will be deleted.
The following parameters are listed for each run: Date and time, run number, patient ID/control ID, operator
ID, lot numbers of Test Cartridge and Test Result.
If a printer is connected, the test results can be printed one by one.
Viewing patient results
Start-up menu
Main menu
Viewing control results
Start-up menu
Main menu
1
Touch
to enter
main menu.
2
Touch
to enter
control records.
3
The last control result is
displayed.
To view more results touch
or
If a printer is connected,
touch
to print the
result.
1
Touch
to enter
main menu.
2
Touch
to enter
patient records.
3
The last patient result is
displayed.
To view more results touch
or
If a printer is connected,
touch
to print the
result.
09-38555
29
INFORMATION CODES AND TROUBLESHOOTING
When an information code appears
Information codes that might appear during use of the Afinion™ AS100 Analyzer refer to specific information
or error messages.The code numbers, the possible causes and actions to take are listed below.
If the Analyzer detects a problem during processing of a Test Cartridge, the test will automatically be aborted
and the Test Cartridge will be safely moved to the cartridge chamber. Proceed as follows:
Do not re-use a Test Cartridge that has been rejected by the Analyzer. Collect a new sample and
repeat the test with a new Test Cartridge.
Information messages caused by assay specific limitations
[#]
Cause
Action to take
103
Hemoglobin too low
Consult the Afinion™ HbA1c Package Insert,
section Test result reporting.
104
Hemoglobin too high
Consult the Afinion™ HbA1c Package Insert,
section Test result reporting.
105
HbA1c too low
Consult the Afinion™ HbA1c Package Insert,
section Test result reporting.
106
HbA1c to high
Consult the Afinion™ HbA1c Package Insert,
section Test result reporting.
107
Creatine too high
Consult the Afinion™ ACR Package Insert
108
Blood in urine
Consult the Afinion™ ACR Package Insert
1
Record the code number (#)
and touch
to accept.
The lid opens automatically.
2
Remove the Test Cartridge.
If the Test Cartridge is not
ejected, restart the Analyzer.
Do not re-use the Test
Cartridge.
3
Look up the possible cause
from the table below, and take
actions to solve the problem.
If the problem persists, contact
your local Afinion™ supplier
(see “Service information”
page 31).
09-38555
30
INFORMATION CODES AND TROUBLESHOOTING
Error messages caused by sample or Test Cartridge failure
[#]
Cause
Action to take
201
Insufficient sample volume:
Repeat the test with a new sample and Test Cartridge.
- Empty capillary
Ensure that the capillary is completely filled with no air bubbles.
- Air bubble in capillary
- Capillary incompletely filled
202
Excess sample on the sampling Repeat the test with a new sample and Test Cartridge.
device exterior
Ensure that only the tip of the capillary is in contact with
the sample.
203
Wrong sample material
Repeat the test with a new sample and Test Cartridge.
Ensure that proper sample material is used (see Package
Insert for the Afinion™ test in use, section Specimen collection
and storage).
204
Coagulated sample
Repeat the test with a new sample and Test Cartridge.
The time from filling the capillary until analyzing the Test
Cartridge should be as short as possible.
Hemolysed blood sample
Consult the Afinion™ HbA1c Package Insert.
(Afinion™ HbA1c)
Repeat the test with a new sample and Test Cartridge.
Test Cartridge or Analyzer
Repeat the test with a new sample and Test Cartridge.
failure
If the problem persists, restart the Analyzer and run controls.
205
Capillary cracked or damaged
Repeat the test with a new sample and Test Cartridge.
Inspect the sampling device before use and handle with care.
206
Barcode label not readable
Repeat the test with a new sample and Test Cartridge.
(dirty or damaged)
If the problem persists, restart the Analyzer and run controls.
207
No sampling device inserted
Repeat the test with a new sample and Test Cartridge.
Used sampling device belongs
Repeat the test with a new sample and Test Cartridge. Ensure
to another Afinion™ test
that the sampling device and Test Cartridge have the same label
color.
Label on sampling device not
Repeat the test with a new sample and Test Cartridge. Ensure
readable (dirty or damaged)
that the label is clean.
208
Test Cartridge previously used
Repeat the test with a new sample and Test Cartridge.
209
Test Cartridge has passed
Check expiration date on the Cartridge pouch or box.
expiration date
Repeat the test using a new sample and a new Test Cartridge
from another lot.
The date in the Analyzer is
Check the date in the Analyzer to make sure it is set correctly.
incorrectly set
Repeat the test with a new sample and Test Cartridge.
210
Test Cartridge temperature
Repeat the test with a new sample and a new Test Cartridge
too low
within recommended operating temperature range
(see Package Insert for the Afinion™ test in use).
211
Test Cartridge temperature
Repeat the test with a new sample and a new
too high
Test Cartridge within recommended operating temperature
range (see Package Insert for the Afinion™ test in use, section
Test procedure).
212
Test Cartridge not recognized
The software version required for this test is not installed.
by the Analyzer
Contact your local supplier for assistance.
09-38555
31
[#]
Cause
Action to take
213
Test Cartridge or Analyzer
Repeat the test with a new sample and Test Cartridge.
214
failure
If the problem persists, restart the Analyzer and run controls.
215
216
Controls are not run after the
Run one Afinion™ Control following the procedure
software upgrade of the
on page 27 “Analyzing a control”.
Analyzer
INFORMATION CODES AND TROUBLESHOOTING
Error messages caused by Analyzer failure
[#]
Cause
Action to take
301
Self-test failed
Restart the Analyzer.
302
Analyzer failure
Restart the Analyzer and run controls.
Repeat the test with a new sample and Test Cartridge.
303
Analyzer temperature is
Wait until the Analyzer has cooled down.
too high
Ensure that the operating temperature is within
recommended range (15-32ºC, 59-89ºF).
Repeat the test with a new sample and Test Cartridge.
304
Analyzer temperature is
Ensure that the operating temperature is within
too low
recommended range (15-32ºC, 59-89ºF).
Repeat the test with a new sample and Test Cartridge.
305
- Printer improperly connected Switch off the Analyzer, reconnect the printer and restart
- Malfunction of the printer
the Analyzer.
If the message persists, see the printer User Manual.
ERROR
Touch screen failure
Restart the Analyzer.
27
28
Start-up procedure failed
Contact your local supplier for assistance.
29
Self-test error.
Analyzer in
Severe instrument failure Contact your local supplier for assistance.
non-operative
mode
Service information
The laboratory must notify the manufacturer of this test system of any performance, perceived or
validated, that does not meet the performance specifications as outlined in the instructions.
The manufacturer provides a toll free line for technical support.
Call 1-877-4-Afinion or 1-877-423-4646.
The toll free number is available for use only in the United States of America.
E-mail: techsupport@us.axis-shield.com
Before asking for assistance, please record the following information:
• Afinion™ AS100 Analyzer serial number (SN) – see page 1
• Software version number – see page 1
• Afinion™ test type
• Test Cartridge lot number – see foil pouch or kit container
• Control name and lot number – see vial label
• Control results obtained
• Description of the problem with reference to information codes and error messages
09-38555
32
MAINTENANCE AND WARRANTY
Cleaning and maintenance
No maintenance of the Afinion
TM
AS100 Analyzer is required other than cleaning the exterior and cartridge chamber.
Cleaning the exterior
Cleaning the exterior of the Afinion
TM
AS100 Analyzer should be performed whenever necessary. Most spills
and stains can be removed with water or a mild detergent.
- Switch off the Analyzer. Unplug the power supply when the shut down procedure is completed.
- Clean the outside of the Analyzer and the touch display with a clean, lint-free and non-abrasive
cloth dampened in water or a mild detergent.
- To disinfect the exterior of the instrument, first remove as much as possible of the spilled material
with a cloth dampened in the disinfectant (2% glutaraldehyde or 0.5% sodium hypochlorite).The
surface of the Analyzer should be exposed to the disinfectant for at least 10 minutes.
1
- Allow the Analyzer to air dry.
- Plug in the power supply and switch on the Analyzer.
• The Analyzer must be switched off and unplugged before cleaning.
• Do not use any cleaning liquid or equipment other than those recommended above.
• Do not immerse the Analyzer in water or other liquids.
Cleaning the cartridge chamber
The Afinion
TM
AS100 Analyzer Cleaning Kit (
1115269) should always be used for cleaning the cartridge
chamber.
The cartridge chamber should be cleaned immediately if materials or liquids are spilled in the cartridge chamber.
For regular maintenance (removal of dust particles etc.), the cartridge chamber should be cleaned every 30 days.
- Touch
to open the lid.
- Unplug the power supply.
- Wet a Cleaning Swab with 3 drops of water or a disinfectant (2% glutaraldehyde or 0.5% sodium
hypochlorite). Do not soak.
- Carefully remove spills and particles from the cartridge chamber using the moistened swab.
- To disinfect the cartridge chamber, the surface of the chamber should be exposed to the disinfectant
for at least 10 minutes.
1
- Wipe off any residual liquid from the cartridge chamber using a new, dry Cleaning Swab.
- Close the lid.
- Plug in the power supply and switch on the Analyzer.
• The Analyzer must be unplugged before cleaning.
• Do not use any cleaning liquid or equipment other than those recommended above.
• Do not allow liquid to drip off the Cleaning Swab into the Analyzer. If liquid drips into
the Analyzer, optics can be destroyed.
• Do not immerse the Analyzer in water or other liquids.
1
Clinical and Laboratory Standards Institute (CLSI) Guideline M29-A3: ”Protection of Laboratory Workers
From Occupationally Acquired Infections; Approved Guideline - Third Edition”. ISBN 1-56-238-567-4
09-38555
33
MAINTENANCE AND WARRANTY
Disposal of the Analyzer
For correct disposal according to the Directive 2002/96/EC (WEEE), contact your local
Afinion™ supplier.
Software upgrade
Consult the Afinion™ AS100 Analyzer SW Upgrade Package Insert.
Warranty
Axis-Shield warrants solely to the Buyer that the Afinion™ AS100 Analyzer will be free from defects
in materials and workmanship, when given normal, proper and intended usage, and will perform in
accordance with Axis-Shield’s specifications for a period of twelve months from the date of delivery.
At its expense, Axis-Shield agrees to repair, or at Axis-Shield’s option, replace with a new or recon-
ditioned unit, any Afinion™ AS100 Analyzer which is under warranty and not performing substantially
in accordance with applicable product specifications, provided that the Buyer has given Axis-Shield
notification of such warranty claim within the warranty period. If Axis-Shield is unable after reason-
able efforts to repair or replace the Afinion™ AS100 Analyzer not performing substantially in accor-
dance with applicable product specifications, the Buyer’s sole remedy shall be the refund of an
amount not to exceed the actual purchase price paid by the Buyer for the Afinion™ AS100 Analyzer.
All repairs will be done during normal working hours. All replaced parts shall become Axis-Shield’s
property. Axis-Shield may require the Buyer to ship the Afinion™ AS100 Analyzer to Axis-Shield or
elsewhere at Axis-Shield’s expense, for warranty service to be performed.
Notwithstanding the foregoing, Axis-Shield shall have no obligation to make repairs, replacements or
corrections which result, in whole or in part, from (i) an act of God or other unforeseen catastrophe,
(ii) any error, omission or negligence of the Buyer, (iii) improper or unauthorized use of the Afinion™
AS100 Analyzer, (iv) operating errors or the disregard of warnings and precautions described in this
Afinion™ AS100 Analyzer User Manual; (v) repairs performed to the Afinion™ AS100 Analyzer by
any person other than an authorized Axis-Shield service representative; (vi) use of the Afinion™
AS100 Analyzer in a manner for which it was not designed, (vii) causes external to the Afinion™
AS100 Analyzer such as, but not limited to, power failure or electric power surges, or (viii) use of the
Afinion™ AS100 Analyzer in combination with equipment, components or software not supplied by
Axis-Shield.
EXCEPT AS STATED IN THIS SECTION OF THE USER MANUAL, AXIS-SHIELD DISCLAIMS ALL
WARRANTIES,WHETHER EXPRESS OR IMPLIED,WRITTEN OR ORAL,WITH RESPECT TO THE
AFINION™ AS100 ANALYZER, INCLUDING ANY WARRANTY OF MERCHANTABILITY OR
FITNESS FOR A PARTICULAR PURPOSE. AXIS-SHIELD’S MAXIMUM LIABILITY ARISING OUT
OF THE SALE OF THE AFINION™ AS100 ANALYZER OR ITS USE,WHETHER BASED UPON
WARRANTY, CONTRACT,TORT OR OTHERWISE, SHALL NOT EXCEED THE ACTUAL
PURCHASE PRICE PAID BY THE BUYER FOR THE AFINION™ AS100 ANALYZER. IN NO
EVENT SHALL AXIS-SHIELD BE LIABLE FOR SPECIAL, INCIDENTAL OR CONSEQUENTIAL
DAMAGES, INCLUDING, BUT NOT LIMITED TO, LOSS OF PROFITS, LOSS OF DATA OR LOSS
OF USE DAMAGES, ARISING HEREUNDER OR FROM THE SALE OF THE AFINION™ AS100
ANALYZER. THIS WARRANTY MAY NOT BE TRANSFERRED BY THE BUYER.
The acknowledgement of claims shall be reported to your Axis-Shield Technical Care Specialist at
1-877-4-Afinion or 1-877-423-4646 or by E-mail to techsupport@us.axis-shield.com
09-38555
34
TECHNICAL SPECIFICATIONS
Afinion
TM
AS100 Analyzer
Analyzer
Size
12.6 x 6.7 x 6.7 in. / 320 x 170 x 170 mm
Weight
11 lbs. / 5 kg
Display
Standard LCD color display with back light and integrated touch panel.
Resolution: 240 x 320 pixels.Visible area: 2.3 x 3.0 in. / 58 x 77 mm.
Camera
640 x 480 pixels
Capacity of result records
500 patient results and 500 control results
SW update
via USB flash drive
Communication interface
RS 232C, USB 1.1
Power supply
Power cord adapter
Separate AC to DC power cord adapter. Double insulated.
Input
100-240 VAC, 50/60 Hz, 42 W
Output
24 VDC ± 5%, 1.75A
Output connector
0.2 x 0.1 in. / 5.5 x 2.5 mm plug. Positive (+) on inner pin.
Adapter standards
IEC/EN-60950/UL 1950 (USA and Europe). EN-60601/UL 2601
(Japan). Approved for in vitro diagnostic medical devices.
Operating conditions
Temperature
15-32ºC (59-89ºF)
Relative humidity
10-90%, non-condensing
Location
Dry, clean, horizontal surface. Avoid direct sunlight.
Test Cartridge temperature
According to specifications for the Afinion
TM
test in use.
Storage and transport (in the original container)
Temperature
-40 to 70ºC (-40 to 158ºF)
Relative humidity
10-93% at 40ºC (104ºF)
Accessories
For information regarding recommended barcode reader and printer please call 1-877-4-Afinion (1-877-423-4646)
or email techsupport@us.axis-shield.com.
09-38555
The following symbols and abbreviations are used in the product labeling and instructions for the Afinion™
AS100 Analyzer System.
Symbol/Abbreviation
Explanation
Conformity to the European directive 98/79/EC on in vitro diagnostic
medical devices
In Vitro Diagnostic Medical Device
Catalog number
Lot number
Serial number
Test Cartridge
Control C I
Control C II
Cleaning kit
Waste Electrical and Electronical Equipment (WEEE)
Contents are sufficient for one test
Contents are sufficient for 15 tests
Expiration date (year-month)
Storage temperature limitations
Manufacturer
Operator's handling
Warnings and precautions
Biological risk
Consult the Afinion™ user instructions
CLIA
Clinical Laboratory Improvement Amendment of 1988
LED
Light Emitting Diode
PC
Personal Computer
ID
Identification
HIS
Hospital Information System
LIS
Laboratory Information System
LCD
Liquid Crystal Display
AC
Alternating Current
DC
Direct Current
SYMBOLS AND ABBREVIATIONS
CONTROL C II
CLEANING KIT
CONTROL C I
35
09-38555
www
.signatur
.no
Axis-Shield PoC AS
P.O. Box 6863 Rodeløkka
NO-0504 Oslo, Norway
ISO 9001 and ISO 13485 certified company
Axis-Shield PoC USA
10 Commerce Way
Norton, MA 02766
USA
1115384
2007
-0
3