58
TUKAN 8k
that the program performs control of number of counts in selected region. Region
is defined by two parameters: Home channel and End channel 
Acquisition is stopped when number of counts in selected region exceeds value
set in Limit of counts field (available range is from 1 to 2
32
-1 of counts)
Controling data acquisition
3.5
Controling data acquisition
Acquisition parameters panel is a frame placed at the right side of a screen in
HARDWARE and ANALYZER module. It contains fields describing actual status
of acquisition, its parameters and the spectrum preview window at the bottom of
it; all data are periodically refreshed: 
Acquisition status:  
ON – acquisition is running, 
OFF – acquisition is stopped or not started, second field below contains
message describing cause of acquisition stop
Time acquisition time in sec:  
Real: duration of the acquisition run (in seconds)
58


59
Measurement
Live:  global amount of time (in seconds) when the device is active, i.e.
the time when ADC is active or ready
Dead [%]:  proportional ratio of subtraction of real and live time to real
time
CPS: (counts per second) number of counts registered by the analyzer in a
period of 1 second of real time 
Stop criteria  acquisition time control parameters:
Live Time or Real Time - when set, the acquisition is stopped after
expiration of selected time defined in Limit[s] 
Counts in ROI -  when set, the acquisition is stopped after reaching count
limit criterion in a reagion defined by  From, To and Limit 
Integral of counts - actual value of counts in selected region
Measurement session: control panel of 
measurement session
 
In ANALYZER module, if acquisition is switched off or when the user does not
pay attention to current acquisition control, he can remove (hide) the acquisition
panel from the screen.
Hide acquisition parameters panel 
menu: View | Measurement panel
Measurement control in MCS mode
Start/Stop/Reset of data acquisition
Acquisition automatic stop criteria
3.6
Continuation of acquisition after program shut
down
After starting the acquisition, the program can be closed. Exiting the program
does not result in acquisition stop – acquisition once started runs autonomous
until one circumstance will occur:
exceeding limit of automatic stop criterion (time flow or counts in region)
counts overflow in a channel.
The program, when started, recognizes actual status of acquisition. If is on,
actual parameters of current acquisition will be displayed.
Hardware setup
3.7
Measurement session
Term 'Measurement session' defines sequence of common measurements
performed in the same conditions and at the same analyzer parameter settings.
Each cycle consists of the following steps:
·
acquisition start
59
68
56
57
48


60
TUKAN 8k
·
acquisition stop after defined period
·
saving spectrum to a file
·
removing spectrum from the buffer of the analyzer
·
acquisition restart after defined period (break)
Measurement session command can be performed in HARDWARE and
ANALYZER modules from Measurement menu or from context menu of
acquisition parameters panel.
Configurable parameters of measurement session are as follows:


61
Measurement
Start type: - manual, performed when acquisition is started by e.g. selecting
'Acquisition start' icon, or scheduled, performed at selected date and time
Sweeps count: - number of sequential measurements (which defines at the
same time a number of spectra files being saved to a disk)
Length of sweep time: - time limit selected in this place is transferred
automatically to the set of parameters
 
Acquisition automatic stop criteria
 and
is obligatory to all measurement, not necessary related to measurement session
Break between sweeps:- period between end of last and begin of next
acquisition
Saving acquisition spectra to separate files 
Spectra files: - sequential spectra are saved to files with a given name
composed with Base name of a file
 
followed by current number of sweep. The
files are saved in a folder defined in Folder name box.
Note: spectra can be saved in binary format (*.wdm) or in any of available ASCII
formats (see.
 
Export to ASCII format
 
).
Saving numerical result parameters of acquisition in separate text files
57
86


62
TUKAN 8k
File with mathematical results tab provides possibility to save additional
information about acquisition after each end of measurement session. In a form
opened when Parameter selection button is pressed, the user can define set of
regions of a spectrum. For each region set of analyze parameters can be
defined. Those parameters are calculated and saved to specified ASCII file after
finish of every sweep.
Generally, for 
acquisition performed in MCS mode of work
, the measurement
sessions are defined in the same manner. Spectra are saved to disk as “.wds”
files and set of available parameters is different in comparison to MCA mode of
work.
Measurement
3.8
Measurement spectrum
"Acquisition" spectrum
Displaying "Live" spectrum
Time backup of acquisition data
Spectrum 'ACQUISITION"
3.8.1
The spectrum „Acquisition" is created automatically by the program. This is a
spectrum which collects data taken directly from the buffer of the analyzer card. 
In 
spectra container
 is always on upper position – it can not be moved or
deleted from it.
66
48
62
63
64
81


Yüklə 6,31 Kb.

Dostları ilə paylaş:
1   ...   12   13   14   15   16   17   18   19   ...   33




Verilənlər bazası müəlliflik hüququ ilə müdafiə olunur ©genderi.org 2024
rəhbərliyinə müraciət

    Ana səhifə